Chapter 20 – Technical Specifications
20.7 Electromagnetic Immunity
The t:slim X2 Pump is intended for use in the electromagnetic environment specified below. Always make sure that the pump is
used in such an environment.
Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
Immunity Test
IEC 60601 Test Level
Compliance Level
Electromagnetic Environment –
Guidance
Electrostatic Discharge (ESD)
IEC 61000-4-2
± 8 kV contact
± 15 kV air
± 8 kV contact
± 15 kV air
Floors should be wood, concrete or ceramic tile. If floors
are covered with synthetic material, the relative humidity
should be at least 30%.
Electrical Fast Transient/burst
IEC 61000-4-4
± 2 kV for power supply lines
± 1 kV for input/ output lines
± 2 kV for power supply lines
± 1 kV for input/output lines
Mains power quality should be that of a typical commercial
or hospital environment.
Surge
IEC 61000-4-5
± 1 kV differential mode
± 2 kV common mode
± 1 kV differential mode
± 2 kV common mode
Mains power quality should be that of a typical commercial
or hospital environment.
Voltage Dips, Short Interruptions,
and Voltage Variations on Power
Supply Input Lines IEC 61000-4-11
<5% Ur (>95% dip in Ur) for 0,5 cycle
40% Ur (60% dip in Ur) for 5 cycles
70% Ur (30% dip in Ur) for 25 cycles
<5% Ur (>95% dip in Ur) for 5 sec
<5% Ur (>95% dip in Ur) for 0,5 cycle
40% Ur (60% dip in Ur) for 5 cycles
70% Ur (30% dip in Ur) for 25 cycles
<5% Ur (>95% dip in Ur) for 5 sec
Mains power quality should be that of a typical commercial
or hospital environment. If the user of the pump requires
continued operation during power mains interruptions,
it is recommended that the pump be powered from an
uninterruptible power supply or a battery.
NOTE: Ur is the a.c. mains voltage prior to application of
the test level.
Power Frequency (50/60 Hz)
Magnetic Field IEC 61000-4-8
3 A/m
400 A/m (IEC 60601-2-24)
Power frequency magnetic fields should be at levels
characteristic of a typical location in a typical commercial
or hospital environment.
table continued on next page …
Содержание t:slim X2
Страница 2: ......
Страница 14: ...Section 1 Before You Begin...
Страница 15: ...Chapter 1 Introduction...
Страница 19: ...Chapter 2 Important Safety Information...
Страница 31: ...Chapter 3 Getting to Know Your t slim X2 Pump...
Страница 54: ...Section 2 Key Pump Features...
Страница 55: ...Chapter 4 Getting Started...
Страница 60: ...Customer Technical Support 877 801 6901 60 Chapter 4 Getting Started THIS PAGE IS INTENTIONALLY LEFT BLANK...
Страница 61: ...Chapter 5 Infusion Site Care and Loading Cartridge...
Страница 75: ...Chapter 6 Personal Profiles...
Страница 84: ...Customer Technical Support 877 801 6901 84 Chapter 6 Personal Profiles THIS PAGE IS INTENTIONALLY LEFT BLANK...
Страница 85: ...Chapter 7 Bolus...
Страница 92: ...Customer Technical Support 877 801 6901 92 Chapter 7 Bolus THIS PAGE IS INTENTIONALLY LEFT BLANK...
Страница 93: ...Chapter 8 Stop Resume Insulin...
Страница 96: ...Section 3 Additional Pump Operations...
Страница 97: ...Chapter 9 Temporary Basal Rate...
Страница 100: ...Customer Technical Support 877 801 6901 100 Chapter 9 Temporary Basal Rate THIS PAGE IS INTENTIONALLY LEFT BLANK...
Страница 101: ...Chapter 10 Quick Bolus...
Страница 104: ...Customer Technical Support 877 801 6901 104 Chapter 10 Quick Bolus THIS PAGE IS INTENTIONALLY LEFT BLANK...
Страница 105: ...Chapter 11 t slim X2 Pump Settings...
Страница 108: ...Customer Technical Support 877 801 6901 108 Chapter 11 t slim X2 Pump Settings THIS PAGE IS INTENTIONALLY LEFT BLANK...
Страница 109: ...Chapter 12 t slim X2 Pump Info and History...
Страница 111: ...Chapter 13 t slim X2 Pump Reminders...
Страница 116: ...Customer Technical Support 877 801 6901 116 Chapter 13 t slim X2 Pump Reminders THIS PAGE IS INTENTIONALLY LEFT BLANK...
Страница 117: ...Chapter 14 User Settable Alerts and Alarms...
Страница 120: ...Section 4 t slim X2 Pump Safety Alerts and Alarms...
Страница 121: ...Chapter 15 t slim X2 Pump Alerts...
Страница 143: ...Chapter 16 t slim X2 Pump Alarms...
Страница 157: ...Chapter 17 t slim X2 Pump Malfunction...
Страница 160: ...Section 5 Living With and Caring for Your t slim X2 Pump...
Страница 161: ...Chapter 18 Lifestyle Issues and Travel...
Страница 164: ...Customer Technical Support 877 801 6901 164 Chapter 18 Lifestyle Issues and Travel THIS PAGE IS INTENTIONALLY LEFT BLANK...
Страница 165: ...Chapter 19 Taking Care of Your t slim X2 Pump...
Страница 168: ...Section 6 Technical Specifications and Warranty...
Страница 169: ...Chapter 20 Technical Specifications...