2074
User Manual
6-16
Sequence
operation
This tests the operation of the sequence generators. Each channel
has its own sequence generator and therefore operation is tested on
each channel separately. This also checks the sequence advance
options.
Automatic Advance
Equipment: Counter
Preparation:
1. Configure the Counter as follows:
Function: TOTB
Measurement
2. Connect the counter channel B to the 2074 output
3. Configure the 2074 as follows (all channels):
SCLK 200
MS/s
Waveform: Sequence
Run Mode:
Trigger
Amplitude: 2
V
Output: On
4. Using ArbConnection prepare and download the following
waveform to all channels:
Segments:
1 to 5
Wavelength: 128
points
Waveform:
1 cycle square
Trigger Source: BUS
5. Using ArbConnection, build and download the following
sequence table:
Step 1:
Segment 1, loop 100,000
Step 2:
Segment 2, loop 100,000
Step 3:
Segment 3, loop 100,000
Step 4:
Segment 4, loop 100,000
Step 5:
Segment 5, loop 100,000
Test Procedure
1. From ArbConnection, click on the Manual trigger button and
observe that counter reading is 500,000 counts. Reset counter
and repeat the test a few times. Every time the counter reading
should be 500,000 counts exactly
Test Results
Pass
Fail
2. Remove the cable from 2074 channel 1
3. Repeat the test procedure as above for channels 2, 3 and 4
Test Results
Pass
Fail
Содержание 5064
Страница 26: ...1 5 Figure 1 2 ArbConnection The Control Panels Figure 1 3 ArbConnection The Wave Composer...
Страница 71: ...2074 User Manual 2 18 This page intentionally left blank...
Страница 145: ...2074 User Manual 3 74 Figure 3 48 Reading the 2074 Internal Temperature...
Страница 207: ...2074 User Manual 4 62 Figure 4 47 Complete Pulse Train Design Figure 4 48 Section 5 of the Pulse Train Design...
Страница 220: ...4 ArbConnection Logging SCPI Commands 4 75 Figure 4 58 Log File Example...
Страница 221: ...2074 User Manual 4 76 This page was intentionally left blank...
Страница 333: ...2074 User Manual 7 2...
Страница 358: ...7 Adjustments and Firmware Update Updating 2074 Firmware 7 27 Figure 7 9 Firmware Update Completed...
Страница 359: ...2074 User Manual 7 28 This page was intentionally left blank...
Страница 360: ...Appendices Appendix Title Page A Specifications A 1 1...
Страница 361: ...2074 User Manual 2...