6
Performance Checks
Test Procedures
6-17
Step Advance
Equipment: Oscilloscope, function generator
Preparation:
1. Configure the Oscilloscope as follows:
Termination:
20 dB, 50
Ω
feedthrough attenuator at the
oscilloscope input
Setup:
As required for the test
2. Connect 2074 Channel 1 output to the oscilloscope input
3. Configure the function generator as follows:
Frequency 10
kHz
Run Mode:
Triggered
Waveform:
TTL level square
4. Connect the function generator output to the 2074 TRIG IN
connector
5. Connect 2074 Channel 1 to the Oscilloscope input
6. Configure the 2074 as follows:
SCLK 200
MS/s
Waveform: Sequence
Seq Advance:
Step
Amplitude: 2
V
Trigger Source: External
Output: On
7. Using ArbConnection prepare and download the following
waveform to all channels:
Segment 1:
Sine, 1000 points
Segment 2:
Triangle, 1000 points
Segment 3:
Square, 1000 points
Segment 4:
Sinc, 1000 points
Segment 5:
Gaussian Pulse, 1000 points
8. Using ArbConnection, build and download the following
sequence table:
Step 1:
Segment 1, loop 1
Step 2:
Segment 2, loop 1
Step 3:
Segment 3, loop 1
Step 4:
Segment 4, loop 1
Step 5:
Segment 5, loop 1
Test Procedure
1. Press the manual trigger button on the function generator and
observe that the waveforms advance through the sequence
table repeatedly
Test Results
Pass
Fail
2. Remove the cable from 2074 channel 1
3. Repeat the test procedure as above for channel 2, 3 and 4
Содержание 5064
Страница 26: ...1 5 Figure 1 2 ArbConnection The Control Panels Figure 1 3 ArbConnection The Wave Composer...
Страница 71: ...2074 User Manual 2 18 This page intentionally left blank...
Страница 145: ...2074 User Manual 3 74 Figure 3 48 Reading the 2074 Internal Temperature...
Страница 207: ...2074 User Manual 4 62 Figure 4 47 Complete Pulse Train Design Figure 4 48 Section 5 of the Pulse Train Design...
Страница 220: ...4 ArbConnection Logging SCPI Commands 4 75 Figure 4 58 Log File Example...
Страница 221: ...2074 User Manual 4 76 This page was intentionally left blank...
Страница 333: ...2074 User Manual 7 2...
Страница 358: ...7 Adjustments and Firmware Update Updating 2074 Firmware 7 27 Figure 7 9 Firmware Update Completed...
Страница 359: ...2074 User Manual 7 28 This page was intentionally left blank...
Страница 360: ...Appendices Appendix Title Page A Specifications A 1 1...
Страница 361: ...2074 User Manual 2...