26
Continued on the next page
e
2
Press the Update (F3) button to manually load
the TEDS data for the selected channel.
3
Press the OK (F1) button to set TEDS data as
the physical quantity conversion values.
Press the CANCEL button (F2) instead to cancel the
setting change.
8-3-2. MODULE SETTINGS
1
Turn the VALUE dial to select MODULE
SETTINGS, and press the ENTER button to open
the MODULE SETTINGS screen.
2
Use the VALUE dial and the ON/OFF (F2) button
to put check marks next to the modules to use
for measurement.
The data for modules that do not have check marks
will not be measured.
3
Select the module to set first and press the
ENTER button to open the settings screen.
4
Select the channel to set first and press the
ENTER button to open the settings screen.
Refer to the following explanations for each of the
input/output module settings.
8-3-2-1. PA Amp settings
INPUT V RANGE
Select a range that covers the input signal changes.
COUPLING
The options are DC and AC.
DC: Use when recording signals that include direct
currents
AC: Use when recording signals of 1 Hz or more
o
When using an IEPE sensor current, set Coupling
to AC.
IEPE CURRENT
Set the IEPE sensor current.
WEIGHTING
Select the weighting filter.
o
When using HPF, set WEIGHTING to FLAT.
HPF
Set the high pass filter.
CHANNEL NAME
Set the input channel signal name.
UNIT
This is the unit after physical quantity conversion. This
can be set to the desired characters.
COEFFICIENT
Set the parameters used to convert signals to physi-
cal quantities.
8. Settings
Содержание ER-LXRC1000