6. Adjust Menu
6.3 Adjust Liquid Sensors (Optional)
Caution
This procedure must be performed when the instrument is installed.
This procedure allows a warning level to be determined. This informs the
user when the level of liquid in the bottles is too low or too high.
When the level of liquid in the liquid bottles 1 & 2 is below 600ml and the level of
the liquid in the waste bottle is above 2000ml the instrument will emit an
accoustic signal to alert the user.
Fill the bottles before performing the adjustment procedure. The bottles marked
green (wash solution) and gray (rinse solution) should be filled with 600ml of
water and the bottle marked red (waste) should be filled with 2000ml of water.
Place all three bottles on the corresponding colored bottle holders. Go to the
Adjust menu and select Adj. Liq. Sens.
Liq. Sens. 1
- + exit yes
XX = displayed sensor value (does not directly
correlate to the volume of liquid in the bottles).
1 = values above limit
Press
yes
to confirm the limit.
XX = displayed value
0 = values below limit
Press
exit
to continue the adjustment of other
sensors.
Using the - and + keys select the liquid sensor.
1 = Green (wash solution)
2 = Gray (rinse solution)
W = Red (waste)
Press
yes
to continue the adjustment procedure.
Empty 1 XX 1
exit yes
Empty 1 XX 0
exit yes
Adj. Liq. Sens.:
other yes
Press
yes
to start the adjustment procedure
Use the cursor keys to repeat this procedure
for the grey (rinse solution) and red (waste)
bottles. See overleaf for more details.
6-6
Instructions for Use for Columbus Washer No. I 109 006 Rev No. 2.2
May 2003