background image

 Making Settings

By making settings for auto-upload, it is possible to automatically 

upload recorded data to the online T&D WebStorage Service.

• In the auto-upload, the maximum number of readings which can be stored in 

T&D WebStorage Service is 80,000. When this limit is exceeded, the oldest data will be 
overwritten and deleted.

• Wireless LAN settings via operation of buttons on the logger can be made only when using 

a wireless LAN access point supporting WPS. 

• If your access point does not support WPS, use the T&D application "TR-7wf for Windows" 

or "ThermoWEB" to make wireless LAN settings. Make sure you have your Wireless LAN 
SSID, Security Mode, and Security Key (Password) information.

• The time difference setting can be changed via TR-7wf for Windows or ThermoWEB. The 

default time difference is as follows:  
Serial number "No.4XXXXXXX" : GMT +1:00 
Serial number "No.3XXXXXXX" : GMT -8:00

11 

Press-and-hold the 
<POWER> and <DISPLAY> 
buttons together until 

 

appears on the LCD screen.

21 

Set your wireless LAN access point to "WPS" mode.

For details on the wireless LAN access point, contact the manufacturer of your access point.

31 

If the setup is successfully completed, the 

 mark on the LCD 

display will turn ON.

If 

 keeps blinking:

It means the device is unable to connect to the wireless network. After adjusting the 

distance between the device and the access point, start from step 1 again.

41 

Click the <INTERVAL> button 
to display the current upload 
interval.

51 

Press-and-hold the <INTERVAL> button until the display flashes.

61 

With each pressing of the <INTERVAL> button the upload inter-
val time will change. When the desired interval appears, stop 
pressing the button.

If you wish NOT to use the auto-upload feature, select "OFF".

Below are some examples of upload intervals and the corresponding 
estimated battery life1

In general, the shorter the upload interval, the shorter the battery life.

Upload Interval

1 min.

10 min.

1 hr.

12 hrs.

Battery Life

About 10 days

About 2 months

About 1 year

About 1.5 years

*  All estimates are based on operations carried out with a new battery and are in no way a 

guarantee of actual battery life.

*  The auto-upload interval can also be changed during a recording session.

 Installing the Device

Install the device to the actual place it will be used.
To perform auto-upload, make sure to install the device at a place 

that has wireless LAN Internet access.

Check Wireless LAN Connection Status

Internet

Wireless LAN 
Access Point

Checking Connection Status

By pressing the <REC/STOP> button, recorded data will be sent to T&D WebStorage Service. 

If the transmission is successful, 

 will still be ON even after communication is complete 

and 

  turns OFF.

 is blinking : Wireless LAN connection failed. Try adjusting the position of the device or its 

distance from the access point.

 is blinking : Internet connection failed. Confirm whether your access point can connect 

to the Internet.

 Checking Recorded Data

Check recorded data uploaded to the T&D WebStorage Service.

11 

Access T&D WebStorage Service from the web browser on your 
PC or mobile device.
T&D WebStorage Service : http://www.webstorage-service.com/

21 

Follow the directions as they appear to get a User ID.
If you already have your User ID, go to 3 below.

31 

Login by entering the registered User ID and Password.

41 

Clicking the [Account] button will take you to the Account Man-
agement page.

51 

Select [Device Registration], and enter the serial number and 
registration code for the target TR-7wf Series and then click the 
[Register this device] button.
The registration code can be found on the supplied Registration 
Code Label.

61 

At first, only the added TR-7wf Series device will appear in the 
[Current Readings] tab page.

Measurements will appear after clicking the <REC/STOP> button 
on the device or when data is auto-uploaded from the device.

If measurements are not displayed:

*  Check the device screen to make sure 

 is ON. Recorded data will not be uploaded 

when recording is not in progress.

*  See "Checking Connection Status" in  , and check if the device has a wireless LAN connection.

If the display of measurements is not refreshed:

We recommend making the recording interval shorter than the auto-upload interval. If the 

recording interval is set to longer than the upload interval, the same measurement is displayed 

until the next recording occurs and the new data is uploaded.

View Data from other than a Web Browser

Recorded data which was sent to the T&D WebStorage Service can 

also be viewed with ThermoWEB and/or T&D Graph.

Saving Auto-Uploaded Data

Recorded data which has been uploaded to the 

T&D WebStorage Service can be saved in the PC by using 

T&D Graph and accessing T&D WebStorage Service.
For details, please see 

 printed on the back of this paper.

