7. Maintenance and Cleaning
7.5.2 Rinse:
Day
This procedure is used to rinse the wash solution system if the instrument is to be
left standing for a longer time. Alternatively it can be used to rinse the wash
solution system after the instrument has been left for a long time.
With this procedure you can use either distilled water or wash solution.
If the instrument is left to stand for 1 hr to 1/2 day, wash solution can be used, for
longer times use distilled water.
Confirm rinsing solution and tube in bottle
Select and confirm Day Rinse procedure
Select required channel (1 - 4)
Rinse: Day
- + other yes
Day: CH1
- + exit yes
Rinse Solution ?
exit yes
Waste Bottle OK?
exit yes
Confirm waste bottle is empty
Soaking
Stop
The waste sink is filled with solution and
the manifold is lowered into it
Manifold left in solution until
Stop
key pressed
Run XX:yyyyyyy
- + other yes
Standby mode
Washing Procedure can be performed
Rinse ChX YY sec
Stop
Wash system rinsed for set rinse time
May 2003
Instructions for Use for Columbus Washer No. I 109 006 Rev No. 2.2
7-11