TDK-Lambda G10-500 Скачать руководство пользователя страница 1

 

 

 

 

 

 Series

 

 

Programmable DC Power Supplies 

5kW in 1U 0-600V/ 0-500A 

Built in LAN, USB, RS-232 & RS-485  Interface 

Optional Interface: IEEE488.2 (GPIB) 

 

USER MANUAL 

 

This manual covers models 

G10-500 

G80-65 

GB10-500 

GB80-65 

G20-250 

G100-50 

GB20-250 

GB100-50 

G30-170 

G150-34 

GB30-170 

GB150-34 

G40-125 

G300-17 

GB40-125 

GB300-17 

G60-85 

G600-8.5 

GB60-85 

GB600-8.5 

 

 

 

 

 

 

 

 

IA761-04-02 

 

 

 

 

Содержание G10-500

Страница 1: ...B RS 232 RS 485 Interface Optional Interface IEEE488 2 GPIB USER MANUAL This manual covers models G10 500 G80 65 GB10 500 GB80 65 G20 250 G100 50 GB20 250 GB100 50 G30 170 G150 34 GB30 170 GB150 34 G4...

Страница 2: ...OPERATION 26 3 1 Introduction 26 3 2 Load Connections Options 26 3 2 1 Connecting Single Loads Local Sensing Default 26 3 2 2 Connecting Single Loads Remote Sensing 26 3 2 3 Connecting Multiple Loads...

Страница 3: ...ocal Remote Analog Control 46 4 3 Local Remote Analog Indication 46 4 4 Remote Voltage Programming of Output Voltage and Current 47 4 5 Remote Resistor Programming of Output Voltage and Output Current...

Страница 4: ...79 5 7 7 Programming Using VISA Drivers 80 5 7 7 1 VISA Description 80 5 7 7 2 VXI 11 Compatibility 80 5 7 7 3 Opening the VISA Connection 80 5 7 7 4 Communicating Using VISA 80 5 7 8 Programming Usi...

Страница 5: ...01 5 13 SCPI Common Commands 101 5 14 SCPI Subsystem Commands 106 5 14 1 Display Subsystem 106 5 14 2 Initiate Subsystem 107 5 14 3 Instrument Subsystem 107 5 14 4 Global Subsystem 107 5 14 5 Measure...

Страница 6: ...6 9 4 Advanced Parallel operation 148 6 9 5 Slave units operation in Advanced Parallel connection 149 6 9 6 Advanced Parallel fault system 149 6 9 7 Advanced Parallel system identification idn 149 CH...

Страница 7: ...eral 168 9 2 Specifications 168 9 3 Maintenance 168 9 3 1 Cleaning Instructions 168 9 3 2 Assembly Instructions for Standard Power Supplies 169 9 3 3 Assembly Instructions for Blank Power Supplies Wit...

Страница 8: ...This page intentionally left blank...

Страница 9: ...1 CHAPTER 1 SPECIFICATIONS 1 1 5000W Series Specifications...

Страница 10: ...2...

Страница 11: ...3...

Страница 12: ...4...

Страница 13: ...5 1 2 Standard Unit Low Voltage Outline...

Страница 14: ...6 1 3 Standard Unit High Voltage Outline...

Страница 15: ...7 1 4 Blank Panel Unit Low Voltage Outline...

Страница 16: ...8 1 5 Blank Panel Unit High Voltage Outline...

Страница 17: ...ional accessories 1 6 1 Printed User manual Printed User manual order P N G M 1 6 2 Serial Port Cables For ordering serial port cables refer to CHAPTER 5 1 6 3 Paralleling Cable Paralleling cable orde...

Страница 18: ...Refer to Figure 2 1 and Table 2 1 for description of the Front Panel controls Figure 2 1 Front Panel Controls No Control Indicator Description 1 Power Switch POWER ON OFF control 2 Power Supply Model...

Страница 19: ...Indicator Activates the Program Sequencer menu The Program menu provides Sequencer function control Trigger function control and loads a sequence stored inside the power supply memory Green LED lights...

Страница 20: ...otection menu is active If the Protection menu is active press the PROT button to exit to the main menu 14 COMM Button Indicator Activates the Communication menu The Communication menu provides commun...

Страница 21: ...rns back to show actual output Voltage and Current Green LED lights when the PREV mode is active 17 OUT Button Indicator Output ON OFF control Press OUTPUT to set the output ON or OFF Green LED lights...

Страница 22: ...Panel indicator LFP is on if the Front Panel is locked 5 AUTO Indicator AUTO Start indicator AUTO is on if the Auto Start mode is active 6 SAFE Indicator SAFE Start indicator SAFE is on if the Safe S...

Страница 23: ...N Indicator Local Area Network indicator LAN communication type is selected 14 OPT Indicator Optional communication type indicator Optional communication type is selected 15 TRIG Indicator Trigger Ind...

Страница 24: ...or Indicators LAN interface connector RJ 45 type LAN status indicators Green LED on RJ45 connector Link Activity Amber LED on RJ45 connector Speed Lit 100Mbps otherwise 10Mbps Green status indicator c...

