Strain gauge load cell Instructions for Use

TC-XR(T)-G6

TC-KR(T)-G6

Compression Load Cell

Introduction

Thank you for purchasing the TC-XR(T)-G6/TC-KR(T)-G6 
load cell.
Please read this document completely before using this 
load cell to achieve its best performance and ensure safe 
and proper operation.

Included accessories

If anything is missing or damaged, contact the retailer 
where you purchased the product.

Test report × 1
Instructions for Use (this document) × 1

 

o

Company names and product names in this document are the 
trademarks or registered trademarks of their respective owners.

IMPORTANT SAFETY INSTRUCTIONS

V

WARNING

If something abnormal occurs

Request repair from the retailer where you purchased 
the product.

Do not open the cover.

Never remove the cover from this unit. Doing so could 
cause malfunction. Request inspection and repair from the 
retailer where you purchased the product. Do not alter this 
unit. Doing so could cause malfunction.

Do not put foreign objects or water, for example, 
into the unit.

Do not place a container that holds water, for example, on 
top of this unit. If liquid is spilled, for example, and enters 
the unit, this could cause malfunction.

Do not use the unit with any power supply voltage 
other than that specified.

Do not use the unit with any power supply voltage other 
than that specified. Doing so could cause malfunction.

V

CAUTION

Unsuitable installation locations

Do not place the unit in the following types of locations. 
Doing so could cause malfunction.

 

o

Locations where it might be exposed to smoke or steam, 
such as near a kitchen table or humidifier

 

o

Unstable locations, including unsteady stands and 
tilted places

 

o

Locations that are very humid or dusty

 

o

Locations that are exposed to direct sunlight

When not using the unit for a long time

For safety, cut the power supply when not using this unit 
for a long time.

Do not operate a damaged unit.

Precautions for use

 

o

This unit is not built to be water or splash resistant, 
and it cannot be used in conditions when the relative 
humidity is high. Moreover, use in atmospheres with 
corrosive gases should be avoided.

 

o

Be careful to prevent water, oil and other substances 
from getting on the unit.

 

o

Avoid use in conditions where condensation could occur.

 

o

Connect cores to the load cell after discharging (elim-
inating) static electricity from your body.

 

o

If the surrounding temperature changes suddenly, the 
values output by this device could become unstable, 
making accurate measurement impossible. 
(This could occur, for example, in a location blown by 
warm or cold air.)

 

o

If a cable of this unit needs to be bent and shifted, make 
the curvature of the bent part at least 60 mm. Do not 
apply tension to the cable.

 

o

Conduct load calibrations periodically.

Installation procedures

 

o

Install this unit in a place where the structure is level 
and can sufficiently bear the load being used.

 

o

Screw holes for installation are located in four places 
each on the top and bottom surfaces.

 

o

The top and bottom screw hole dimensions are different 
according to the load cell rated capacity. The screw 
dimensions are shown in the tables below.

TC-XR(T)-G6

Rated 

capacity

Screw 

dimensions

Screw hole 

depth

Tightening 

torque

20 kN

M5X0.8

7 mm

7 N·m

50 kN

M5X0.8

7 mm

7 N·m

100 kN

M5X0.8

7 mm

7 N·m

200 kN

M8X1.25

10 mm

29.5 N·m

300 kN

M8X1.25

12 mm

29.5 N·m

TC-KR(T)-G6

Rated 

capacity

Screw 

dimensions

Screw hole 

depth

Tightening 

torque

5 kN

M4X0.7

8 mm

3.6 N·m

10 kN

M5X0.8

8 mm

7 N·m

20 kN

M5X0.8

8 mm

7 N·m

30 kN

M5X0.8

8 mm

7 N·m

50 kN

M5X0.8

8 mm

7 N·m

100 kN

M5X0.8

8 mm

7 N·m

200 kN

M8X1.25

12 mm

29.5 N·m

300 kN

M8X1.25

15 mm

29.5 N·m

Precautions when placing loads on the unit

 

o

If the load receiving area is contacted by something that 
is at a different temperature and the load is increased, 
the values output by this device could become unstable, 
making accurate measurement impossible. In such a 
case, wait until the temperature difference ceases to 
exist before measuring.

 

o

Make sure the load is perpendicular to the surface to 
which this unit is attached.

 

o

Place the load so that it is centered on the center of the 
unit. If the load is not centered (eccentric load), twisting, 
for example, and measurement errors could occur. This 
could even result in damage.

F is the correct load orientation
F1 is an eccentric inclined load
F2 is an eccentric load

F1

F F2

 

o

Be careful to avoid turning and twisting from lateral 
loads. This could cause troubles like those described 
in the previous item.

 

o

Be careful to avoid applying loads that exceed the 
rated capacity. In particular, use caution when there are 
vibrations because loads that exceed the rated capacity 
could occur due to sympathetic vibrations, for example.

 

o

Make sure coaxiality and perpendicularity tolerances 
of 0.02 mm are maintained for the center hole and the 
top and bottom surfaces of these products. When using 
them, please consider attachment methods as well as 
jigs and other parts being used so that coaxiality and 
perpendicularity tolerances of 0.02 mm are maintained.

 

o

Use load cells so that loads are distributed evenly across 
the entire surfaces of both the top and bottom.

Electrical connection of load cell with 

built-in TEDS

 

o

Connect as shown in the illustration below. Incorrect con-
nections could result in inability to balance and in errors 
occurring in the output voltage when loads are applied.

Using a cable with bare lead wires

Red 

Input (+)

Black 

Output (−)

Blue 

Input (−)

White  Output (+)

Orange  TEDS signal (+)

Green  TEDS signal (COM)

Yellow Shield

TEDS

 

o

This unit has a built-in TEDS function.

 

o

The orange and green cores in the cable and the F and 
G pins in the connector are wired for TEDS.

 

o

This unit does not support remote sensing.

 

o

Since products that support remote sense. 

 

o

See the operation manuals of indicators and strain amps 
that support remote sense for how to connect sensors 
with those units.

 

o

The shield is not connected to the main body of this 
product. For this reason, if grounding is necessary because 
of external noise or another issue, arrange to ground the 
shield to a part other than the body of this unit, for example.

 

o

Since the cable is directly connected to this unit, use 
a specialized cable to increase the length. (Please 
consult with us.)

 

o

When conducting insulation resistance tests, limit them 
to the red, black, blue and white cores. Do not apply to 
the TEDS cores (orange and green).

Using a connector (optional)

A  Input (+)

B  Output (−)

C  Input (−)

D  Output (+)

F  TEDS signal (+)

G  TEDS signal (COM)

E Shield

NDIS-compatible connector

TEDS

Precautions when using during calibra-

tion or as a master load cell

Please take the following precautions when calibrating 
one of these products yourself or when using one as a 
master load cell.

 

o

We recommend attaching plates to the top and bottom 
sides and using them when applying loads.

 

o

Consider rigid materials and their hardness for the 
plates used for top and bottom so that they are able 
to maintain the flatness and parallelism of the load cell 
in relation to the load.

 

o

Moreover, by making the load-receiving surface of the 
top plate curved, the affects of unbalanced loads, for 
example, can be reduced, enabling accurate measurements.

 

o

Refer to the following illustration as an example of use 
when considering top and bottom plates.