7-3-2. Area output
Each decision output switches on when the following
conditions are met.
(with hysteresis set to 0)
HH: HH value < reading
HI: HI value < reading < HH value
LO: LL value < reading < LO value
LL: Reading < LL value
Time
Reading
OFF
ON
HH
OFF
ON
HI
OFF
ON
OK
OFF
ON
LO
OFF
ON
LL
Decision output
HH
HI
LO
LL
NOTE
The operations of decision outputs vary depending on the
setting for Hysteresis (F1-6). For detailed information, see the
section 7-2. “Hysteresis.”
1. Enter the setup screen number F1-15.
Press once
Press once
2. To select a comparative output pattern,press
5
or
b
.
or
The following two settings are available.
0.nor: Standard output
1.ArE: Area output
3. To save the selection, press ENTER, and then to exit
the setup mode, press ESC.
To save, press:
7. Setting Functions