Страница 25: ...n 1 Daisy In SO Input for Series Operation Input for Shut Off control of the power supply output 2 Daisy Out PS_OK 2 Output for Series Operation Output 2 for indication of power supply status High lev...

Страница 26: ...MON Return for all signals 18 COM COMMON Return for all signals 19 ILC Enable Disable the power supply output by dry contact short open or voltage source 20 Programmed Signal 2 General Purpose Open Dr...

Страница 27: ...ENA ENABLE IP IP ENA PL ENABLE POLARITY MAC MAC NORM NORMAL BAUD BAUD RATE REV REVERSE LANG LANGUAGE PSO DL POWER SUPPLY OK DELAY SCPI SCPI SAVE SAVE GEN GENESYS RECAL RECALL REV REVISION FRST FACTOR...

Страница 28: ...ON VSL DN VOLTAGE SLEW DOWN RES RESISTANCE Table 2 5 Front Panel Display Messages 2 7 Menu Navigation 2 7 1 Introduction Power Supply series contains the following five independent menus COMMUNICATION...

Страница 29: ...meter Current Display LAN Inter fac e Ac tiv e NO COMM E N TE R E N TE R CURRENT ENCODER VOLTAGE ENCODER 6 COMMUNICATION MENU E N TE R E N TE R Level down YES Note Press Current encoder to select requ...

Страница 30: ...L OFF ON UVP OFF FOLD OFF Voltage encoder 1 step rotate Current encoder 1 step rotate Function Voltage Display Parameter Current Display E N TE R Level down Note Press Current encoder to select requir...

Страница 31: ...UP 0 0001 999 99 VSL UP CSL DN 0 0001 999 99 VSL DN SLEW OFF SLEW VOLT SLEW CURR C POW OFF R INT OFF R INT ON Voltage encoder 1 step rotate Current encoder 1 step rotate Function Voltage Display Para...

Страница 32: ...0 115 2 60 BRT 1 4 DM BRT 0 3 LOCK ULOCK Present mode LOCAL Present mode REMOTE LOCKED ENA PL REV NORM Note Menu displays alternative parameter only no selection OFF E N TE R E N TE R E N TE R DM BRT...

Страница 33: ...meter accepted Voltage encoder multiple steps rotate Current encoder multiple steps rotate Note Menu displays alternative parameter only no selection Figure 2 9 PROGRAM Menu Diagram 2 7 2 Exiting a Me...

Страница 34: ...which the power supply operates at any given time depends on the output voltage setting output current limit setting and the load resistance The third operation mode is the Constant Power mode In this...

Страница 35: ...the output terminals Figure 3 2 shows recommended remote sensing connection for single loads Remote sensing is used when in Constant Voltage mode the load regulation is important at the load terminal...

Страница 36: ...motely located output distribution terminals are used the power supply output terminals should be connected to the distribution terminals by a pair of twisted and or shielded wires Each load should be...

Страница 37: ...e the Voltage encoder to a required value Press the Voltage encoder to select the required value The display will blink once to acknowledge the setting Exit the PREV menu by pressing the Back button o...

Страница 38: ...stment a Press the FINE button to select between the lower about 1 of rated voltage and higher setting least significant digit seen on the voltage display resolution b The FINE LED illuminates when th...

Страница 39: ...nable the output momentarily press the OUTPUT button 3 2 10 Viewing Software Revision Via the Front Panel Menu it is possible to view the installed software revision 1 Press the COMM button COMM GREEN...

Страница 40: ...ars on the display 2 Rotate the Current encoder to adjust the OVP level 3 Press the Current encoder to select the required OVP level The display blinks once to acknowledge the setting 4 Exit the PROT...

Страница 41: ...minimum value setting is zero UVP must be turned ON to be operational 1 Press the PROT button PROT GREEN LED illuminates 2 Rotate the Voltage encoder until UVL appears on the Voltage display 3 Rotate...

Страница 42: ...on of Foldback state The display blinks once to acknowledge the setting If the CC state is selected the FOLD I indicator appears on the display If the CV state is selected the FOLD V indicator appears...

Страница 43: ...frequency Resetting the OTP circuit can be automatic non latched or manual latched depending on the power supply mode Safe start mode or Auto restart mode Safe start mode The power supply stays off a...

Страница 44: ...0VDC von 200VDC gegen ber Erde aufweisen Bei den Ausgangsmodellen 150 600VDC kann das maximale Spannungspotential bis zu 600VDC betragen 3 4 1 Series Connection for Increased Output Voltage Two units...

Страница 45: ...circuits of this power supply are isolated from the output potentials Therefore the circuits used to control each series connected unit don t have to be separated and floated from each other Using th...

Страница 46: ...hod may be applied to systems with more units COM SELV J1 11 J1 2 COM SELV Daisy out J1 11 J1 2 Daisy out Daisy in J1 1 POWER SUPPLY 3 J1 11 J1 2 Daisy out POWER SUPPLY 1 Daisy in J1 1 Daisy in J1 1 P...

