Quick Guide
Refilling Disposable Tips
22
Refilling Disposable Tips
The
)(
Application Software recognizes the DiTi fill
status. The buttons representing empty racks are red.
1.
The “Refill DiTis” dialog box is displayed by clicking on the
appropriate icon
n
.
To refill a DiTi Rack before it is empty, click the corre-
sponding button.
2.
Remove the DiTi Carrier
o
.
3.
Remove the empty DiTi Trays
p
.
Replacing the empty DiTi trays with full ones
4.
Put the right side of the Rack on the two bolts and move it
to the right as far as it will go
q
.
5.
Press the left side down until the Rack clicks into place
(two springs).
CAUTION
•
Do not start the following steps before the
“Refill DiTis” dialog box is displayed.
•
Do not open Front Safety Cover as this will stop
all instrument activities.
2
1
CAUTION
Make sure the Trays are placed properly.
Front Side
Marking of
the left side
3
4