Medix T2x Series
Battery Contact PCB Replacement Instructions
2
How to replace the Battery Contact PCB for
Medix T22B & T24B
Medical-Cart Computer: Some of the pictures in this
instructions may not be exactly the same as the product you
have but the procedures should be the same.
NOTE:
PRE
PARE A CLEAN WORKING SURFACE TO PROTECT SYSTEM’S
DISPLAY & CASE FROM SCRATCHES & DAMAGE.
If a Stand is not installed on the unit, please skip items 1 & 2, & go directly to
item 3.
1.
Slide the Stand Cap up. Remove the Stand Cap.
2. Remove 4- M4 screws. Remove system from the stand.