61
16. Specifications
Bridge voltage
DC 10 V, 5 V±10% (30mA current maximum, can be used with remote sense)
Signal input range
±5.0 mV/V
Equivalent input/
TEDS
Calibration
range
0.05 mV/V – 5.0 mV/V
Calibration
precision
Within 0.1% F.S. (when using 1m standard TEAC Ø8, 6-core shielded cable with 350Ω
impedance, 10V BV and 5mV/V setting)
Precision
Linearity
Within 0.01% F.S. +1 digit (when input is 5 mV/V)
Zero drift
Within 0.5 μV/°C (input conversion value)
Gain drift
Within ±0.005% F.S./°C
A/D conversion
24-bit, 20000 times/second
Digital filter
Select 3, 10, 30, 100, 300, 1000 Hz(−6 dB/oct) or none
D/A output**
4000 times per second, isolated output, ±1–±10V voltage output (set in 1V steps) and
about 1/59000 resolution (when set to ±10 V), or 4–20mA current output and about
1/43000 resolution
TEDS function
IEEE1451.4 class 2 mix mode interface
Indicator value
Display range
−99999 to 99999
Decimal point
Display position selectable
Times displayed
Select 4, 6, 10 or 20 times/second
Setting items
Calibration
settings
Zero Balancing, TEDS Calibration, Actual Load Calibration, Equivalent Input Calibration
Function
settings
High limit, low limit, comparison mode, hysteresis, nearly zero, moving average, low pass
filter, motion detect, zero tracking, static strain, digital zero, digital offset, zone definition,
hold mode, lock, minimum grid, display times, bridge voltage, digital zero limit, clear
digital zero, comparison output pattern, select data output**, D/A converter**, remote
sense
Hold functions
Sample hold, peak hold, bottom hold, zone definition hold (peak, bottom)
External input
and output
signals
Input
Hold, clear, digital zero (isolated from main unit circuits using a photocoupler)
Output
HI, OK, LO open collector output (isolated from main unit circuits using a photocoupler)
RS-485
*
A+, B− (isolated from main unit circuits using a photocoupler), TRM, SG
Power supply
Ratings: 24V DC ±10%, 7 W
Operating temperature range
0°C to 40°C
Storage temperature range
−20°C to 60°C
Operating humidity range
85% RH or less (without condensation)
Applicable standards
CE marking, FCC (Class A), UL61010-1, UKCA
External dimensions (W × H × D)
Approximately 30 mm × 85 mm × 110 mm (without protrusions)
Weight
About 210 g
*TD-SC1 (485) only
**TD-SC1 (D/A) only
o
Specifications and appearance are subject to change without notice.
o
Weight and dimensions are approximate.
o