3 - Software Communication
Communication Protocols
Cavro® XLP 6000 Modular Syringe Pump Operating Manual, 734237-C, en
3 - 19
Note: For query commands, no acknowledge frame is needed.
Report Command
The host sends command report 23 of frame type 6 to a pump, and the pump is
set to address 1.
Note: For a multi-frame reply, the start frame is type 3, the middle frame is type 4,
and the last frame is type 6.
Host sends:
0
XXXXXXX
110
0
0010
29
Dir
Slave ID
Frame Type
RTR
DLC
Data bytes
Pump reports:
1
XXXXXXX
110
0
0010
<20h><60h>
Dir
Slave ID
Frame Type
RTR
DLC
Host sends:
0
XXXXXXX
110
0
0010
23
Dir
Slave ID
Frame
Type
RTR
DLC
Data bytes
Pump reports:
1
XXXXXXX
011
0
1000
<20h><60h><XLP60>
Dir
Slave ID
Frame
Type
RTR
DLC
1
XXXXXXX
100
0
0111
00 - Vx.x.
x<20h>x/200x (frame type 4)
<20h>734604<20h>
(frame type 4)
Rev <20h>X (frame type 6)
Dir
Slave ID
Frame
Type
RTR
DLC
Data bytes