 Changing the Settings

By using the T&D WebStorage Service, it is possible to change the 

following settings:

• Auto-upload Interval

• Warning Settings

The changes in settings will be applied when the device communicates with the 
T&D WebStorage Service. To apply the changes immediately, press the <REC/STOP> 
button on the device after settings are completed.

Thermo Recorder  

TR-7wf Series Introductory Manual

What can TR-7wf Series devices do?

Auto-upload of Recorded Data

*1

Wireless LAN  
Communication

View with a Web Browser

Setup and Data Download

Setup and Data Download

Graph Display

Graph Display

Setup

USB  
Communication

*2

Smartphone
Tablet device

PC

TR-7wf Series

TR-7wf Series

TR-7wf Series

Graph Display 
Download Data

Wireless  
Communication

*3

(Wireless LAN)

Apply Settings

Via Cloud

Via USB

Via Wireless

*1:  To auto upload recorded data to the T&D WebStorage Service, it is necessary to have Internet access via wireless LAN and a T&D WebStorage Service User ID.

*2:  A USB communication cable (sold separately) is required for connecting the TR-7wf series device to your computer.

*3:  A mobile device with wireless LAN capability is required for direct communication with the TR-7wf series device.

 Setting up the Device

11 

Connect sensors to the 
device.

To ensure a proper connection, make sure that the sensor connector is completely 

inserted.

21 

Remove the battery cover and 
insert batteries.

*  Make sure not to m / -.
*  Always use new batteries.

31 

The measurement will be displayed on the LCD screen.

Under Normal Conditions:

During Sensor Error (no sensor connected / 

improper connection / damaged):

 Starting Recording

• Upon the start of a new recording session all stored data will be deleted from the device.
• Recording Interval settings cannot be changed while a recording session is in progress. 

Stop recording first and then change settings.

• During recording, the power cannot be turned off.

11 

Click the <INTERVAL> button 
to display the current record-
ing interval.

21 

Press-and-hold the <INTERVAL> button until the display flashes.

31 

With each pressing of the <INTERVAL> button the recording 
interval time will change. When the desired interval appears, stop 
pressing the button.

41 

Start recording by pressing 
and holding the <REC/STOP> 
button until the 

 mark 

appears in the display.

Below are some examples of recording intervals and maximum recording times1

1 sec.

30 sec.

5 min.

15 min.

60 min.

About 2 hours

About 2 days

About 27 days

About 83 days

About 333 days

In order to stop a recording session, press and hold the <REC/STOP> button until the 

 mark disappears from the display.

Details about the LCD Display

Other than the measurement display, the following is displayed on 

the LCD monitor.

Shows the status of wireless LAN connection.

ON: Connected to the wireless network. (Signal strength: 1 to 

3 bars)

BLINKING : Unable to connect to the wireless network.

OFF : Settings have not been made for the wireless LAN 

connection.

Shows the status of the Internet connection.

ON : Connected to the Internet

BLINKING : Unable to connect to the Internet

OFF : Auto-upload is set to “OFF”.

 

Shows the channel number of the measurement being displayed.

This appears when direct communication with mobile devices is possible.

This appears when the wireless LAN setup using WPS is possible.

This appears when the device is in wireless LAN communication or USB communication.
Shows the recording status.

ON : Recording in progress   BLINKING : Waiting for programmed start

OFF : Recording stopped
Shows the recording mode, which can be changed via TR-7wf for Windows or 

ThermoWEB.

ENDLESS : Upon reaching the logging capacity of 8,000 readings, the oldest data will be 

overwritten and recording will continue.

ONETIME : Upon reaching the logging capacity of 8,000 readings, [FULL] will appear on 

the LCD display and recording will automatically stop.
The amount of recorded data in the device is shown here.

This appears when the battery level is low. Please change the batteries as soon as you see 

this mark. If the batteries are not changed but remain in use, the device will enter sleep 

mode ([SLP] will appear on LCD) and stop all operations. If the batteries are further left 

unchanged, all previously recorded data will be lost.
This appears while the auto-upload interval is displayed by using the <INTERVAL> button.
This appears while the recording interval is displayed by using the <INTERVAL> button.

Common Operations  

Operations Via Cloud

© Copyright  T&D Corporation. All rights reserved.   
2013. 09. 16504930004 1st Edition

http://www.tandd.com/

Distributed by MicroDAQ.com, Ltd 

www.MicroDAQ.com 

(603) 746-5524

Отзывы: