The PZ5M NeuroDigitizer
Fast Facts
Amp
Configuration
Test
Impedance
Preview
Waveform
Using the Front Panel Display.
The front panel touch
screen display (icons shown below) can be used for
impedance testing, waveform preview, and on-the-fly
device configuration.
Open
System Setup
screen to access wifi
and device information.
Hardware Setup.
Use the provided paired fiber optic
cable to connect the PZ5M to the RZ processor, as
shown below. The connectors are color coded and
keyed to ensure proper connections. The PZ5M-512 is
shown below. If using the PZ5M-256, connect only to
the primary connection on the RZ2 (located adjacent to
the zBUS connection).
Create a
new logical amplifier
. If all banks
are configured, the green plus sign is not
displayed.
Toggle
LED indicators
off/on.
Lock/Unlock
configuration settings
.
Touch the icon in the
logical amp
configuration area
(shown
below)
to display the desired screen for the
corresponding logical amplifier.
LED Channel Indicators Touch Screen
Caution:
Touch screen logical amp
configuration settings are overwritten at
run time when using the PZ5M_Control
macro.
Referencing
Coupling
Sample Rate
EMG
Diff (true differential) AC
750 Hz
EEG
Shared
AC
750 Hz
LFP
Shared
AC
3 kHz
SU
Local
AC
25 kHz
Power Status LEDs
This fast fact sheet provides basic information for the PZ5M.
See the
System 3 Manual for more information.
The table below lists default settings for each amp type.
PZ Input
DSP-P
Banks are color coded to indicate current
configuration.
Configuration
Color Codes
EMG
Green
EEG
Brown
LFP
Purple
SU
Teal
Not configured
Gray
Warning.
A red outline indicates that the bank is
configured as part of a logical amplifier but no headstage
is currently detected on that bank.
Software Configuration.
Logical amplifiers can be
configured on the Synapse, PZ5 options page. The table
below lists default settings for each analog input amp
type.