9-4. Digital offset
The set value is subtracted from the measured value (digital
zero).
9-5. Digital Zero
9-5-1. Enable Digital Zero
When Enable Digital Zero is set to OFF, the digital zero function
will not be executed even if the D/Z signal is ON for the control
input terminal.
9-5-2. Digital Zero Limit Value
This sets the digital zero adjustment range. (The setting value
unit is the same as for the indicator value.)
ATTENTION
If the current sensor input value exceeds the digital zero limit
setting value, the result of Digital Zero Limit Value command
execution will end with an error and the indicator value will not
become zero.
9-5-3. Clear Digital Zero
The display adjustment that has been made with Digital Zero
is reverted so that the unadjusted value is shown again.
When zero balancing has been conducted, the value is shown
as zero.
9-6. Static Strain Display Switching
(static strain display mode)
When a Static Strain Display Switching execution command is
issued, the input signal is shown using the strain amount unit
(μST) (page 27).
Use when checking sensor output and when investigating
unsteadiness in indicator values, including unsteadiness
caused by sensors and cables, as well as for other problems.
NOTE
o
To exit static strain display mode, issue a Switch to Home
Screen execution command.
o
Static strain is shown using 1-gauge method with a gauge
factor of 2.0.
9-7. Select Data Output
Select the data output from the D/A converter of the TD-SC1
(D/A).
Select Data Output has the following two options.
The hold value is output in coordination with the display.
The input is output as is.