Страница 47: ...ss the SYST button SYST GREEN LED illuminates 2 Rotate the Voltage encoder until ILC OFF appears 3 Rotate Current encoder to select ILC ON or ILC OFF Press Current encoder to accept The display blinks...

Страница 48: ...Press Current encoder to accept The display blinks once to acknowledge the setting 4 Exit SYST menu by pressing the Back button or the SYST button 3 6 3 Enable_In Polarity This function provides the...

Страница 49: ..._2 J1 20 are open drain maximum input voltage 25V and maximum sink current 100mA Programmed signals can be controlled via the Front Panel or via communications Selection of OFF via front panel or comm...

Страница 50: ...uired delay time Press on the current encoder to select the desired delay value 3 The display will blink once to acknowledge the setting 4 Exit SYST menu by pressing the Back button or the SYST button...

Страница 51: ...ets basic parameters to their default state as defined in Table 3 5 It is set via communication only The following states are cleared OVP Foldback and UVP 3 7 3 Last Setting Memory Last setting stores...

Страница 52: ...RATED 0 Over Voltage Protection OVP V 1 1 RATED or 1 2 RATED Model Dependent 1 1 RATED or 1 2 RATED Model Dependent Under Voltage Level UVL V 0 0 Under Voltage Protection UVP Mode OFF OFF Under Volta...

Страница 53: ...0 Depends on PSC SCPI only Event registers 00000 00000 Power On Status Clear 1 Display Brightness 3 Display Dim Delay S 30 Display Dim Brightness 2 Factory Calibration Parameters Factory Calibration T...

Страница 54: ...oltage or by programming resistors via J1 contacts Refer to Analog Programming Setting Table 4 1 Figure 4 1 for remote voltage programming connection or Figure 4 2 for remote resistor programming conn...

Страница 55: ...selection display blinks once to acknowledge the setting 5 Rotate the Voltage encoder until RANGE appears on the voltage display 6 Rotate the Current encoder to set 5 or 10 External 5V or 10V Press t...

Страница 56: ...nge For Resistor Analog Programming wiring refer to Figure 4 2 Set the Remote Programming settings as follows 1 Short the pin J1 6 to J1 11 2 Press the CONF button CONF GREEN LED illuminates 3 Rotate...

Страница 57: ...r supply 5 Front panel PREV button Use the PREV button to display the output voltage and current settings defined by the Encoders or Communication 4 6 Monitoring of Output Voltage V_MON and Current I_...

Страница 58: ...ddress MAC MAC1 MAC6 MAC1 MAC6 LAN Reset LAN RESET RESET Table 5 1 Front Panel Setup 5 2 1 Default Setting Refer to Table 3 5 Non Volatile Memory Parameters and Table 3 6 LAN Non Volatile Memory Param...

Страница 59: ...hese four fields represent an IPv4 address Each IP field can be set to any value in the range of 0 to 255 1 Press the COMM button COMM LED illuminates INTFC message appears on the Voltage display 2 Ro...

Страница 60: ...d rate is locked to 115200 5 2 6 Language Selection RS232 RS485 USB Power Supply Series implement the SCPI standard interface to control the unit by remote communication In addition GEN legacy command...

Страница 61: ...ters cannot be changed via the front panel If an attempt is made to change parameters via the front panel Current display shows REM message for 2 seconds It is possible to return to Local mode by pres...

Страница 62: ...ns Refer to Figure 5 1 for the connector description Figure 5 1 J4 Rear Panel RS232 485 IN Connector Pinout RS232 485 IN Pinout 1 RS232 RX 2 RS232 TX 3 RS485 RXD 4 RS485 TXD 5 RS485 TXD 6 RS485 RXD 7...

Страница 63: ...onfiguration Refer to Figure 5 2 for the connector description Figure 5 2 J3 Rear Panel RS485 OUT Connector Pinout RS485 OUT Pinout 1 NC Not Connected 2 NC Not Connected 3 RS485 TXD 4 RS485 RXD 5 RS48...

Страница 64: ...es Db 9 Connector 8 Pin Connector Remarks Pin No Name Pin No Name Housing Shield Housing Shield 2 RX 2 TX TWISTED PAIR 3 TX 1 RX 5 SG 8 SG Figure 5 3 RS232 Cable With DB9 Connector P N GEN 232 9 DB 9...

Страница 65: ...rted Figure 5 6 USB Connection 1 Insert the software CD ROM shipped with the product into a CD ROM Drive The Auto Run menu automatically appears 2 Install the appropriate USB driver according to the o...

Страница 66: ...UI that programs and reads the power supply output and status Security settings to block multiple controllers and risky protocols Optional password protection to prevent unauthorized operation LAN pro...

Страница 67: ...over cable connection Auto Negotiate Selects fastest of 10Base T or 100Base T networks 10 or 100 Megabits per second NETWORK CONFIGURATION MAC Address TDK Lambda assigned 00 19 f9 xx xx xx xx xx xx is...

