User Manual

Series

Programmable DC Power Supplies

Half-Rack GH1kW/1.5kW in 1U Height

Full-Rack G1kW/1.7kW/2.7kW/3.4kW/5kW in 1U Height

GSP10kW/15kW in 2U/3U Height

Built-in LAN (

 1.5), USB, RS-232 & RS-485 Interfaces

Built-in Remote Isolated Analog Program/Monitor/Control Interface

Optional Interface: IEEE488.2 SCPI (GPIB), Modbus TCP or EtherCAT

Sie haben Fragen oder wünschen eine Beratung? 

Angebotsanfrage unter

 07121 / 51 50 50

 oder über 

[email protected]

Distributed by:

Содержание GH10-100

Страница 1: ...ight GSP10kW 15kW in 2U 3U Height Built in LAN 1 5 USB RS 232 RS 485 Interfaces Built in Remote Isolated Analog Program Monitor Control Interface Optional Interface IEEE488 2 SCPI GPIB Modbus TCP or E...

Страница 2: ...mpliant LAN USB RS 232 RS 485 Interface Optional Interface IEEE488 2 GPIB MODBUS TCP or EtherCAT USER MANUAL For units equipped with MODBUS TCP interface option refer to MODBUS TCP User manual IA761 0...

Страница 3: ...This page intentionally left blank...

Страница 4: ...HB40 38 GHB300 5 GH60 25 GH600 2 6 GHB60 25 GHB600 2 6 1kW G10 100 G80 12 5 GB10 100 GB80 12 5 G20 50 G100 10 GB20 50 GB100 10 G30 34 G150 7 GB30 34 GB150 7 G40 25 G300 3 5 GB40 25 GB300 3 5 G60 17 G6...

Страница 5: ...8 5 10kW GSP10 1000 GSP100 100 GBSP10 1000 GBSP100 100 GSP20 500 GSP150 68 GBSP20 500 GBSP150 68 GSP30 340 GSP200 50 GBSP30 340 GBSP200 50 GSP40 250 GSP300 34 GBSP40 250 GBSP300 34 GSP50 200 GSP400 26...

Страница 6: ...unication Settings 21 1 9 8 Reset Blank Front Panel Power Supply to Default Settings 22 CHAPTER 2 LOCAL OPERATION 23 2 1 Introduction 23 2 2 Load Connections Options 23 2 2 1 Connecting Single Loads L...

Страница 7: ...Setting Memory 41 2 13 4 Save 1 4 41 2 13 5 Recall 1 4 41 CHAPTER 3 REMOTE ANALOG PROGRAMMING 44 3 1 Introduction 44 3 2 Local Remote Analog Control 44 3 3 Local Remote Analog Indication 44 3 4 Remot...

Страница 8: ...The Home Page 68 4 7 6 4 Login Rules 69 4 7 6 5 DC Power Page 70 4 7 6 6 LAN Page 73 4 7 6 7 HELP Page 77 4 7 7 Programming Using VISA Drivers 77 4 7 7 1 VISA Description 77 4 7 7 2 VXI 11 Compatibil...

Страница 9: ...rchy 97 4 12 6 Header 97 4 12 7 Data Formats 97 4 12 8 Commands Notes 98 4 13 SCPI Common Commands 98 4 14 SCPI Subsystem Commands 103 4 14 1 Display Subsystem 103 4 14 2 Initiate Subsystem 104 4 14 3...

Страница 10: ...lel System Acknowledge 146 5 9 4 1 Acknowledge via the front panel 146 5 9 4 2 Acknowledge via communication 146 5 9 4 3 Acknowledge via communication Blank Panel master power supply 146 5 9 5 Advance...

Страница 11: ...eral 167 7 2 IEEE 488 2 Interface 167 7 3 Point to Point Connection 167 7 4 Multi Drop Connection 168 7 4 1 Selecting a Single Power Supply in a Multi Drop Chain 168 7 5 Communication Cables 169 7 6 I...

Страница 12: ...TER 1 FRONT REAR PANEL CONTROL CONNECTORS 1 1 Introduction The power supply series has a full set of controls indicators and connectors that allow the user to setup and operate the unit Before startin...

Страница 13: ...lected function 5 Operation Mode Indicator CV CC CP Operation mode indicator 6 Current Display 4 digit 16 segment Current display Normally displays the output current In Preview mode the display indic...

Страница 14: ...reen LED lights when System menu is active If System menu is active press the SYST button to exit to the main menu Lock Unlock Front Panel buttons by pressing the SYST button followed by current encod...

Страница 15: ...voltage current rated scale Green LED lights when the unit is in Fine mode 16 PREV Button Indicator Press the PREV button to display the Output Voltage and Current Limit settings The display shows the...

Страница 16: ...ation Mode Indicator CV CC CP operation mode indicator 3 Current Display 4 digit 16 segment Current display Normally displays the output current In preview mode the display indicates the program setti...

Страница 17: ...emote communication RS232 485 USB LAN OPTional 11 RS Indicator Recommended Standard indicator RS232 or RS485 communication type is selected 12 USB Indicator Universal Serial Bus indicator USB communic...

Страница 18: ...tus LED Green LED lights when power supply is ON Power Switch ON 4 REM LED REMOTE status LED Green LED lights if power supply is controlled by a remote communication RS232 485 USB LAN OPTional 5 OUT L...

