Making Settings
By making settings for auto-upload, it is possible to automatically
upload recorded data to the online T&D WebStorage Service.
• In the auto-upload, the maximum number of readings which can be stored in
T&D WebStorage Service is 80,000. When this limit is exceeded, the oldest data will be
overwritten and deleted.
• Wireless LAN settings via operation of buttons on the logger can be made only when using
a wireless LAN access point supporting WPS.
• If your access point does not support WPS, use the T&D application "TR-7wf for Windows"
or "ThermoWEB" to make wireless LAN settings. Make sure you have your Wireless LAN
SSID, Security Mode, and Security Key (Password) information.
• The time difference setting can be changed via TR-7wf for Windows or ThermoWEB. The
default time difference is as follows:
Serial number "No.4XXXXXXX" : GMT +1:00
Serial number "No.3XXXXXXX" : GMT -8:00
11
Press-and-hold the
<POWER> and <DISPLAY>
buttons together until
appears on the LCD screen.
21
Set your wireless LAN access point to "WPS" mode.
For details on the wireless LAN access point, contact the manufacturer of your access point.
31
If the setup is successfully completed, the
mark on the LCD
display will turn ON.
If
keeps blinking:
It means the device is unable to connect to the wireless network. After adjusting the
distance between the device and the access point, start from step 1 again.
41
Click the <INTERVAL> button
to display the current upload
interval.
51
Press-and-hold the <INTERVAL> button until the display flashes.
61
With each pressing of the <INTERVAL> button the upload inter-
val time will change. When the desired interval appears, stop
pressing the button.
If you wish NOT to use the auto-upload feature, select "OFF".
Below are some examples of upload intervals and the corresponding
estimated battery life1
In general, the shorter the upload interval, the shorter the battery life.
Upload Interval
1 min.
10 min.
1 hr.
12 hrs.
Battery Life
About 10 days
About 2 months
About 1 year
About 1.5 years
* All estimates are based on operations carried out with a new battery and are in no way a
guarantee of actual battery life.
* The auto-upload interval can also be changed during a recording session.
Installing the Device
Install the device to the actual place it will be used.
To perform auto-upload, make sure to install the device at a place
that has wireless LAN Internet access.
Check Wireless LAN Connection Status
Internet
Wireless LAN
Access Point
Checking Connection Status
By pressing the <REC/STOP> button, recorded data will be sent to T&D WebStorage Service.
If the transmission is successful,
will still be ON even after communication is complete
and
turns OFF.
is blinking : Wireless LAN connection failed. Try adjusting the position of the device or its
distance from the access point.
is blinking : Internet connection failed. Confirm whether your access point can connect
to the Internet.
Checking Recorded Data
Check recorded data uploaded to the T&D WebStorage Service.
11
Access T&D WebStorage Service from the web browser on your
PC or mobile device.
T&D WebStorage Service : http://www.webstorage-service.com/
21
Follow the directions as they appear to get a User ID.
If you already have your User ID, go to 3 below.
31
Login by entering the registered User ID and Password.
41
Clicking the [Account] button will take you to the Account Man-
agement page.
51
Select [Device Registration], and enter the serial number and
registration code for the target TR-7wf Series and then click the
[Register this device] button.
The registration code can be found on the supplied Registration
Code Label.
61
At first, only the added TR-7wf Series device will appear in the
[Current Readings] tab page.
Measurements will appear after clicking the <REC/STOP> button
on the device or when data is auto-uploaded from the device.
If measurements are not displayed:
* Check the device screen to make sure
is ON. Recorded data will not be uploaded
when recording is not in progress.
* See "Checking Connection Status" in , and check if the device has a wireless LAN connection.
If the display of measurements is not refreshed:
We recommend making the recording interval shorter than the auto-upload interval. If the
recording interval is set to longer than the upload interval, the same measurement is displayed
until the next recording occurs and the new data is uploaded.
View Data from other than a Web Browser
Recorded data which was sent to the T&D WebStorage Service can
also be viewed with ThermoWEB and/or T&D Graph.
Saving Auto-Uploaded Data
Recorded data which has been uploaded to the
T&D WebStorage Service can be saved in the PC by using
T&D Graph and accessing T&D WebStorage Service.
For details, please see
printed on the back of this paper.
