G - 2
Cavro® XLP 6000 Modular Syringe Pump Operating Manual, 734237-C, en
G - Command Quick Reference
Initialization Commands
G.2
Initialization Commands
Command
Values of <n>
Description
Z
<n
1
>
0 = initializes at full plunger force
1 = initializes at half plunger force
2 = initializes at one-third plunger force
10 – 40 = initializes at the defined speed
Initializes the plunger drive
and homes the valve in a
clockwise
direction.
<n
2
>
0 = Set initialization input port to Port 1
1..X = Set initialization input port for distribution
valves, where X is the number of ports on the valve
Sets initialization input port
<n
3
>
0 = Set initialization output port to Port X
1..X = Set initialization output port for distribution
valves, where X is the number of ports on the valve
Sets initialization output port
Y
<n
1
>
0 = initializes at full plunger force
1 = initializes at half plunger force
2 = initializes at one-third plunger force
10 – 40 = initializes at the defined speed
Initializes the plunger drive
and homes the valve in a
counterclockwise
direction.
<n
2
>
0 = Set initialization input port to Port 1
1..X = Set initialization input port for distribution
valves, where X is the number of ports on the valve
Sets initialization input ports
<n
3
>
0 = Set initialization output port to Port X
1..X = Set initialization output port for distribution
valves, where X is the number of ports on the valve
Sets initialization output port
W
<n
1
>
0 = initializes at full plunger force
1 = initializes at half plunger force
2 = initializes at one-third plunger force
10 – 40 = initializes at the defined speed
Initializes the plunger drive
only (commonly used for
valveless pumps).