Страница 19: ...SCHURTER IEC C16 1 5kW 3 4kW 1 Phase PC 5 3 G 7 62 PHOENIX CONTACT shown 2 7kW 5kW 3 Phase PC 5 4 G 7 62 PHOENIX CONTACT shown 10kW 15kW DFK PC 16 4 ST 10 16 PHOENIX CONTACT shown 2 Ground Stud Functi...

Страница 20: ...5 seconds default communication interface following reset is set to optional communication interface i e IEEE MODBUS TCP or EtherCAT Power Switch must be set to ON position 6 Paralleling Connectors Ma...

Страница 21: ...Manual for any connect disconnect of any connector on the rear panel 1 6 J1 Connector Terminal and Function Control and monitoring signals are SELV Connector Technical Information Connector type 6180...

Страница 22: ...nals 12 COM COMMON Return for all signals 13 COM COMMON Return for all signals 14 COM COMMON Return for all signals 15 NOT USED 16 NOT USED 17 COM COMMON Return for all signals 18 COM COMMON Return fo...

Страница 23: ...ENA ENABLE IP IP ENA PL ENABLE POLARITY MAC MAC NORM NORMAL BAUD BAUD RATE REV REVERSE LANG LANGUAGE PSO DL POWER SUPPLY OK DELAY SCPI SCPI SAVE SAVE GEN GENESYS RECAL RECALL REV REVISION FRST FACTOR...

Страница 24: ...DOWN RES RESISTANCE RESET RESET WD WATCHDOG WD TIM WATCHDOG TIMER Table 1 6 Front Panel Display Messages 1 8 Menu Navigation 1 8 1 Introduction power supply series contains the following five independ...

Страница 25: ...AGE ENCODER 6 COMMUNICATION MENU E N TE R E N TE R Level down YES Note Press Current encoder to select required parameter Current display blinks Parameter accepted For LAN or OPT Interfaces GEN langua...

Страница 26: ...L OFF ON UVP OFF FOLD OFF Voltage encoder 1 step rotate Current encoder 1 step rotate Function Voltage Display Parameter Current Display E N TE R Level down Note Press Current encoder to select requir...

Страница 27: ...UP 0 0001 999 99 VSL UP CSL DN 0 0001 999 99 VSL DN SLEW OFF SLEW VOLT SLEW CURR C POW OFF R INT OFF R INT ON Voltage encoder 1 step rotate Current encoder 1 step rotate Function Voltage Display Para...

Страница 28: ...0 115 2 60 BRT 1 4 DM BRT 0 3 LOCK ULOCK Present mode LOCAL Present mode REMOTE LOCKED ENA PL REV NORM Note Menu displays alternative parameter only no selection OFF E N TE R E N TE R E N TE R DM BRT...

Страница 29: ...meter accepted Voltage encoder multiple steps rotate Current encoder multiple steps rotate Note Menu displays alternative parameter only no selection Figure 1 10 PROGRAM Menu Diagram 1 8 2 Exiting a M...

Страница 30: ...lation procedure Following driver installation connect power supply to PC using USB cable Power supply COM port is detected in Device Manager under Ports COM LPT category as G serial port COMX where X...

Страница 31: ...face NOTES 1 Following SYST INT RS232 command power supply communication interface immediately switches to RS232 Execution time is 25mS 2 Power supply has to be re addressed using INST NSEL 6 command...

Страница 32: ...pply Communication Settings power supply default communication parameters are as follows Interface USB Language SCPI Baud rate 115200 Address 6 Genesys power supply default communication parameters ar...

Страница 33: ...LANG GEN switch communication language to GEN language NOTE There are differences between power supply series GEN language commands to Genesys power supply series GEN language commands Verify commands...

Страница 34: ...refer to section 4 2 2 The power supply has three operating modes The two basic operating modes are Constant Voltage mode and Constant Current mode The mode in which the power supply operates at any g...

Страница 35: ...vermeiden CAUTION When using shielded sense wires ground the shield in one place only The location can be the power supply chassis or one of the output terminals Figure 2 2 shows recommended remote s...

Страница 36: ...tput distribution terminals are used the power supply output terminals should be connected to the distribution terminals by a pair of twisted and or shielded wires Each load should be separately conne...

Страница 37: ...by encoder press Exit the PREV menu by pressing the Back button or the PREV button c When the output is disabled press the PREV button and then rotate the Voltage encoder to a required value Press th...

Страница 38: ...to acknowledge the setting Exit the PREV menu by pressing the Back button or the PREV button 2 Set current programming resolution to Coarse or Fine adjustment a Press the FINE button to select between...

Страница 39: ...nable the output momentarily press the OUTPUT button 2 8 Viewing Software Revision Via the Front Panel Menu it is possible to view the installed software revision 1 Press the COMM button COMM GREEN LE...

Страница 40: ...utton PROT GREEN LED illuminates OVP message appears on the display 2 Rotate the Current encoder to adjust the OVP level 3 Press the Current encoder to select the required OVP level The display blinks...

Страница 41: ...f UVL setting value is limited at the maximum level to approximately 5 below the Output Voltage setting Attempting to adjust the value above this limit will result in no response to the adjustment att...

Страница 42: ...shut down the power supply output if power supply operation mode crosses over from CC to CV or from CV to CC according to a selected operation mode There are three states of Foldback protection OFF d...

