57
13. TEDS settings
By connecting a sensor that supports IEEE1451.4 Transducer
Electronic Data Sheet (TEDS) to the unit, the rated output
stored in the sensor can be loaded and this can be applied to
calibration of the indicator with this function.
In addition, the value calibrated by the unit itself can be loaded
to the TEDS sensor and its original value can be restored.
These functions are supported by TEDS sensors that comply
with the standard shown in the table below.
Note, however that equipment with TEDS memory includes
both 1kbit and 4kbit devices, but the unit only supports 4kbit.
TEDS standard
IEEE
1451.4
(V0.9)
IEEE1451.4 (V1.0)
Template ID
Bridge
Sensors (33)
Strain Gauge
(35)
Other
-
-
TEDS calibration, TEDS data writing and TEDS data resto-
ration supported
TEDS calibration supported
-
Not supported
13-1. Rewrite TEDS Data
The current calibration values (rated output and rated capac-
ity) and calibration date will be written to the TEDS memory.
The indicator unit will not be written.
13-2. Restore TEDS data
This restores the factory default calibration values of a product
using the data from “Rewrite TEDS Data” above.
13-3. Read TEDS data
Values can be read by executing the following commands
from the TEDS data display group.
Serial number (6001)
Maximum rated capacity (6002)
Maximum rated output (6003)
Sensor Impedance (6004)
Maximum excitation level (6005)
Calibration date (6006)
Model number (6007)