Balance Kit Operating Manual, 393278, en, V1.0
4 - Putting into Operation
Setting up the Balance Kit and the SAG 285/01
ATTENTION
Do not reset the parameters unnecessarily, otherwise you will have to set them
again!
4.1.8 Balance Setup with the Instrument Software
Purpose
The purpose of this setup is to establish the proper communication with the
SAG 285/01.
Nomenclature
The Instrument Software uses the term “Balance” for the Weighing Module,
regardless of whether the Weighing Cell and the Control Unit are separated or
incorporated in the same device.
Required for
the Setup
To run the setup procedure you need the following:
Instrument Software Manual that corresponds to the installed version.
–
A PDF-file of the manual is stored on the installation CD of the Instrument
Software.
–
Operating Manual of the SAG 285/01 Weighing Module.
If version V4.7 or later is installed on the computer you need a user account
with a user name and a password.
–
Required access rights: SnS_Customer access level or higher.
–
If necessary turn to the system administrator for an appropriate user
account and the necessary access rights.
Procedure
The procedure varies slightly, depending on the version of the installed Instrument
Software: Always follow the instructions provided in the Instrument Software
Manual and/or the SAG 285 Operating Manual.
To perform the Balance Setup:
1
Switch on the computer and the SAG 285/01.
2
Tare the Weighing Cell (see SAG 285 Operating Manual).
Tab. 4-2
Required settings of balance parameters
Property
Set to
Comment
Vibration adapter
2
Medium weighing speed
Weighing process adapter
1
Suitable for small quantities
Reproducibility
Std
Standard reproducibility
Weighing unit 1
unit 1 mg
Both weighing units must be mg
Weighing unit 2
unit 2 mg
Automatic zero correction
az. off
No automatic zero correction
Automatic off
aoff -
No automatic switch off
Start mode
qu start
Quick start. No display check at
startup.