Страница 43: ...s off The display shows FOLD FAULT and the alarm LED blinks with 1 2 Hz frequency 2 9 5 Protection Delay Foldback protection delay is the time between feedback transition CV to CC transition or vice v...

Страница 44: ...600VDC rated output No point may be at a greater potential of 600VDC from ground WARNUNG Wenn Stromversorgungen in Reihe geschaltet werden und die Last oder eine der Ausgangsklemmen geerdet ist darf...

Страница 45: ...damage It is recommended that diodes be connected in parallel with each unit output to prevent reverse voltage during start up sequence or in case one of the units shuts down Each diode should be rate...

Страница 46: ...ter unit The signal is referenced to the isolated interface COM_SELV J1 11 12 13 14 The function is active after initial high to low transition PS_OK_OUT signal is Power Supply OK Open Collector type...

Страница 47: ...sy in J1 1 Daisy in J1 1 POWER SUPPLY 2 COM SELV Figure 2 7 Daisy Chain Connection 2 11 1 Daisy In Function DAISY_IN signal is used only for Daisy Chain application connected to DAISY_OUT signal of a...

Страница 48: ...SYST GREEN LED illuminates 2 Rotate the Voltage encoder until ILC OFF appears 3 Rotate Current encoder to select ILC ON or ILC OFF Press Current encoder to accept The display blinks once to acknowled...

Страница 49: ...ton 2 12 2 1 Activated ENA Alarm When the ENA is activated the power supply output shuts down The display shows ENA FAULT RED alarm LED blinks 1 2Hz frequency 2 12 3 Enable_In Polarity This function p...

Страница 50: ...3 Rotate Current encoder to select PIN 1 ON or PIN 1 OFF PIN 2 ON or PIN 2 OFF Press Current encoder to accept The display blinks once to acknowledge the setting 4 Exit the SYST menu by pressing the...

Страница 51: ...10mA sink current CAUTION Do not connect CV CC signal to a voltage source higher than 30VDC Always connect CV CC signal to the voltage source with a series resistor to limit the sink current to less t...

Страница 52: ...et via the Front Panel or communication User can save up to 4 sets of parameters Refer to Table 2 6 for available Store parameter sets To save the Front Panel setting perform the following steps 1 Pre...

Страница 53: ...NT RATED VOLTAGE MULTIPLYED BY RATED CURRENT Slew Mode OFF OFF Voltage Slew Up Down V mS 999 99 999 99 Current Slew Up Down A mS 999 99 999 99 Internal Resistance Function OFF OFF Internal Resistance...

Страница 54: ...Non Volatile Memory Parameters NOTES 1 REM mode is set after AC recycle if LLO mode was set prior AC recycle 2 Refer to Table 1 4 Reset Button if Reset operation is performed by Reset Button located...

Страница 55: ...oltage or by programming resistors via J1 contacts Refer to Analog Programming Setting Table 3 1 Figure 3 1 for remote voltage programming connection or Figure 3 2 for remote resistor programming conn...

Страница 56: ...encoder until RANGE appears on the voltage display 6 Rotate the Current encoder to set 5 or 10 External 5V or 10V Press the Current encoder for selection display blinks once to acknowledge the settin...

Страница 57: ...For Resistor Analog Programming wiring refer to Figure 3 2 Set the Remote Programming settings as follows 1 Short the pin J1 6 to J1 11 2 Press the CONF button CONF GREEN LED illuminates 3 Rotate Vol...

Страница 58: ...put voltage and current settings defined by the Encoders or Communication 6 To limit the maximum remote analog programming value of the current programming channel refer to Section 5 7 OCL Analog Prog...

Страница 59: ...MAC1 MAC6 MAC1 MAC6 LAN Reset LAN RESET RESET Table 4 1 Front Panel Setup 4 2 1 Default Setting Refer to Table 2 6 Non Volatile Memory Parameters and Table 2 7 LAN Non Volatile Memory Parameters for...

Страница 60: ...MM LED illuminates INTFC message appears on the Voltage display 2 Rotate the Voltage encoder 2 steps clockwise IP message appears on the Voltage display 3 Rotate the Current encoder to select a requir...

Страница 61: ...lection RS232 RS485 USB power supply Series implement the SCPI standard interface to control the unit by remote communication In addition GEN legacy commands subset is provided for ease of use and bac...

Страница 62: ...ters cannot be changed via the front panel If an attempt is made to change parameters via the front panel Current display shows REM message for 2 seconds It is possible to return to Local mode by pres...

Страница 63: ...ns Refer to Figure 4 1 for the connector description Figure 4 1 J4 Rear Panel RS232 485 IN Connector Pinout RS232 485 IN Pinout 1 RS232 RX 2 RS232 TX 3 RS485 RXD 4 RS485 TXD 5 RS485 TXD 6 RS485 RXD 7...

Страница 64: ...isy chain configuration Refer to Figure 4 2 for the connector description Figure 4 2 J3 Rear Panel RS485 OUT Connector Pinout RS485 OUT Pinout 1 NC Not Connected 2 NC Not Connected 3 RS485 TXD 4 RS485...

Страница 65: ...s Db 9 Connector 8 Pin Connector Remarks Pin No Name Pin No Name Housing Shield Housing Shield 2 RX 2 TX TWISTED PAIR 3 TX 1 RX 5 SG 8 SG Figure 4 3 RS232 Cable with DB9 Connector P N GEN 232 9 DB 9 C...

Страница 66: ...Lambda Web Site decompress zipped file if compressed navigate to x86 folder for 32 bit operating system or x64 folder for 64 bit operating system launch dpinst exe file Follow on screen instructions t...

Страница 67: ...s and reads the power supply output and status Security settings to block multiple controllers and risky protocols Optional password protection to prevent unauthorized operation LAN protocols VISA Tel...

Страница 68: ...ne Feed characters as a terminator 4 7 2 Specifications 4 7 2 1 LAN Specifications ELECTRICAL Ethernet Meets IEEE 802 3u specifications Auto MDIX Accepts patch or cross over cable connection Auto Nego...

Страница 69: ...er Verify LAN connection to instrument HTTP Web page server with Java scripts COMMANDS SCPI SCPI 1999 compliant command set measure and status IEEE 488 2 Condition and event register tree MULTIPLE CON...

Страница 70: ...nel LEDs SECURITY Web Page Password Can set password to prevent unauthorized or accidental changes to LAN settings or power supply output settings Single Client Only Set to prevent multiple programs f...

Страница 71: ...uter using the power supply s built in web pages or using SCPI programming commands To select the LAN mode 1 Press the COMM button on the front panel 2 Rotate the Current encoder to select LAN LAN mes...

Страница 72: ...sed A standard straight patch CAT 5 or better network cable A crossover cable The cable type is auto detected by the power supply NOTE The serial link RS 485 cable 0 5m long provided with the power su...

Страница 73: ...power supply 2 Connect the LAN cable before or after the power supply is switched ON 3 For a network with a DHCP server wait about 10 seconds The rear panel green LAN Status LED turns ON 4 For a peer...

Страница 74: ...icate LAN IP SRD command refer to section 4 14 10 Range Any address 169 254 xxx xxx Any address Lifetime Address may change as the DHCP server assigns addresses dynamically to many instruments Fixed p...

Страница 75: ...h a different way to use the host name This is shown in the table below DHCP Auto IP Static IP Hostname Default Hostname xxxxvvvv aaaa Refer to the above example xxxxvvvv aaaa Refer to the above examp...

Страница 76: ...isplay 3 Press Current encoder to enter IP configuration Rotate Voltage encoder to view the IP address The voltage display shows IP1 IP4 by rotating Voltage encoder The current display shows the corre...

Страница 77: ...STATIC IP addressing DHCP and Auto IP addressing is disabled 2 Refer to SYSTem COMMunicate LAN IP SRD command to modify the IP address by communication 4 7 5 3 LAN Reset To reset the power supply LAN...

Страница 78: ...ble 1 Read the IP address from the front panel Refer to section 4 7 5 1 2 Open a web page browser program such as Internet Explorer Enter the power supply IP address as shown below 3 Press the Enter k...

Страница 79: ...to section 4 7 4 4 VISA Name Using Hostname An alternate VISA resource descriptor may be the hostname Refer to section 4 7 4 5 RS 485 Address This address applies to a multi drop chain of power suppl...

Страница 80: ...ult the password field is empty Figure 4 10 LOGIN Button 7 Click Login 8 The password may be set or removed on the LAN Users web page 9 LAN Reset clears the password 4 7 6 4 Login Rules Up to 2 users...

Страница 81: ...ng the RS 485 address Select RS 485 Multi drop Address This address is described in section 4 7 1 1 This address applies to a multi drop chain of RS 485 supplies only If there is no multi drop chain o...

Страница 82: ...ify 0 Faults For the selected supply these boxes show any fault which may be active in the supply B DC Power Protection Page On the DC Power tab the Protection button is available at the top of the pa...

Страница 83: ...lity Page On the DC Power tab the Utility button is available at the top of the panel This page is used for sending any SCPI command and viewing the response It is a learning tool for SCPI language op...

Страница 84: ...Address Displays the IP address assigned to the power supply through either DHCP Auto IP or Static IP sources Subnet Mask Displays the subnet mask assigned to the power supply through either DHCP Auto...

Страница 85: ...he LAN settings The available fields depend on the selection of DHCP Enabled AUTO IP or Static IP Changes to these setting will take place only after clicking the Apply button The settings can be chan...

Страница 86: ...requirements of the network server These settings do not change if the power supply is transferred to a different LAN connection DNS server field is disabled gray NOTE In this mode there is no hostnam...

Страница 87: ...which instruments are connected to the LAN It may be disabled for security reasons Auto MDIX This service is always enabled in the power supply The power supply LAN connection will always detect a pa...

Страница 88: ...es of communication busses are supported VISA drivers are available from several instrument vendors Any programming language that supports Windows COM or DLL libraries can call the VISA functions 4 7...

Страница 89: ...ocket Description The VISA drivers for the power supply are commonly used in the Test and Measurement industry For customers who cannot use VISA because of installation licensing issues or because the...

Страница 90: ...wo VISA ports may be open at a time Mixed VISA and Sockets N A Up to two VISA ports and 3 TCP UDP Sockets may be open at a time 4 7 8 4 Input Buffer Requirements With a controller using TCP or UDP soc...

Страница 91: ...acknowledgement that they have been received Open UDP socket port 8005 to send SCPI commands Responses to queries are sent back automatically with a line feed terminator and carriage return appended...

Страница 92: ...2 Other units connection The other units on the bus are connected via RS485 interface Refer to Figure 4 21 for typical connections 3 Using the Linking cable supplied with each unit Refer to Figure 4 2...

Страница 93: ...to any one of the supplies in an RS 485 chain by first sending the INSTrument N SELect NR1 address command All commands and queries will then apply only to the selected supply until a new INSTrument...

Страница 94: ...data format is 8 bit one start bit and one stop bit No parity bit 4 9 2 End of Message The End of Message is the Carriage Return character ASCII 13 0x0D Power supply ignores the Line Feed ASCII 10 0x...

Страница 95: ...n implied decimal point assumed at the right of the least significant digit Example 255 NR2 Digits with an explicit decimal point Example 0253 NRh Digits in the hexadecimal format Bool Boolean data Ex...

Страница 96: ...m dd Returns SRD Example 2017 12 17 4 10 5 Initialization Commands CLS Function Clear Status Sets FEVE and SEVE registers to zeroes Returns DSC OK ADR NR1 Function Send ADR and address 0 to 31 to open...

Страница 97: ...Execution time for this command is 250mS 3 FRST command sets USB communication interface 4 10 6 Output Commands PV NR2 Function Set digital programming output voltage value in Volts Parameters Range i...

Страница 98: ...UT ON will respond with E07 if the output cannot be turned on because of a latching fault OTP AC ENA ILC POFF SO shutdown Parameters 0 1 OFF ON Query OUT Returns Bool 0 1 FLD DSC Function Sets the Fol...

Страница 99: ...e 2 1 UVL NR2 Function Sets the Under Voltage Limit Value n may be equal or less than 95 of PV setting Returns E06 if higher Parameters Range is limited by the UVL programming range and voltage progra...

Страница 100: ...ster Slave setting Single Master number of slaves units connected or Slave Return SRD SINGLE MASTER SLAVE NOTE Power supply does not execute commands in the Slave mode 4 10 7 Global Output Commands Ge...

Страница 101: ...reply GPV NR2 Function Refer to PV command description Returns No reply Model Rated Output Voltage V Minimum V Maximum V 10 0 10 20 0 20 30 0 30 40 0 40 50 0 50 60 0 60 80 0 80 100 0 100 150 0 150 20...

Страница 102: ...5 6 Table 4 5 Current Programming Range 1kW 3 4kW Model 5kW Maximum A Model 10kW Maximum A Model 15kW Maximum A 10 500 500 10 1000 1000 10 1500 1500 20 250 250 20 500 500 20 750 750 30 170 170 30 340...

Страница 103: ...communication protocol to SCPI language Parameter SCPI RIE Bool Function Enable Disable ILC function If disabled power supply ignores the ILC pin state Parameters 0 1 OFF ON Query RIE Returns Bool 0...

Страница 104: ...the complete power supply status Returns ASCII characters representing the following data separated by commas MV actual measured voltage PV programmed set voltage MC actual measured current PC program...

Страница 105: ...ister that holds the real time status of the power supply Returns NRh Example 09FA SENA NRh Function Sets the value of the Operation Group Enable register This register is a mask to enable specific bi...

Страница 106: ...4 Communication PC write ADR 06 Power supply response OK PC write OUT 1 Power supply should turn output ON Power supply response OK PC write PV n for n values refer to Table 4 4 Power supply response...

Страница 107: ...ne There is no CR between the command string and the sign Example STT 3A STAT 7B STT 3A checksum is calculated as follows S 0x53 T 0x54 T 0x54 0x3F 0x53 0x54 0x54 0x3F 0x013A checksum is the least sig...

Страница 108: ...sensitive It will recognize any case mixture such as TRIGGER Trigger and TRIGger Short form headers result in a faster program execution 4 12 7 Data Formats Data Formats Description NR1 Digits with an...

Страница 109: ...ENT or MEASURE CURRENT DC Can be used Expressions enclosed in angle brackets are programming values entered without the For example ESE 255 Vertical bars separate alternative parameters For example BU...

Страница 110: ...the Standard Event Status Enable register contents through power cycles The CLS clear status command does not clear the Standard Event Status Enable register Any or all conditions can be reported to...

Страница 111: ...mand error or an execution error QYE Query Error Set to 1 if the output queue is empty or if the data is missing even after a query has been sent OPC Operation Complete CLS Clear Status command clears...

Страница 112: ...Power ON condition occurs Parameters 0 1 OFF ON Query PSC Returns Bool 0 1 RCL NR1 Function This command restores power supply to a state stored in the memory locations 1 through 4 previously stored b...

Страница 113: ...on Bit Position 7 6 5 4 3 2 1 0 Bit Name OPR ESB MAV QUE SYS Bit Weight 128 32 16 8 4 OPR Operation Condition Status Register Summary ESB Standard Event Status Register Summary MAV Output Queue Messag...

Страница 114: ...Those can be a single command or a group of commands Groups are comprised of commands that extend one or more levels below the root Commands followed by a question mark take only the query form 4 14...

Страница 115: ...ming parameter Send ABORt commands to allow parameters change 4 14 3 Instrument Subsystem INSTrument COUPle DSC Function Defines coupling between all Power Supplies connected in a multi drop connectio...

Страница 116: ...GLOBal CURRent AMPLitude NRf Function Refers to SOURce CURRent LEVel IMMediate AMPLitude NRf command description Units A GLOBal OUTPut STATe Bool Function Refers to OUTPut STATe Bool command descripti...

