3. Installation
3.2 Removal of the Transport Lock
The instrument is delivered with the plate support locked into place, so that it can
not be damaged.
Before the instrument can be used the transport lock must be removed using the
following procedure:
1. Ensure that the instrument is disconnected from the mains power supply.
2. Open the plate compartment flap.
3. Using the 2mm Allen key supplied, remove the two screws holding the
transport lock in place. (See diagram below).
2 screws
4. Remove the transport lock.
Note:
The transport lock should be saved for further transportation purposes.
3-2
Operating Manual for GENios, GENios FL, and GENios
Plus
No: I 112 904 Rev No: 1.1
June 2002