DTC-300 Getting Started Guide
Page 52
7
Carefully lift the bottom stack off of the securing posts.
NOTE
: The thermocouples are secured to the
bottom plate,
DO NOT
lift the bottom stack more than the thermocouples will allow.
CAUTION: Damage can occur if care is not taken when lifting the bottom stack.
MISE EN GARDE: Des dégâts peuvent survenir si vous ne faites pas attention lors du levage de la
cheminée inférieure.
8
Using a hex key, loosen the stack spacer, then remove it from under the bottom stack.
9
Temporarily place the bottom stack back on the securing posts.
Installing the Stack Spacer
1
Clean both the top and bottom stack surfaces with a lint-free cloth and a small amount of isopropyl alcohol.
2
Lift the bottom stack off of the securing posts and insert the stack spacer using a hex key.
3
Place the bottom stack back onto the securing posts. Make sure that all of the bottom stack components
are situated correctly.
4
Place a clean sample onto the bottom stack surface or simply put the two stack faces together.
5
Lower the top stack to apply pressure to the bottom stack. Make sure that the bottom stack is correctly
aligned and proper contact between all surfaces is maintained. If a problem is found, release the pressure
to the bottom stack.
6
Replace the two retaining screws directly under the bottom stack.
Figure 29
Replace screws.
7
Raise the top stack and remove the sample.
Stack Spacer