Страница 117: ...signal polarity NORM same as interlock Output ON if ENA pin signal J1 10 is low 0 REV Output ON if ENA pin signal J1 10 is high 1 Parameter 0 1 REV NORM Query OUTPut ENA POLarity Return DSC REV NORM...

Страница 118: ...power supply output if a transition between operation modes occurs Power supply turns output off after a specified delay if there is a transition into CV mode or into CC mode This feature is particul...

Страница 119: ...ted Parameter 0 1 2 OFF FSTR TRIG Query OUTPut TTLTrg MODE Return DSC OFF FSTR TRIG 4 14 7 Program Subsystem PROGram COUNter NR1 INFinity Function Sets the sequencers iterations counter Parameters 1 9...

Страница 120: ...nts are separated by commas The function accepts up to 100 parameters Parameters The range is limited by the power supply voltage programming range including UVP OVP settings Unit V Query PROGram LIST...

Страница 121: ...EP Return DSC AUTO ONCE PROGram STORe NR1 Function Stores a sequence in a specified memory cell 1 2 3 or 4 Stored parameters mode voltage current dwell time step and counter Parameters 1 2 3 4 NOTE A...

Страница 122: ...s The range is limited by the power supply voltage programming range including UVP OVP settings Unit V Query PROGram WAVE VOLTage Return NR2 NR2 5 digits format Example WAVE VOLT 2 0 2 5 3 0 WAVE VOLT...

Страница 123: ...turn DSC DIG VOL RES SOURce CURRent LEVel IMMediate AMPLitude NRf Function Sets a digital programming output current value Parameter The range is limited by the power supply model Unit A Query SOURce...

Страница 124: ...eters 0 1 2 NONE LIST WAVE Query SOURce CURRent MODE Return DSC NONE LIST WAVE SOURce VOLTage EXTernal MODE DSC Function Sets the voltage programming source Front panel Communication Digital external...

Страница 125: ...EW UP Return NR2 Example VOLT SLEW UP 1 Sets a digital voltage reference up programming slew to 1V mSec VOLT SLEW UP returns digital voltage up programming slope value VOLT SLEW UP MAX and VOLT SLEW U...

Страница 126: ...UVP mode If enabled the output is disabled when measured output voltage reaches the UVL level UVP bit is set in the Questionable Condition register Parameters 0 1 OFF ON Query SOURce VOLTage PROTecti...

Страница 127: ...d Slew Rate functions should be disabled to enable Constant Power SOURce POWer LEVel NRf Function Sets the constant power limit level Parameters The range is limited to a multiplication of maximum vol...

Страница 128: ...current programming channel is set EVR External Voltage Reference Set to 1 when analog voltage programming channel is set CFB Current Foldback Enabled Set to 1 when Constant Current operation mode Fol...

Страница 129: ...nd for the complete registers list Returns NR1 0 65535 Example 00136 STATus OPERation ENABle NR1 Function Sets the value of the Operational Condition Group Enable register This register is a mask for...

Страница 130: ...Slave Set to 1 when master power supply is waiting for slaves to become ready PERR Parallel Error Set to 1 when an error occurs in Advanced Parallel system GERR General Error Unrecoverable system fau...

Страница 131: ...be masked Parameters 0 65535 Query STATus QUEStionable ENABle Returns NR1 0 65535 4 14 10 SYSTem Subsystem SYSTem COMMunicate ADDRess NR1 Function Sets the power supply communication address Parameter...

Страница 132: ...etc according to the number of Power Supplies from the same type A number is added only if at least 2 Power Supplies of the same type are connected to the same network SYSTem COMMunicate LAN IDLED Bo...

Страница 133: ...hdog Parameter 0 1 OFF ON Query SYSTem COMMunicate WATCHdog STATe Return Bool 0 1 SYSTem COMMunicate WATCHdog TIMe NR1 Function Set communication watchdog timeout in seconds Parameter 1 3600 MIN MAX S...

Страница 134: ...d above SYSTem FIRMware VERSion Query Returns firmware versions string Interface Master Control Display and Option if available Return SRD xx xxx xx xx x x xx xxx NOTE Option firmware revision last xx...

Страница 135: ...S_OK signal delay following output ON Parameters 0 000 10 MIN MAX Resolution 0 001 Unit S Query SYSTem PSOK DELay SYSTem PSOK DELay MAX SYSTem PSOK DELay MIN Return NRf SYSTem RANGe DSC Function Sets...

Страница 136: ...vates LLO only by communication or AC recycle Parameters 0 1 2 or LOC REM LLO Query SYSTem REMote STATe Return DSC LOC REM LLO NOTES System state changes from LOC to REM only if a communication comman...

Страница 137: ...mand TRIG generates an immediate trigger signal TRIGger DELay NRf Function Sets a time delay between trigger events of a specified trigger source to the start of any corresponding trigger action over...

Страница 138: ...ttings Saves an instrument state Set service request enable register Return status byte Trigger Performs self test returns a result Wait for all operations to complete Aborts the triggered action Desc...

Страница 139: ...t trigger depends execution step Store program in memory Set WAVE output current points Set slopes time Set WAVE output voltage points Enable disable analog mode current limit External source of curre...

Страница 140: ...efault Set communication language Enable disable communication watchdog Set communication watchdog timeout Returns system calibration date Enable error message Read system error messages Factory reset...

