41
8-5. Remote Sense
Before connecting the sensor, set which type will be con-
nected in advance.
Options: Used (6-wire format)
Unused (4-wire format)
ATTENTION
Incorrect connections or settings could cause damage to sensors.
8-6. Minimum Scale Selection
Set the minimum digital change of the indicator value.
Options: 1, 2, 5, 10
8-7. Display Frequency Selection
Select the number of times that the indicator value is shown
per second.
Options: 4, 6, 10, 20
8-8. Maximum Display Value
Set the highest displayed value.
The default value is set to 110% of the rated capacity.
If this value is exceeded, the STATUS indicator will light magenta.
ATTENTION
The input value will be checked with a value converted to
match the input signal.
If the value converted to match the input exceeds 5.0 mmV/V,
the setting value will be adjusted to be equivalent to 5.0 mmV/V.
The default value will be checked in the same manner and
adjusted if necessary.
8-9. Sensor Input Logic
The sensor input logic can be reversed artificially.
Normally, “Standard” should be used.
NOTE
o
“Reversed” does not electrically reverse the input.
o
After changing this setting, always redo zero.
8-10. Zero Balancing
You can conduct zero balancing even without resetting the
calibration value.
1
Set Calibration Value Lock to OFF.
See “8-1-1. Locking and unlocking calibration values” on
page 36.
2
With no load on the sensor, press and hold the
ZERO button or send a Zero Balancing command.
During zero balancing, the STATUS indicator will become
magenta.
3
Set Calibration Value Lock to ON.
ATTENTION
Zero balancing cannot be conducted during static strain display.
8-11. Zero Point Input Calibration
Set the zero point output value.
o
Any desired output value can be set as the zero point.
8-12. Digital Zero
The current indicator value becomes zero.
o
The range in which the digital zero can be adjusted is the
range set by the Digital Zero Limit Value.
o
Digital zero is cleared by turning off the power supply or
using Clear Digital Zero.
ATTENTION
The digital zero function cannot be executed during static
strain display.
8. Calibration