Страница 68: ...server with Java scripts COMMANDS SCPI SCPI 1999 compliant command set measure and status IEEE 488 2 Condition and event register tree MULTIPLE CONTROLLERS Single Client Setting Only one may be open a...

Страница 69: ...nel LEDs SECURITY Web Page Password Can set password to prevent unauthorized or accidental changes to LAN settings or power supply output settings Single Client Only Set to prevent multiple programs f...

Страница 70: ...omputer using the power supply s built in web pages or using SCPI programming commands To select the LAN mode 1 Press the COMM button on the front panel 2 Rotate the Current encoder to select LAN LAN...

Страница 71: ...sed A standard straight patch CAT 5 or better network cable A crossover cable The cable type is auto detected by the power supply NOTE The serial link RS 485 cable 0 5m long provided with the power su...

Страница 72: ...ch ON the power supply 2 Connect the LAN cable before or after the power supply is switched ON 3 For a network with a DHCP server wait about 10 seconds The rear panel green LAN Status LED turns ON 4 F...

Страница 73: ...IP SRD command refer to section 5 14 10 Range Any address 169 254 xxx xxx Any address Lifetime Address may change as the DHCP server assigns addresses dynamically to many instruments Fixed prefix for...

Страница 74: ...with a different way to use the host name This is shown in the table below DHCP Auto IP Static IP Hostname Default Hostname xxxxvvvv aaaa Refer to the above example xxxxvvvv aaaa Refer to the above e...

Страница 75: ...isplay 3 Press Current encoder to enter IP configuration Rotate Voltage encoder to view the IP address The voltage display shows IP1 IP4 by rotating Voltage encoder The current display shows the corre...

Страница 76: ...STATIC IP addressing DHCP and Auto IP addressing is disabled 2 Refer to SYSTem COMMunicate LAN IP SRD command to modify the IP address by communication 5 7 5 3 LAN Reset To reset the Power Supply LAN...

Страница 77: ...s accessible 1 Read the IP address from the front panel Refer to section 5 7 5 1 2 Open a web page browser program such as Internet Explorer Enter the Power Supply IP address as shown below 3 Press th...

Страница 78: ...alternate VISA resource descriptor may be the hostname Refer to section 5 7 4 5 RS 485 Address This address applies to a multi drop chain of power supplies only The RS 485 address of the LAN supply i...

Страница 79: ...N Button 2 Click Login 3 The password may be set or removed on the LAN Users web page 4 LAN Reset clears the password 5 7 6 4 Login Rules Up to 2 users may view the web pages of a Power Supply at the...

Страница 80: ...ng the RS 485 address Select RS 485 Multi drop Address This address is described in section 5 7 1 1 This address applies to a multi drop chain of RS 485 supplies only If there is no multi drop chain o...

Страница 81: ...ify 0 Faults For the selected supply these boxes show any fault which may be active in the supply B DC Power Protection Page On the DC Power tab the Protection button is available at the top of the pa...

Страница 82: ...box The settings can be changed only by logging in as admin Figure 5 13 DC Power System Tab D DC Power Utility Page On the DC Power tab the System button is available at the top of the panel This page...

Страница 83: ...ure the Power Supply s LAN settings A LAN Configure Page When the LAN tab is selected the Configure page opens Figure 5 15 LAN Configure Tab The following settings are available on the LAN Configure w...

Страница 84: ...tionless UDP sockets The multiple client setting is needed to allow more than one controller connection at a time and to enable UDP sockets connections Multicast DNS mDNS is a protocol which allows de...

Страница 85: ...ields are disabled gray on the web page If the server cannot make the assignment the Power Supply reverts to the Auto IP method described in section 5 7 4 4 on page 65 In this mode the user can change...

Страница 86: ...onds 30 minutes Ping Server Ping is a network utility that allows the computer to verify communication with the Power Supply This service can be disabled on the Modify panel for security reasons Auto...

Страница 87: ...There is no password protection for automation programming with VISA or sockets By default the old password is blank The new password must contain between 6 and 16 characters Password is case sensitiv...

Страница 88: ...ce_write Write text to an instrument VXI 11 Device_read Read text from an instrument VXI 11 Destroy_link Close link to an instrument 5 7 7 3 Opening the VISA Connection Test and automation programs ma...

Страница 89: ...The web page has a security setting to limit or enable the types of connections and the numbers of control computers called clients that may be connected at a time The single or multiple clients setti...

Страница 90: ...nds through a TCP socket the socket drivers may combine all the messages into one long packet Therefore it is necessary to add a terminator character to the end of each SCPI command NOTE All SCPI comm...

Страница 91: ...onnect over the Wide Area Network i e the global internet the following settings must be made in the network server 5 7 9 1 View Web Pages Over WAN The Power Supply LAN interface has a server for runn...

Страница 92: ...gure 5 21 for typical connections 3 Using the Linking cable supplied with each unit Refer to Figure 5 21 connect each unit s OUT connector to the next unit s IN connector Figure 5 21 Multi Power Suppl...

Страница 93: ...Supply in a Multi drop Chain All the SCPI commands may be sent to any one of the supplies in an RS 485 chain by first sending the INSTrument N SELect NR1 address command All commands and queries will...

Страница 94: ...drop chain will be set to 70 volts except for the supply at RS 485 address 4 that will be set to 90 volts NOTE RS 485 retransmit baud rate is automatically set to 115 200 bps if USB LAN or optional co...

Страница 95: ...D Power Supply ignores the Line Feed ASCII 10 0x0A character 5 9 3 Command Repeat The backslash character causes the last command to be repeated 5 9 4 Checksum The user may optionally add a checksum t...

Страница 96: ...OFF ON Boolean parameters represent a single binary condition that is either true or false I e for a false condition the Power Supply will accept OFF or 0 For a true condition the Power Supply will ac...

Страница 97: ...m dd Returns SRD Example 2017 12 17 5 10 5 Initialization Commands CLS Function Clear Status Sets FEVE and SEVE registers to zeroes Returns DSC OK ADR NR1 Function Send ADR and address 0 to 31 to open...

Страница 98: ...gle Master or Slave roles does not change 5 10 6 Output Commands PV NR2 Function Set digital programming output voltage value in Volts Parameters Range is limited by the Power Supply model OVP and UVP...

Страница 99: ...T ON will respond with E07 if the output cannot be turned on because of a latching fault OTP AC ENA ILC POFF SO shutdown Parameters 0 1 OFF ON Query OTP Returns Bool OFF ON FLD DSC Function Sets the F...

Страница 100: ...ameters Range is limited by the UVL programming range and voltage programmed value Refer to Table 5 7 for UVL programming range Query UVL Returns NR2 4 digit format UVP Bool Function Enable Disable Un...

Страница 101: ...commands are received by all Power Supplies connected to the BUS without any need of individual addressing commands All Power Supplies execute commands immediately There is no acknowledgment back to...

Страница 102: ...Voltage Programming Range NOTE The Power Supply can accept values higher by 5 than the table values However it is not recommended to program the Power Supply above the rated values Model Minimum A Max...

Страница 103: ...xiliary Commands LANG SCPI Function Set a communication protocol to the SCPI language The Power Supply addressing is lost after the communication language change RIE Bool Function Enable Disable ILC f...

Страница 104: ...MOSFET OFF Parameter 0 1 OFF ON Query REL1 Return Bool 0 1 REL2 Bool Function Set the Programmable Pin 2 state J1 20 0 MOSFET ON 1 MOSFET OFF Parameter 0 1 OFF ON Query REL1 Return Bool 0 1 5 10 9 Sta...

Страница 105: ...ts are updated as a transition from 0 to 1 Reading the register clears it All Zeros Returns NRh STAT Query Returns the value of the Operation Group Condition register which is a read only register tha...

Страница 106: ...4 Communication PC write ADR 06 Power Supply response OK PC write OUT 1 Power Supply should turn output ON Power Supply response OK PC write PV n for n values refer to Table 5 4 Power Supply response...

Страница 107: ...dd a checksum to the end of the command The checksum is followed by two hex characters If a command or a query has checksum the response will also have one There is no CR between the command string an...

Страница 108: ...sensitive It will recognize any case mixture such as TRIGGER Trigger and TRIGger Short form headers result in a faster program execution 5 12 8 Data Formats Data Formats Description NR1 Digits with a...

Страница 109: ...re brackets are optional entered without the Expressions enclosed in angle brackets are programming values entered without the Vertical bars separate alternative parameters For example BUS EXT BUS or...

Страница 110: ...s the Standard Event Status Enable register contents through power cycles The CLS clear status command does not clear the Standard Event Status Enable register Any or all conditions can be reported to...

Страница 111: ...ror QYE Query Error Set to 1 if the output queue is empty or if the data is missing even after a query has been sent OPC Operation Complete CLS Clear Status command clears all the bits in the Standard...

Страница 112: ...condition occurs Parameters 0 1 OFF ON Query PSC Returns Bool 0 1 RCL NR1 Function This command restores Power Supply to a state stored in the memory locations 1 through 4 previously stored by the SAV...

Страница 113: ...on Bit Position 7 6 5 4 3 2 1 0 Bit Name OPR ESB MAV QUE SYS Bit Weight 128 32 16 8 4 OPR Operation Condition Status Register Summary ESB Standard Event Status Register Summary MAV Output Queue Messag...

Страница 114: ...Those can be a single command or a group of commands Groups are comprised of commands that extend one or more levels below the root Commands followed by a question mark take only the query form 5 14...

Страница 115: ...meters change 5 14 3 Instrument Subsystem INSTrument COUPle DSC Function Defines the coupling between various Power Supplies All coupled power supplies receive communication commands Parameters NONE A...

Страница 116: ...OUTPut STATe Bool Function Refers to OUTPut STATe Bool command description Parameter 0 1 OFF ON GLOBal VOLTage AMPLitude NRf Function Refers to SOURce VOLTage LEVel IMMediate AMPLitude NRf command de...

