Strain gauge load cell Instructions for Use

TC-FSRSP(T)-G3

Compression Load Cell

Introduction

Thank you for purchasing the TC-FSRSP(T)-G3 load cell.
Please read this document completely before using this 
load cell to achieve its best performance and ensure safe 
and proper operation.

Included accessories

If anything is missing or damaged, contact the retailer 
where you purchased the product.

Test report × 1
Instructions for Use (this document) × 1
Curved load button × 1
Flat load button × 1

 

o

Company names and product names in this document are the 
trademarks or registered trademarks of their respective owners.

IMPORTANT SAFETY INSTRUCTIONS

V

WARNING

If something abnormal occurs

In the unlikely event that the product produces smoke, a 
strange smell or noise, for example, continuing to use it in 
this abnormal state could cause fire or electric shock. After 
cutting off the power, confirm that smoke is no longer being 

produced. Then, request repair from the retailer where you 
purchased the product.

Do not open the cover.

Never remove the cover from this unit. Doing so could 
cause electric shock. Request inspection and repair from 
the retailer where you purchased the product. Do not 
alter this unit. Doing so could cause fire or electric shock.

Do not put foreign objects or water, for example, 
into the unit.

Do not place a container that holds water, for example, 
on top of this unit. Liquid overflowing or entering the unit 
could cause fire or electric shock.

Do not use the unit with any power supply voltage 
other than that specified.

Do not use the unit with any power supply voltage other than 
that specified. Doing so could cause fire or electric shock.

V

CAUTION

Unsuitable installation locations

Do not place the unit in the following types of locations. 
Doing so could cause fire or electric shock.

 

o

Locations where it might be exposed to smoke or steam, 
such as near a kitchen table or humidifier

 

o

Unstable locations, including unsteady stands and 
tilted places

 

o

Location that are very humid or dusty

 

o

Locations that are exposed to direct sunlight

When not using the unit for a long time

For safety, cut the power supply when not using this unit 
for a long time.

Do not operate a damaged unit.

Precautions for use

 

o

This unit is not built to be water or splash resistant, 
and it cannot be used in conditions when the relative 
humidity is high. Moreover, use in atmospheres with 
corrosive gases should be avoided.

 

o

Be careful to prevent water, oil and other substances 
from getting on the unit.

 

o

Avoid use in conditions where condensation could occur.

 

o

Connect cores to the load cell after discharging (elimi-
nating) static electricity from your body.

 

o

If the surrounding temperature changes suddenly, 
the values output by this device could become 
unstable, making accurate measurement impossible.  
(This could occur, for example, in a location blown by 
warm or cold air.)

 

o

Conduct load calibrations periodically.

Installation procedures

To use this product, install one of the included load buttons 
(two types) or a part that you have prepared in the central 
M2 internal thread as shown in the illustration below.

 

o

Tighten the included load button by hand to the position 
where the screw stops.

 

o

Since it could loosen during use, we recommend coating 
the screw part with threadlocker. Doing so, however, 
will make removing the load button more difficult.

  Curved load button  Flat load button

 

o

When attaching a part other than one of the included 
load buttons, make certain that it does not touch the 
cover. (Refer to the dimensional drawings.)
Since the space between the cover and the attachment 
area of the loadbearing part is not always uniform, we 
recommend using a part that is smaller than the diameter 

of the component attachment area.
Furthermore, since the attached part will sink by more 
than 30 μm when a load is applied, leave a sufficient 
gap between the part and the cover.

 

o

Install this unit in a place where the structure is level 
and can sufficiently bear the load being used.

 

o

Screw holes for attachment are located in four places.

 

o

The screw hole dimensions are shown in the table below.

Model

Screw hole 

dimensions

Tightening 

torque

TC-FSRSP(T)-G3 M2.6 (3mm depth)

0.65 N·m

Installation surface

Do not use the 

central indentation 

as an installation 

surface.

Precautions when placing loads on the 

unit

 

o

Make sure the load is perpendicular to the surface to 
which this unit is attached.

 

o

When using a flat load button, use it so that the load is 
equally distributed within a Ø7mm range.

 

o

Apply the load so that it is centered on the center of 
the unit. If the load is not centered (eccentric load), 
twisting, for example, and measurement errors could 
occur. This could even result in damage.

F is the correct 
load orientation
F1 is an eccentric 
inclined load
F2 is an eccentric 

load

F1

F2

F

F1

F2

F

Curved load button 

Flat load button

 

o

Be careful to avoid turning and twisting from lateral 
loads. This could cause troubles like those described 
in the previous item.

 

o

Take extra care if the part that you have prepared is 
cylindrical. Even during installation, do not allow loads 
to be applied from the side or at an angle as shown in 
the illustration below. Doing so could cause the screw 
part to bend and damage internally.
Moreover, when securing the part, tighten with a torque 
of no more than 0.18 Nm or use a threadlocker and 
tighten gently by hand.

F

F

Bad

Bad

Part

 

o

A curved load button is used in load testing during 
production as shown in the illustration below, and the 
results are used in the TEDS data and the test report.
Since the load point, load direction and other condi-
tions differ when using a flat load button, differences in 
performance could occur compared to using a curved 
load button.

Curved load button used

4 screws (class 12.9 strength steel) in 

the bottom attached it to the jig.

0.65N·m tightening torque

F

 

o

Be careful to avoid applying loads that exceed the 
rated capacity. In particular, use caution when there are 
vibrations because loads that exceed the rated capacity 

could occur due to sympathetic vibrations, for example.

 

o

If the load receiving area (spherical surface) is 
contacted by something that is at a different tem-
perature and the load is increased, the values output 
by this device could become unstable, making accu-
rate measurement impossible.  
In such a case, wait until the temperature difference 
ceases to exist before measuring.

Electrical connection of load cell with 

built-in TEDS

 

o

Connect as shown in the illustration below. Incorrect 
connections could result in inability to balance and 
in errors occurring in the output voltage when loads 
are applied.

Using a cable with bare lead wires

Red 

Input (+)

Black 

Output (−)

Blue 

Input (−)

White 

Output (+)

Orange  TEDS signal (+)

Green 

TEDS signal (COM)

Yellow Shield

TEDS

 

o

This unit has a built-in TEDS function.

 

o

The orange and green cores in the cable and the F and 
G pins in the connector are wired for TEDS.

 

o

This unit does not support remote sensing.

 

o

Since products that support remote sensing use the 
same cables and connector pins as TEDS, be careful not 
to try mistakenly to use these pins for remote sensing.

 

o

See the operation manuals of indicators and strain 
amps that support remote sensing for how to connect 
sensors with those units.