7.6.2
Choosing the Sensor Mounting Location
The performance of the sensor will be enhanced through careful consideration of the mounting
location. The following guidelines should be followed:
The sensor should not be mounted in the bottom of a sump since the sensor head may
become
restricted preventing correct operation.
Dynamic oil flow is necessary; do not mount in places where the oil is likely to stagnate
or be static, since the oil in the sensor needs to be representative of the whole system.
When the oil quality sensor is mounted in a pipeline, please ensure that the sensor will
not restrict flow.
When mounting the oil quality sensor in a lubrication system, for maximum performance,
please ensure the sensor is located prior to the oil filters, oil coolers etc. to ensure oil is
representative of the whole system.
7.6.3
Fitting Method
The standard thread is ½ inch BSPP/NPT requiring an 32mm spanner for installation.
Decide on an appropriate location for sensor head installation.
Drain the lubricant sufficiently to allow the sensor to be fitted.
Install the sensor head into the selected location/position, torque to 5Nm being careful
not to over tighten. Ensure the integral cable is not twisted excessively during this
process.
Route the cable, fixing (cable ties) at appropriate intervals.
Avoid sharp edges and hot surfaces.
Connect the sensor to the chosen interface.
Figure 16 - Installing the Sensor
Содержание OQSExISG2
Страница 1: ...UG OQSxG2 23 0025 OIL QUALITY SENSOR ASSOCIATED ACCESSORIES USER GUIDE...
Страница 2: ......
Страница 5: ...Page 3 Page Intentionally Blank...
Страница 37: ...Page 35 Figure 19 Connection to other Tan Delta and 3rd Party Products Interfaces...
Страница 38: ...Page 36 Figure 20 ExOQSxG2 Ex Sensor Installation Drawing Rev 2 STAMPED...
Страница 42: ...Page 40 8 Display Express...
Страница 97: ...Page 95 P4 CANL RS485A P5 0V P6 CANH RS485B...