Strain gauge load cell Instructions for Use

TC-XR(T)-G6

TC-KR(T)-G6

Compression Load Cell

Introduction

Thank you for purchasing the TC-XR(T)-G6/TC-KR(T)-G6 
load cell.
Please read this document completely before using this 
load cell to achieve its best performance and ensure safe 
and proper operation.

Included accessories

If anything is missing or damaged, contact the retailer 
where you purchased the product.

Test report × 1
Instructions for Use (this document) × 1

 

o

Company names and product names in this document are the 
trademarks or registered trademarks of their respective owners.

IMPORTANT SAFETY INSTRUCTIONS

V

WARNING

If something abnormal occurs

In the unlikely event that the product produces smoke, a 
strange smell or noise, for example, continuing to use it in 
this abnormal state could cause fire or electric shock. After 

cutting off the power, confirm that smoke is no longer being 
produced. Then, request repair from the retailer where you 
purchased the product.

Do not open the cover.

Never remove the cover from this unit. Doing so could 
cause electric shock. Request inspection and repair from 
the retailer where you purchased the product. Do not 
alter this unit. Doing so could cause fire or electric shock.

Do not put foreign objects or water, for example, 
into the unit.

Do not place a container that holds water, for example, 
on top of this unit. Liquid overflowing or entering the unit 
could cause fire or electric shock.

Do not use the unit with any power supply voltage 
other than that specified.

Do not use the unit with any power supply voltage other than 

that specified. Doing so could cause fire or electric shock.

V

CAUTION

Unsuitable installation locations

Do not place the unit in the following types of locations. 
Doing so could cause fire or electric shock.

 

o

Locations where it might be exposed to smoke or steam, 
such as near a kitchen table or humidifier

 

o

Unstable locations, including unsteady stands and 
tilted places

 

o

Location that are very humid or dusty

 

o

Locations that are exposed to direct sunlight

When not using the unit for a long time

For safety, cut the power supply when not using this unit 
for a long time.

Do not operate a damaged unit.

Precautions for use

 

o

This unit is not built to be water or splash resistant, 
and it cannot be used in conditions when the relative 
humidity is high. Moreover, use in atmospheres with 
corrosive gases should be avoided.

 

o

Avoid use in conditions where condensation could occur.

 

o

Connect cores to the load cell after discharging (elim-
inating) static electricity from your body.

 

o

Be careful to prevent water, oil and other substances 
from getting on the unit.

 

o

If the surrounding temperature changes suddenly, the 
values output by this device could become unstable, 
making accurate measurement impossible. (This could 
occur, for example, in a location blown by warm or cold air.)

Installation procedures

 

o

Install this unit in a place where the structure is level 
and can sufficiently bear the load being used.

 

o

Screw holes for installation are located in four places 
each on the top and bottom surfaces.

 

o

The top and bottom screw hole dimensions are different 
according to the load cell rated capacity. The screw hole 
dimensions are shown in the tables below.

TC-KR(T)-G6

Rated 
capacity

Screw hole 
dimensions

Screw hole 
depth

Tightening 
torque

5 kN

M4X0.7

8 mm

3.6 N·m

10 kN

M5X0.8

8 mm

7 N·m

20 kN

M5X0.8

8 mm

7 N·m

30 kN

M5X0.8

8 mm

7 N·m

50 kN

M5X0.8

8 mm

7 N·m

100 kN

M5X0.8

8 mm

7 N·m

200 kN

M8X1.25

12 mm

29.5 N·m

300 kN

M8X1.25

15 mm

29.5 N·m

TC-XR(T)-G6

Rated 
capacity

Screw hole 
dimensions

Screw hole 
depth

Tightening 
torque

20 kN

M5X0.8

8 mm

7 N·m

50 kN

M5X0.8

8 mm

7 N·m

100 kN

M5X0.8

8 mm

7 N·m

200 kN

M8X1.25

12 mm

29.5 N·m

300 kN

M8X1.25

15 mm

29.5 N·m

Precautions when placing loads on the 

unit

 

o

If the load receiving area is contacted by something that 
is at a different temperature and the load is increased, 
the values output by this device could become unstable, 
making accurate measurement impossible. In such a 
case, wait until the temperature difference ceases to 
exist before measuring.

 

o

Make sure the load is perpendicular to the surface to 
which this unit is attached.

 

o

Place the load so that it is centered on the center of the 
unit. If the load is not centered (eccentric load), twisting, 
for example, and measurement errors could occur. This 
could even result in damage.

F is the correct load orientation
F1 is an eccentric inclined load
F2 is an eccentric load

F1

F F2

 

o

Be careful to avoid turning and twisting from lateral 
loads. This could cause troubles like those described 
in the previous item.

 

o

Be careful to avoid applying loads that exceed the 
rated capacity. In particular, use caution when there are 
vibrations because loads that exceed the rated capacity 
could occur due to sympathetic vibrations, for example.

 

o

These products are made with coaxiality and perpen-
dicularity tolerances of 0.02 mm for the center hole and 
the top and bottom surfaces. When using them, please 
consider attachment methods as well as jigs and other 

parts being used so that coaxiality and perpendicularity 
tolerances of 0.02 mm are maintained.

 

o

Use load cells so that loads are distributed evenly across 
the entire surfaces of both the top and bottom.

Electrical connection of load cell with 

built-in TEDS

 

o

Connect as shown in the illustration below. Incorrect con-
nections could result in inability to balance and in errors 
occurring in the output voltage when loads are applied.

Using a cable with bare lead wires

Red 

Input (+)

Black 

Output (−)

Blue 

Input (−)

White  Output (+)

Orange  TEDS signal (+)

Green  TEDS signal (COM)

Yellow Shield

TEDS

 

o

This unit has a built-in TEDS function.

 

o

The orange and green cores in the cable and the F and 
G pins in the connector are wired for TEDS.

 

o

This unit does not support remote sensing.

 

o

Since products that support remote sensing use the 
same cables and connector pins as TEDS, be careful not 
to try mistakenly to use these pins for remote sensing.

 

o

See the operation manuals of indicators and strain amps 
that support remote sensing for how to connect sensors 
with those units.

 

o

The shield is not connected to the main body of this 
product. For this reason, if grounding is necessary because 

of external noise, for example, arrange to ground the 
shield to a part other than the body of this unit.

 

o

Since the cable is directly connected to this unit, use a 
specialized cable to increase the length. (Please consult 
with us.)

 

o

When conducting insulation resistance tests, limit them 
to the red, black, blue and white cores. Do not apply to 
the TEDS cores (orange and green).

Using a connector (optional)

A  Input (+)

B  Output (−)

C  Input (−)

D  Output (+)

F  TEDS signal (+)

G  TEDS signal (COM)

E Shield

NDIS-compatible connector

TEDS

Precautions when using during calibra-

tion or as a master load cell

Please take the following precautions when calibrating one of 
these products yourself or when using one as a master load cell.

 

o

We recommend attaching plates to the top and bottom 
sides and using them when applying loads.

 

o

Consider rigid materials and their hardness for the 
plates used for top and bottom so that they are able 
to maintain the flatness and parallelism of the load cell 
in relation to the load.