Страница 141: ...gger refer to section 5 2 3 5 1 1 LIST Mode The Output values change in steps determined by parameters in the LIST Output change is activated by input trigger refer to section 5 2 2 Program LIST param...

Страница 142: ...Set the operation mode by SOURce CURRent MODE DSC and or SOURce VOLTage MODE DSC command s IDLE STATE INITIATE STATE DELAY STATE OUTPUT CHANGE STATE LINEAR STEP WAVE STEP COMPLETE PROG STEP AUTO PROG...

Страница 143: ...2 points in the WAVE mode Time range is 1mSec 36 hours in steps of one milli second Time setting is available via the communication command PROGram WAVE TIME NRf NRf 5 1 3 4 Load Load a sequence from...

Страница 144: ...execution Return sequencer system to idle state Abort function is available via the Front Panel menu refer to section 1 8 or via the communication command ABORt NOTE Consecutive ON OFF ON function vi...

Страница 145: ...Out Mode OFF Table 5 1 Sequencer Memory Parameters 5 2 Trigger System The Trigger system consisting of Trigger In and Trigger Out functions provides the ability to synchronize sequencer arbitrary wav...

Страница 146: ...to SOURce CURRent MODE DSC and SOURce VOLTage MODE DSC commands Programming mode NONE for Current and Voltage mode both set to NONE OFF Mode No trigger output signal TRIG Mode Trigger is generated if...

Страница 147: ...system INIT Power supply is waiting for trigger 7 Turn on the power supply output and send a trigger command TRG NOTE To stop execution of a running sequence use ABORt command refer to section 4 13 5...

Страница 148: ...the PROG button PROG Green LED illuminates TRIG ABORT appears Press Current encoder to abort the running sequence 5 3 4 LIST Mode Example VOLT MODE LIST Select LIST Mode Sequence LIST VOLT 2 4 2 8 5 4...

Страница 149: ...5 Set time values 1 0 5 0 5 0 5 0 5 1 5 1 5 Seconds STEP AUTO Set AUTO step execution mode AUTO COUN 1 Set wave execution iterations 1 TRIG SOUR BUS Select BUS trigger source via communication interf...

Страница 150: ...the communication command SYSTem RIN STATe Bool Internal resistance setting range is 0 001 to 1 ohm in steps of 0 001 ohms Internal resistance setting is available via the Front Panel menu refer to se...

Страница 151: ...by multiplication of rated voltage by rated current in steps of 1 watt Constant power limit setting is available via the Front Panel menu refer to section 1 8 or the communication command SOURce POWe...

Страница 152: ...y This function is mainly used to prevent discharge of batteries connected to the power supply output terminals Setting preload to ON state default configuration keeps the preload active as long as po...

Страница 153: ...lue 10 Volts CURR 300 Set current value 300 Amperes SYST RANG 10 Set analog programming range 10 CURR EXT MODE VOL Set analog current programming source Set external voltage to 4V Current programming...

Страница 154: ...8 or the following communication commands SOURce CURRent SLEW DOWN NRf Set current reference down programming slew SOURce CURRent SLEW UP NRf Set current reference up programming slew SOURce VOLTage S...

Страница 155: ...performed while the Power Supplies are disconnected from the grid disconnected from the AC source 5 9 2 Advanced Parallel Dis Connection System Dis Assembly Advanced parallel disconnection is performe...

Страница 156: ...nals is reliable to prevent disconnection during operation Disconnection may cause damage to the power supply NOTE With local sensing it is important to minimize wire length to decrease wire resistanc...

Страница 157: ...s the Current encoder Display blinks indicating that the parallel system assembly was accepted NOTE If the master power supply is a Blank Panel Option power supply refer to 5 9 4 2 to acknowledge the...

Страница 158: ...ts Slave units allow only query communication commands 5 9 7 Advanced Parallel fault system Advanced parallel fault system combines the faults of the master and slave units The master unit shows on th...

Страница 159: ...el System Three roles are available Single No Slave or Master units are connected Master At least one Slave unit is connected Slave A unit or multiple units connected to a Master unit Prior assembling...

Страница 160: ...ddress Not Affected Preserves Setting Communication Baud Rate Baud Not Affected Preserves Setting Communication Language Not Affected Preserves Setting LAN parameters Not Affected Preserves Setting ex...

Страница 161: ...mS 999 99 Current Slew Up Down A mS SAME AS MASTER POWER SUPPLY Internal Resistance Function OFF Internal Resistance Set Point 0 001 Voltage Reference Programming Source DIG Current Reference Program...

Страница 162: ...ted Preserves Setting Factory Calibration Parameters Not Affected Preserves Setting Table 5 4 Slave Role Non Volatile Memory Parameters NOTE Slave unit parameters read by communication might not refle...

Страница 163: ...ation Watchdog Timeout When communication watchdog timeout period expires power supply output shuts down The display shows WD FAULT Red alarm LED blinks 1 2Hz frequency error message is written into e...

Страница 164: ...r example OUTPut STATe Bool default return value is 0 1 it could be switched to OFF ON Refer to SYSTem COMMunicate BOOL STATe DSC command in SCPI language SYSTem Subsystem Section 4 14 10 Refer to BOO...

Страница 165: ...s errors faults and SRQ register structures The registers can be read or set via communication commands Two individual sets of registers are managed one set for the SCPI language another set for the G...

