7. Maintenance and Cleaning
7.2 Replacing the Mains Fuses
WARNING
TO PREVENT THE RISK OF FIRE, THE MAINS FUSES SHOULD ONLY BE
REPLACED WITH THE SAME TYPE AND RATING OF FUSE.
The following steps must be performed to replace the mains fuses, which are
located by the power cable connection, in the rear panel of the instrument.
1.
Switch off the instrument and unplug the power cord.
Fig. 7-7 The Fuse Compartment
l
0
Fuse
Compartment
2. Open the fuse compartment, by inserting a screw driver into the slot in the
right hand side of the compartment. Remove the compartment out and to the
right.
Fig. 7-8 The Fuses
l
0
Fuses
3. Remove the fuses and replace them.
4. Ensure that the fuses have the correct rating.
T 1,6 A / 250 V (slow blow)
5. Replace the fuse compartment.
6. Reconnect the power cord and switch the instrument on
WARNING
IF THE FUSE CONTINUES TO BLOW, PLEASE CALL FOR SERVICE.
May 2003
Instructions for Use for Columbus Washer No. I 109 006 Rev No. 2.2
7-7