Chapter 5 – Infusion Site Care and Loading Cartridge
Instructions for Filling the Cartridge
1.
Inspect the cartridge package for any signs of damage.
Discard any damaged product.
2.
Open the package and remove the cartridge.
3.
Hold the cartridge upright and gently insert the needle into the white
insulin fill port on the cartridge (see image E). The needle is not intended
to go all the way in, so do not force it.
4.
Keeping the syringe vertically aligned with the cartridge, and the needle
inside the fill port, pull back on the plunger until it is fully retracted
(see image F). This will remove any residual air from the cartridge.
Bubbles will rise toward the plunger.
5.
Make sure the needle is still in the fill port and release the plunger.
Pressure will pull the plunger to its neutral position but it will NOT
push any air back inside the cartridge (see image G).
Содержание t:slim X2
Страница 2: ......
Страница 14: ...Section 1 Before You Begin...
Страница 15: ...Chapter 1 Introduction...
Страница 19: ...Chapter 2 Important Safety Information...
Страница 31: ...Chapter 3 Getting to Know Your t slim X2 Pump...
Страница 54: ...Section 2 Key Pump Features...
Страница 55: ...Chapter 4 Getting Started...
Страница 60: ...Customer Technical Support 877 801 6901 60 Chapter 4 Getting Started THIS PAGE IS INTENTIONALLY LEFT BLANK...
Страница 61: ...Chapter 5 Infusion Site Care and Loading Cartridge...
Страница 75: ...Chapter 6 Personal Profiles...
Страница 84: ...Customer Technical Support 877 801 6901 84 Chapter 6 Personal Profiles THIS PAGE IS INTENTIONALLY LEFT BLANK...
Страница 85: ...Chapter 7 Bolus...
Страница 92: ...Customer Technical Support 877 801 6901 92 Chapter 7 Bolus THIS PAGE IS INTENTIONALLY LEFT BLANK...
Страница 93: ...Chapter 8 Stop Resume Insulin...
Страница 96: ...Section 3 Additional Pump Operations...
Страница 97: ...Chapter 9 Temporary Basal Rate...
Страница 100: ...Customer Technical Support 877 801 6901 100 Chapter 9 Temporary Basal Rate THIS PAGE IS INTENTIONALLY LEFT BLANK...
Страница 101: ...Chapter 10 Quick Bolus...
Страница 104: ...Customer Technical Support 877 801 6901 104 Chapter 10 Quick Bolus THIS PAGE IS INTENTIONALLY LEFT BLANK...
Страница 105: ...Chapter 11 t slim X2 Pump Settings...
Страница 108: ...Customer Technical Support 877 801 6901 108 Chapter 11 t slim X2 Pump Settings THIS PAGE IS INTENTIONALLY LEFT BLANK...
Страница 109: ...Chapter 12 t slim X2 Pump Info and History...
Страница 111: ...Chapter 13 t slim X2 Pump Reminders...
Страница 116: ...Customer Technical Support 877 801 6901 116 Chapter 13 t slim X2 Pump Reminders THIS PAGE IS INTENTIONALLY LEFT BLANK...
Страница 117: ...Chapter 14 User Settable Alerts and Alarms...
Страница 120: ...Section 4 t slim X2 Pump Safety Alerts and Alarms...
Страница 121: ...Chapter 15 t slim X2 Pump Alerts...
Страница 143: ...Chapter 16 t slim X2 Pump Alarms...
Страница 157: ...Chapter 17 t slim X2 Pump Malfunction...
Страница 160: ...Section 5 Living With and Caring for Your t slim X2 Pump...
Страница 161: ...Chapter 18 Lifestyle Issues and Travel...
Страница 164: ...Customer Technical Support 877 801 6901 164 Chapter 18 Lifestyle Issues and Travel THIS PAGE IS INTENTIONALLY LEFT BLANK...
Страница 165: ...Chapter 19 Taking Care of Your t slim X2 Pump...
Страница 168: ...Section 6 Technical Specifications and Warranty...
Страница 169: ...Chapter 20 Technical Specifications...