TDK-Lambda DTM160 Series Скачать руководство пользователя страница 2

DTM160, DTM160-C8 Manual

 

External AC-DC Power Adapters, 160W 

 

1.  P

RODUCT 

D

ESCRIPTION

 

With the two series DTM160 and DTM160-C8 TDK-Lambda offers compact 150-160W medical AC-DC power 

adapters. 
Both, the DTM160 as a Class I version with protective earth on the input and the DTM160-C8 as Class II 

version with double insulated protection uses a robust desktop housing with the dimensions of only 150 x 70 

x 38mm. Both series are members of the DTM family, whose devices are available in various power classes 
between 36W and 300W. 
The two power adapter series are suitable for 2MOPP medical as well as for general industrial applications. 

The DTM160-C8 also has an IEC 60601-1-11 approval, making it ideal for medical, clinical and home 
healthcare applications. 
All devices are approved for installation altitudes up to 5000m and EMC is tested according to IEC/EN 60601-

1-2 Ed 4. 
Both series are available with 12V, 24V, 36V and 48V output voltage. 
Due to the high efficiencies and the low no-load losses, the devices are compliant with EU Regulation 

2019/1782 (Ecodesign requirements for external power supplies) and the US DoE Level VI specification. 
Special features are the compact design, achieved by the high efficiency and the standard locking DIN output 
connector, which prevents unintentional output voltage disconnection. Other connector versions and cable 

assemblies are available on request. 
Safety is assured by IEC/UL/CSA/EN 60601-1 and EN 62368-1. CE and UKCA markings confirm compliance 
with the low voltage, EMC and RoHS directives. 

2.  P

RODUCT 

F

EATURES

 

Small in size, only 150 x 70 x 38mm 

Protection Class I for DTM160 and II for DTM160-C8 

IEC/UL/EN 60601-1 - 2MOPP approved 

4400Vac Input to output isolation 

IEC/EN 60601-1-2 Ed 4 (EMC) tested 

IEC/EN 60601-1-11 home healthcare approval for DTM160-C8 

IEC/BS/EN 62368-1 (I.C.T.) compliant 

Temperature range -20 to +70°C (with derating) 

AC 100-240V universal input voltage 

Locking output connector 

Operating altitude up to 5000m 

High efficiencies up to 93% 

Low no-load power consumption <150mW 

Compliant with EU Regulation 2019/1782 (ecodesign for external power supplies) 

Meets US DoE Level VI specifications 

All parameters are typical values specified at nominal input voltage, nominal output load, 25°C ambient and after a 5 minutes run-in time unless  
otherwise noted. The information presented in this document is believed to be accurate and reliable and may change without notice. 

For additional information, please visit 

https://product.tdk.com/en/power/dtm

 

2/13

DOC-ID: 260933 – Issue 1

Содержание DTM160 Series

Страница 1: ...y Curves 7 11 EMC 8 12 Physical Dimensions Weight and Connectors 9 13 Environment 10 14 Isolation and Dielectric Strength 11 15 Safety and Protection Features 11 16 Approvals and Regulatory Compliance...

Страница 2: ...the compact design achieved by the high efficiency and the standard locking DIN output connector which prevents unintentional output voltage disconnection Other connector versions and cable assemblies...

Страница 3: ...The device is designed to operate in areas with a maximum relative humidity of 93 4 INSTALLATION INSTRUCTIONS Before operating this device please read this manual thoroughly and retain this manual for...

Страница 4: ...itive load is higher than the values specified the device operates in an intermittent mode hiccup mode See short form data table for maximum values The output voltage is fixed and can not be adjusted...

Страница 5: ...xternal Adapter Power Supply Class I 160W 48V 3 33A DTM160CN120C8 External Adapter Power Supply Class II 150W 12V 12 5A DTM160CN240C8 External Adapter Power Supply Class II 160W 24V 6 66A DTM160CN360C...

Страница 6: ...0CN120 DTM160CN120C8 For all remaining models Bandwidth 20MHz 25degC ambient 100nF ceramic capacitors and 10uF aluminium capacitor to output Efficiency Typ Typ Typ Typ 91 0 92 0 92 0 93 0 For DTM160CN...

Страница 7: ...ency vs output load DTM160CN360 DTM160CN360C8 Efficiency vs output load DTM160CN480 DTM160CN480C8 Efficiency vs output load All parameters are typical values specified at nominal input voltage nominal...

Страница 8: ...les class 3 80Vac 250 300 cycles class 3 Criterion A Criterion A Criterion B Criterion A Criterion A Voltage interruptions EN 61000 4 11 EN 60601 1 2 AC 100 120 230Vac 0Vac 250 300 cycles class 3 Crit...

Страница 9: ...KPPx 4P or equivalent P1 P2 Vout P3 P4 Vout Alternative output connectors are available Please contact Sales for further information minimum quantities apply All parameters are typical values specifie...

Страница 10: ...rding to IEC 60664 1 Degree of pollution 2 According to IEC 62477 1 not conductive Vibration sinusoidal 4g 15 2000Hz 60 minutes per axis According MIL STD 810F Table 514 5C VIII sinusoidal vibration u...

Страница 11: ...0 DTM160CN240C8 For DTM160CN360 DTM160CN360C8 For DTM160CN480 DTM160CN480C8 In case of an internal defect a redundant circuit limits the maximum output voltage The output shuts off Cycle input voltage...

Страница 12: ...t 1 General requirements for basic safety and essential performance Part 1 6 Collateral standard Usability Part 1 11 Collateral Standard Requirements for medical electrical equipment and medical elect...

Страница 13: ...oted The information presented in this document is believed to be accurate and reliable and may change without notice For additional information please visit https product tdk com en power dtm 17 PACK...

Отзывы: