Page | 120
13.1.4
Cleaning Procedure
–
Loctite 7063 or Similar Solvent
Unplug and remove the sensor then use absorbent paper to wipe off the excess oil from
the sensor tip and thread.
While rotating the sensor, force a short blast of solvent up into each of the holes in the
sensor nose, and then give the general exterior a wash on both sides as shown in the
pictures below:
Figure 85 - Cleaning the Sensor (1)
Then give a slightly longer spray up the centre hole (2 seconds).
Shake sensor over the absorbent paper to dislodge any solvent remaining around the
electrode.
Leave to dry for a few minutes.
Figure 86 - Cleaning the Sensor (2)