Changing the Settings
By using the T&D WebStorage Service, it is possible to change the
following settings:
• Auto-upload Interval
• Warning Settings
The changes in settings will be applied when the device communicates with the
T&D WebStorage Service. To apply the changes immediately, press the <REC/STOP>
button on the device after settings are completed.
Thermo Recorder
TR-7wf Series Introductory Manual
What can TR-7wf Series devices do?
Auto-upload of Recorded Data
*1
Wireless LAN
Communication
View with a Web Browser
Setup and Data Download
Setup and Data Download
Graph Display
Graph Display
Setup
USB
Communication
*2
Smartphone
Tablet device
PC
TR-7wf Series
TR-7wf Series
TR-7wf Series
Graph Display
Download Data
Wireless
Communication
*3
(Wireless LAN)
Apply Settings
Via Cloud
Via USB
Via Wireless
*1: To auto upload recorded data to the T&D WebStorage Service, it is necessary to have Internet access via wireless LAN and a T&D WebStorage Service User ID.
*2: A USB communication cable (sold separately) is required for connecting the TR-7wf series device to your computer.
*3: A mobile device with wireless LAN capability is required for direct communication with the TR-7wf series device.
Setting up the Device
11
Connect sensors to the
device.
To ensure a proper connection, make sure that the sensor connector is completely
inserted.
21
Remove the battery cover and
insert batteries.
* Make sure not to m / -.
* Always use new batteries.
31
The measurement will be displayed on the LCD screen.
Under Normal Conditions:
During Sensor Error (no sensor connected /
improper connection / damaged):
Starting Recording
• Upon the start of a new recording session all stored data will be deleted from the device.
• Recording Interval settings cannot be changed while a recording session is in progress.
Stop recording first and then change settings.
• During recording, the power cannot be turned off.
11
Click the <INTERVAL> button
to display the current record-
ing interval.
21
Press-and-hold the <INTERVAL> button until the display flashes.
31
With each pressing of the <INTERVAL> button the recording
interval time will change. When the desired interval appears, stop
pressing the button.
41
Start recording by pressing
and holding the <REC/STOP>
button until the
mark
appears in the display.
Below are some examples of recording intervals and maximum recording times1
1 sec.
30 sec.
5 min.
15 min.
60 min.
About 2 hours
About 2 days
About 27 days
About 83 days
About 333 days
In order to stop a recording session, press and hold the <REC/STOP> button until the
mark disappears from the display.
Details about the LCD Display
Other than the measurement display, the following is displayed on
the LCD monitor.
Shows the status of wireless LAN connection.
ON: Connected to the wireless network. (Signal strength: 1 to
3 bars)
BLINKING : Unable to connect to the wireless network.
OFF : Settings have not been made for the wireless LAN
connection.
Shows the status of the Internet connection.
ON : Connected to the Internet
BLINKING : Unable to connect to the Internet
OFF : Auto-upload is set to “OFF”.
Shows the channel number of the measurement being displayed.
This appears when direct communication with mobile devices is possible.
This appears when the wireless LAN setup using WPS is possible.
This appears when the device is in wireless LAN communication or USB communication.
Shows the recording status.
ON : Recording in progress BLINKING : Waiting for programmed start
OFF : Recording stopped
Shows the recording mode, which can be changed via TR-7wf for Windows or
ThermoWEB.
ENDLESS : Upon reaching the logging capacity of 8,000 readings, the oldest data will be
overwritten and recording will continue.
ONETIME : Upon reaching the logging capacity of 8,000 readings, [FULL] will appear on
the LCD display and recording will automatically stop.
The amount of recorded data in the device is shown here.
This appears when the battery level is low. Please change the batteries as soon as you see
this mark. If the batteries are not changed but remain in use, the device will enter sleep
mode ([SLP] will appear on LCD) and stop all operations. If the batteries are further left
unchanged, all previously recorded data will be lost.
This appears while the auto-upload interval is displayed by using the <INTERVAL> button.
This appears while the recording interval is displayed by using the <INTERVAL> button.
Common Operations
Operations Via Cloud
© Copyright T&D Corporation. All rights reserved.
2013. 09. 16504930004 1st Edition
http://www.tandd.com/
Distributed by MicroDAQ.com, Ltd
www.MicroDAQ.com
(603) 746-5524