Страница 117: ...signal polarity NORM same as interlock Output ON if ENA pin signal J1 10 is low 0 REV Output ON if ENA pin signal J1 10 is high 1 Parameter 0 1 REV NORM Query OUTPut ENA POLarity Return DSC REV NORM...

Страница 118: ...used to disable the Power Supply output if a transition between operation modes occurs Power Supply turns output off after a specified delay if there is a transition into CV mode or into CC mode This...

Страница 119: ...ted Parameter 0 1 2 OFF FSTR TRIG Query OUTPut TTLTrg MODE Return DSC OFF FSTR TRIG 5 14 7 Program Subsystem PROGram COUNter NR1 INFinity Function Sets the sequencers iterations counter Parameters 1 9...

Страница 120: ...nts are separated by commas The function accepts up to 100 parameters Parameters The range is limited by the Power Supply voltage programming range including UVP OVP settings Unit V Query PROGram LIST...

Страница 121: ...ingle step Parameters ONCE AUTO Query PROGram STEP Return DSC AUTO ONCE PROGram STORe NR1 Function Stores a sequence in a specified memory cell 1 2 3 or 4 Stored parameters mode voltage current dwell...

Страница 122: ...equired after the PROGram WAVE TIME command if long sequences are used prior sending any additional command PROGram WAVE VOLTage NRf NRf Function Specifies output voltage points in a wave Voltage poin...

Страница 123: ...te AMPLitude NRf Function Sets a digital programming output current value Parameter The range is limited by the Power Supply model Unit A Query SOURce CURRent LEVel IMMediate AMPLitude Return NR2 5 di...

Страница 124: ...eters 0 1 2 NONE LIST WAVE Query SOURce CURRent MODE Return DSC NONE LIST WAVE SOURce VOLTage EXTernal MODE DSC Function Sets the voltage programming source Front panel Communication Digital external...

Страница 125: ...W UP Return NR2 Example VOLT SLEW UP 1 Sets a digital voltage reference up programming slew to 1V mS VOLT SLEW UP returns digital voltage up programming slope value VOLT SLEW UP MAX and VOLT SLEW UP M...

Страница 126: ...ection UVP mode If enabled the output is disabled when measured output voltage reaches the UVL level UVP bit is set in the Questionable Condition register Parameters 0 1 OFF ON Query SOURce VOLTage PR...

Страница 127: ...0 1 NOTE Analog programming internal resistance sequencer functions should be disabled to enable Constant Power SOURce POWer LEVel NRf Function Sets the constant power limit level Parameters The range...

Страница 128: ...rnal Current Reference Set to 1 when analog current programming channel is set EVR External Voltage Reference Set to 1 when analog voltage programming channel is set CFB Current Foldback Enabled Set t...

Страница 129: ...t EVR External Voltage Reference Set to 1 when analog voltage programming channel is set CFB Current Foldback enabled Set to 1 when Current Foldback is active Set to 0 indicates Voltage Foldback is ac...

Страница 130: ...s follows Position Value Name Position 15 7 14 16384 POFF 6 13 5 12 4 11 2048 GERR 3 10 1024 PACK 2 9 512 UVP 1 8 256 ENA 0 Value 128 64 32 16 8 4 2 Name ILC OFF SO OVP FLD OTP AC POFF Power OFF Set t...

Страница 131: ...EVENt command for the complete registers list Returns NR1 0 65535 Example 00136 STATus QUEStionable ENABle NR1 Function Sets the value of the Questionable Condition Group Enable register This registe...

Страница 132: ...1 9600 19200 38400 57600 115200 SYSTem COMMunicate INTerface DSC Function Sets a communication interface The Power Supply addressing is lost after the communication interface change Parameters 0 1 2 3...

Страница 133: ...FF 01 6D SYSTem COMMunicate LAN RESet Function Sets LAN parameters to default Refer to factory reset LAN settings Table 3 6 LAN Non Volatile Memory Parameters SYSTem COMMunicate LANGuage GEN Function...

Страница 134: ...n firmware revision is seen if the query is sent through optional communication interface SYSTem PANel LOCK Query Returns the front panel unlocked 0 locked 1 state Return Bool 0 1 SYSTem PARallel Quer...

Страница 135: ...Power Supply analog programming monitor range Programming 5 10 Volts or 5 10k Monitoring 5 10 Volts Parameters 5 10 Query SYSTem RANGe Return DSC 5 10 SYSTem RIN LEVel NRf Function Sets the internal r...

Страница 136: ...command is received System state changing from LLO to REM after AC power recycle SYSTem SENSe STATe DSC Function Selects the output sensing point local or remote Parameters 0 1 LOC REM OFF ON Query S...

Страница 137: ...a time delay between trigger events of a specified trigger source to the start of any corresponding trigger action over the Power Supply output Parameters 0 000 10 000s Step 0 001 Sec Unit S Query TRI...

Страница 138: ...et settings Saves an instrument state Set service request enable register Return status byte Trigger Performs self test returns a result Wait for all operations to complete Aborts the triggered action...

Страница 139: ...ad stored program from memory Set trigger depends execution step Store program in memory Set WAVE output current points Set slopes time Set WAVE output voltage points Enable disable analog mode curren...

Страница 140: ...C address once Set LAN parameters to default Set communication language Enable error message Read system error messages Factory reset default Read Firmware versions Return F P lock state Accept presen...

Страница 141: ...er refer to section 6 2 3 6 1 1 LIST Mode The Output values change in steps determined by parameters in the LIST Output change is activated by input trigger refer to section 6 2 2 Program LIST paramet...

Страница 142: ...et the operation mode by SOURce CURRent MODE DSC and or SOURce VOLTage MODE DSC command s IDLE STATE INITIATE STATE DELAY STATE OUTPUT CHANGE STATE LINEAR STEP WAVE STEP COMPLETE PROG STEP AUTO PROG C...

Страница 143: ...points in the WAVE mode Time range is 1mSec 36 hours in steps of one milli second Time setting is available via the communication command PROGram WAVE TIME NRf NRf 6 1 3 4 Load Load a sequence from a...

Страница 144: ...the Front Panel Refer to section 5 14 7 to operate the sequencer functions via communication commands NOTE If sequencer system is in the INITIATED STATE following the INIT command or INIT function vi...

Страница 145: ...available triggers in the sources External trigger rear panel DB26 J1 22 Positive edge triggered pulse Front Panel trigger Current or Voltage encoder press Active only when power supply output is ON a...

Страница 146: ...6 3 Sequencer Trigger System Examples A combination of the Sequencer and the Trigger System provides the ability to generate any sequence and synchronize it 6 3 1 WAVE Mode Voltage Programming via Com...

Страница 147: ...1 Set dwell values 0 5 0 5 1 1 1 1 Seconds STEP AUTO Set AUTO step execution mode AUTO COUN 1 Set list execution iterations 1 TRIG SOUR BUS Select BUS trigger source via communication interface or fr...

Страница 148: ...O Set AUTO step execution mode AUTO TRIG SOUR BUS Select BUS trigger source via communication interface or front panel INIT CONT OFF Trigger system is enabled for a single trigger action INIT Trigger...

Страница 149: ...the communication command SYSTem RIN STATe Bool Internal resistance setting range is 0 001 to 1 ohm in steps of 0 001 ohms Internal resistance setting is available via the Front Panel menu refer to se...

Страница 150: ...by multiplication of rated voltage by rated current in steps of 1 watt Constant power limit setting is available via the Front Panel menu refer to section 2 7 or the communication command SOURce POWe...

Страница 151: ...y This function is mainly used to prevent discharge of batteries connected to the power supply output terminals Setting preload to ON state default configuration keeps the preload active as long as po...

Страница 152: ...lue 10 Volts CURR 300 Set current value 300 Amperes SYST RANG 10 Set analog programming range 10 CURR EXT MODE VOL Set analog current programming source Set external voltage to 4V Current programming...

Страница 153: ...the following communication commands SOURce CURRent SLEW DOWN NRf Set current reference down programming slew SOURce CURRent SLEW UP NRf Set current reference up programming slew SOURce VOLTage SLEW...

Страница 154: ...ng advanced parallel connection cable G P 1 Connect J7 M output connector from the master unit to the J6 S input connector of the slave unit 2 To connect additional units connect J7 M connector to the...

Страница 155: ...eve better current balance between power supplies 6 9 3 Advanced Parallel System Acknowledge The acknowledge system prevents unintended changes of advanced paralleling assembled systems Following init...

Страница 156: ...er supply turn on the master unit wait 5 seconds Read questionable condition register by sending STATus QUEStionable CONDition command If paralleling system is ready for acknowledge questionable condi...

Страница 157: ...and communications both its own faults and the slave units faults Each slave unit shows on the display its own faults only Slave units show OUT OFF in case of a fault in any of the other units In cas...

Страница 158: ...s errors faults and SRQ register structures The registers can be read or set via communication commands Two individual sets of registers are managed one set for the SCPI language another set for the G...

Страница 159: ...able condition fault register group holds a snapshot of the actual faults state of the power supply at a present time If a fault occurs the corresponding bit is set If a fault is removed the correspon...

Страница 160: ...roup Bit Number Decimal Value Bit Symbol Description 0 1 CV Constant Voltage operation mode 1 2 CC Constant Current operation mode 2 4 NFLT No Fault 3 8 TWI Trigger Wait 4 16 AST Auto Start Enabled 5...

Страница 161: ...ster is cleared at power supply power on Bit Number Decimal Value Bit Symbol Description 0 1 OPC Operation Complete 1 2 2 4 QYE Query Error 3 8 DDE Device Dependent Error 4 16 EXE Execution Error 5 32...

Страница 162: ...queue holds a message it sets the SYS bit in the status byte register refer to Figure 7 1 Error Number Error Description Error Event 0 No Error No error reported 100 Command Error Unit receives comman...

Страница 163: ...ny SYST ERR messages are stored in error queue Newest messages are discarded 301 PV Above OVP Attempt to program voltage above OVP setting 302 PV Below UVL Attempt to program voltage below UVL setting...

Страница 164: ...n attempt to set Constant Power Mode Internal Resistance Slew Rate or Sequencer while Analog Programming is ON 338 Slew Mode is ON An attempt to set Constant Power Mode Internal Resistance Analog Prog...

Страница 165: ...64 RQS Service request 7 128 OPR Operation condition group summary event Table 7 6 Service Request Enable Register SYS System error message available bit Refer to section 10 6 to read the available er...

Страница 166: ...ing Event register for each summary group to determine which events caused the summary bit to be set When an Event register is read it is cleared This action also clears the corresponding summary bit...

Страница 167: ...Tree Figure 7 2 GEN Registers Tree Diagram GEN register tree shown in Figure 7 2 describes the structure of the status faults messaging and service request registers Condition registers hold a snapsh...

Страница 168: ...of the power supply at a present time If a fault occurs the corresponding bit is set If a fault is removed the corresponding bit is cleared Some of the faults might quickly change their state clear fa...

Страница 169: ...back Enabled 6 64 7 128 LOC Local Remote communication mode 8 256 UVP Under Voltage Protection enabled 9 512 ILC enable Interlock control Enabled 10 1024 ENA enable Enable control Enabled 11 2048 CFB...

Страница 170: ...eturned if an attempt to set an operation mode Internal Resistance Constant Power Sequencer Slew Rate or Analog Programming if another operation mode is enabled or loading an empty memory cell Only on...

Страница 171: ...ht control wires and eight ground wires If the system runs from a personal computer there are numerous vendors of IEEE controller cards and software Because many instruments may be connected and indep...

Страница 172: ...a Multi Drop Chain All of the SCPI commands may be sent to any one of the Power Supplies in RS485 chain by sending the INSTrument NSELect NR1 command All commands and queries will then apply only to t...

Страница 173: ...ce 8 7 Power Supply Configuration To communicate over the IEEE interface Power Supply has to be configured to IEEE If IEEE option is installed by default Power Supply is pre configured to the IEEE com...

Страница 174: ...nal Instruments MAX application 1 Run the National Instruments MAX Measurement Automation Explorer Figure 8 3 National Instruments MAX Desktop Icon 2 On the side tree expand My System and select Devic...

Страница 175: ...roperties 5 In the GPIB Explorer toolbar click Communicate with Instrument NI 488 2 Communicator appears Refer to Figure 8 6 Figure 8 6 ID String Query In the Send String text box IDN appears 6 Click...

Страница 176: ...m For 10V model only Derate 1 C 100m Or 2 Load 100m above 2000m NOTE For power supplies with Air filter option and GPIB option please contact TDK Lambda customer support department or your local sales...

Страница 177: ...169 9 3 2 Assembly Instructions for Standard Power Supplies...

Страница 178: ...170...

Страница 179: ...171 9 3 3 Assembly Instructions for Blank Power Supplies Without Display...

Страница 180: ...172...

Страница 181: ...iodic cleaning To clean the device 1 Disconnect the unit from the AC supply and allow 30sec for discharging internal voltage 2 To clean the front panel and the metal surfaces use a mild solution of de...

Страница 182: ...nuity and replace if necessary 3 6 Is the AC input voltage within range Check input AC voltage Connect to appropriate voltage source Safety Instructions 3 6 Output is present momentarily but shuts off...

Страница 183: ...Front panel CV indicator is on Are sensing wires connected properly Connect the sense wires according to User Manual instructions 3 9 4 Front Panel controls are non functional Is the power supply in L...

Страница 184: ...nden Im Ger t angesammelter Staub kann mit Hilfe von Niedrigdruck Pressluft herausgeblasen werden Bei Ger ten mit optional angebrachtem Luftfilter Aufsatz entnehmen Sie bitte die Wartungshinweise aus...

Страница 185: ...6 Ausgangsspannung l sst sich nicht einstellen CV LED an der Frontplatte leuchtet Pr fen ob die Einstellung der Ausgangsspannung ber der OVP oder unter der UVL Begrenzung liegt OVP bzw UVL so einstel...

Страница 186: ...leuchtet Stromeinstellregler dr cken um zwischen den Modi Frontplatte verriegelt Lock und Frontplatte nicht verriegelt Unlock zu wechseln Das Display zeigt dann den gew hlten Modus an 5 2 7 Kein Ausga...

Страница 187: ...Daisy Chain Connection 38 to RS232 84 Data Formats 100 Enable In Function 40 Enable_in Polarity 40 Execution Time 165 Front Panel Display Messages 19 Fuse Rating 175 GEN Command Set Description 87 Re...

Страница 188: ...ds 106 SCPI Protocol 99 Sequencer Functions 135 Serial Communication Test Set Up 98 Serial RS232 RS485 50 Series Connection 36 Output Voltage 37 Series Operation Remote Programming 37 Service Request...

Отзывы: