57
12-1-1-2. Stop conditions
See “Recording stopping conditions” on page 32 for
details about each setting.
Level trigger
Sets the level and UP/DOWN conditions for each
channel.
External trigger
The options are ON and OFF.
Recording time
Sets the recording time.
12-1-1-2-1. Level trigger
When Use is ON, the setting items are shown.
Logic
Set when there are multiple conditions to determine
whether one or all must be fulfilled.
AND:
The trigger requirements will be judged ful-
filled when all set conditions are met.
OR:
The trigger requirements will be judged ful-
filled when any set condition is met.
Count
Measurement stops when the set count logic is fulfilled.
CH1, CH2, CH3, CH4
Touch the icon for a channel to turn use ON/OFF for it.
Edge
UP:
The trigger requirement will be judged to
be fulfilled when the input value changes
from lower than the set level value to
higher than that value. The trigger require-
ment will not be judged to be fulfilled if the
level already exceeds the set level when
the unit becomes record ready.
DOWN:
The trigger requirement will be judged to
be fulfilled when the input value changes
from higher than the set level value to
lower than that value. The trigger require-
ment will not be judged to be fulfilled if
the level is already lower than the set level
when the unit becomes record ready.
Level (%)
The set channel range value is 100%.
12. TRIGGER settings