Страница 166: ...lds a snapshot of the actual faults state of the power supply at a present time If a fault occurs the corresponding bit is set If a fault is removed the corresponding bit is cleared Some of the faults...

Страница 167: ...ation of the operational condition status register group Bit Number Decimal Value Bit Symbol Description 0 1 CV Constant Voltage operation mode 1 2 CC Constant Current operation mode 2 4 NFLT No Fault...

Страница 168: ...ter is cleared at power supply power on Bit Number Decimal Value Bit Symbol Description 0 1 OPC Operation Complete 1 2 2 4 QYE Query Error 3 8 DDE Device Dependent Error 4 16 EXE Execution Error 5 32...

Страница 169: ...ble messages until 0 No error message is received Whenever the queue holds a message it sets the SYS bit in the status byte register refer to Figure 6 1 Error Number Error Description Error Event 0 No...

Страница 170: ...elow PV Attempt to set OVP below voltage setting 306 UVL Above PV Attempt to set UVL above voltage setting 307 On During Fault Attempt to turn power supply output on when a fault exists 320 UVP Shutdo...

Страница 171: ...mming or Sequencer while Slew rate is ON 340 Advanced Parallel Wait Acknowledge Advanced Parallel Waiting for Acknowledge system has changed 341 Input Overflow Received over 1500 characters 342 Commun...

Страница 172: ...event Table 6 6 Service Request Enable Register SYS System error message available bit Refer to Section 6 2 6 to read the available error messages QUE Questionable condition group summary bit Questio...

Страница 173: ...ing Event register for each summary group to determine which events caused the summary bit to be set When an Event register is read it is cleared This action also clears the corresponding summary bit...

Страница 174: ...163 6 3 GEN Language 6 3 1 GEN Register Tree Figure 6 2 GEN Register Tree Diagram...

Страница 175: ...ter holds a snapshot of the actual faults state of the power supply at a present time If a fault occurs the corresponding bit is set If a fault is removed the corresponding bit is cleared Some of the...

Страница 176: ...Enabled 6 64 7 128 LOC Local Remote communication mode 8 256 UVP Under Voltage Protection enabled 9 512 ILC enable Interlock control Enabled 10 1024 ENA enable Enable control Enabled 11 2048 CFB CC op...

Страница 177: ...Returned if an attempt to set an operation mode Internal Resistance Constant Power Sequencer Slew Rate or Analog Programming if another operation mode is enabled or loading an empty memory cell Only...

Страница 178: ...t control wires and eight ground wires If the system runs from a personal computer there are numerous vendors of IEEE controller cards and software Because many instruments may be connected and indepe...

Страница 179: ...a Multi Drop Chain All of the SCPI commands may be sent to any one of the Power Supplies in RS485 chain by sending the INSTrument NSELect NR1 command All commands and queries will then apply only to t...

Страница 180: ...ce 7 7 Power Supply Configuration To communicate over the IEEE interface power supply has to be configured to IEEE If IEEE option is installed by default power supply is pre configured to the IEEE com...

Страница 181: ...asurement Automation Explorer Figure 7 3 National Instruments MAX Desktop Icon 2 On the side tree expand My System and select Devices and Interfaces GPIB0 3 In the GPIB Explorer toolbar click Scan for...

Страница 182: ...ument NI 488 2 Communicator appears Refer to Figure 7 6 Figure 7 6 ID String Query In the Send String text box IDN appears 6 Click Query Refer to Figure 7 6 The ID string indicates the following Compa...

Страница 183: ...f detergent and water The solution should be applied onto a soft cloth and not directly to the surface of the unit CAUTION Do not use aromatic hydrocarbons or chlorinated solvents for cleaning Use low...

Страница 184: ...applied Check input AC voltage Connect to appropriate voltage source Safety Installation Manual Output voltage does not adjust Front panel CC indicator is on Is the unit in constant current mode Check...

Страница 185: ...y Connect the sense wires according to User Manual instructions Safety Installation Manual Front Panel controls are non functional Is the power supply in Local Lockout mode Press SYST Button SYST GREE...

Страница 186: ...reinigt werden Die L sung auf ein weiches Tuch geben nicht direkt auf die Oberfl che des Ger tes aufbringen VORSICHT Zur Reinigung keine aromatischen Kohlenwasserstoffe oder chlorhaltige L sungsmittel...

Страница 187: ...AC Spannung der Quelle ab wenn ein Verbraucher ange schlossen ist Eingangsnetzspannung pr fen An entsprechende Versorgungsspannung anschlie en Safety Installation Manual Ausgangsspannung l sst sich ni...

Страница 188: ...Installation Manual Die Bedienelemente an der Frontplatte sind au er Funktion Wird das Netzger t im Local Lockout Modus betrieben SYST Taste dr cken SYST GREEN LED leuchtet Stromeinstellregler dr cke...

Страница 189: ...169 Front Panel Display Messages 12 Fuse Rating 174 GEN Command Set Description 84 Register Tree 163 GEN Protocol 83 HOME Page 68 Opening 67 IEEE 488 2 Interface 167 Controller Configuration 169 IEEE...

Страница 190: ...Serial RS232 RS485 48 Series Connection 34 Output Voltage 34 Series Operation Remote Programming 35 Service Request Group Structure 161 Setting Baud Rate 50 Default 40 Signal See CV CC Power Supply 39...

Отзывы: