4. Operating Instructions
4.5.2
Instrument with Liquid Sensor (Optional)
Waste Bottle Full at Start
When the selected program is confirmed, the instrument checks if the waste
bottle is full. If it is full, the following message is displayed:
Wastebottle Full
yes
Press
yes
to remove the message and return to the start and the following
message is displayed:
Run XX
- + other yes
Empty the waste bottle and restart the program.
Waste Bottle Full during Procedure
If the waste bottle becomes full during the washing procedure, the following
message is displayed after the instrument has completed the washing procedure:
Wastebottle Full
yes
Note:
The instrument does not stop a procedure when the waste bottle becomes full.
Therefore, ensure that the waste full sensor is calibrated, so that there is
enough space in the waste bottle to accept the remaining solution.
(See chapter 6)
Press
yes
to remove the message and the instrument proceeds to the standby
mode and the following message is displayed:
Run XX
- + other yes
Empty the waste bottle and the instrument is ready to start the next procedure.
Liquid Bottle Empty
If the liquid bottle is empty before a program is started, the following message is
displayed:
Liq. Empty Ch: X
yes
Fill the liquid bottle and press
yes
to remove the message and continue with the
program.
May 2003
Instructions for Use for Columbus Washer No. I 109 006 Rev No. 2.2
4-11