70
14. About TEDS
Connecting a sensor that conforms to IEEE1451.4 (Transducer
Electronic Data Sheet (TEDS)) to the TD-260T, you can load the
rated output values that are saved on the sensor into the TD-
260T and use the values for calibration with the TD-260T.
You can also use the TD-260T to save on your TEDS sensor
values that are used for calibration with the TD-260T or to
restore the TEDS sensor's factory settings.
TEDS memory is available in 1-kilobit and 4-kilobit sizes. Note
that the TD-260T supports the 4-kilobit memory only.
TEDS standard
IEEE
1451.4
(V0.9)
IEEE1451.4(V1.0)
Template ID
Bridge
Sensors
(33)
Strain gage
(35)
Others
: TEDS calibration, Save TEDS Data (F4-6), and Restore
TEDS Data (F4-7) all supported
: TEDS calibration supported
: None of the above functions supported
14-1. Save TEDS Data
You can save the current calibration values into the TEDS
memory.
1. Enter the setup screen number F4-6.
Press 4 times
Press 5 times
2. To enter 00015, change the blinking digit of the
value, press
5
or
b
. To move to the previous or next
digit, press
g
or
t
.
To change the digit, press:
To move to the previous
or next digit, press:
NOTE
<
Be sure to enter this value to prevent the original TEDS data
from being overwritten by mistake.
<
If you press ESC while “tEdS” is not displayed, the setting is
interrupted, and then the TD-260T exits the setup mode.
3. Press ENTER. The TD-260T displays “tEdS” and starts
verifying data in the TEDS memory,
4. To enter a calibration date;
enter a year by pressing
g
,
t
,
5
, or
b
, and then
press ENTER. For month and day settings, too, press
ENTER after you enter values.
When the date setting is completed, the TD-260T starts
saving the calibration values.
Enter year and press
Enter month and press
Enter day and press
NOTE
If you press ESC while “tEdS” is not displayed, the setting is
interrupted, and then the TD-260T exits the setup mode.
5. The display shows “tEdS” during the save process.
6. The TD-260T displays “F4-6,” showing that the save
process is completed, and then waits for a key to be
pressed.
To exit the setup mode, press ESC.