4. Motor Errors
Error messages:
Exchange Waste Tank
Status:
Not Ready
Probable Cause
1. Waste container is full.
Corrective Action
1. Replace the waste container. (Refer to Chapter 9: 9.12
Replacing the waste container (option).)
Error messages:
WB Asp Motor Error
Status:
Not Ready
Category:
Analysis Error
Probable Cause
1. The load on the whole blood aspiration pump is abnormally
high.
Corrective Action
1. Check the WB aspiration pump:
Check that no fluid switch connector or tubing is touching
the upper/lower part of WB aspiration pump. Then press the
HELP key on the panel keypad of the Main Unit. Select
“OK” on the function menu to check the operation.
Error messages:
RBC Sheath Motor Error
Status:
Not Ready
Category:
Analysis Error
Probable Cause
1. The load on the RBC sheath injector piston is abnormally
high.
Corrective Action
1. Check the RBC seath injector piston:
Check that no fluid switch connector or tubing is touching
the upper/lower part of RBC sheath injector piston. Then
press the HELP key on the panel keypad of the Main Unit.
Select “OK” on the function menu to check the operation.
Error messages:
FCM Sheath Motor Error
Status:
Not Ready
Category:
Analysis Error
Probable Cause
1. The load on the FCM sheath injector piston is abnormally
high.
Corrective Action
1. Check the FCM seath injector piston:
Check that no fluid switch connector or tubing is touching
the upper/lower part of FCM sheath injector piston. Then
press the HELP key on the panel keypad of the Main Unit.
Select “OK” on the function menu to check the operation.
Error messages:
Rinse Motor Error
Status:
Not Ready
Category:
Analysis Error
Probable Cause
1. The load on the rinse cup driving part is abnormally high.
Corrective Action
1. Check the rinse cup:
Check that no fluid switch connector or tubing is touching
the upper/lower part of the rinse cup. Then press the HELP
key on the panel keypad of the Main Unit. Select “OK” on
the function menu to check the operation.
Содержание XE-5000
Страница 2: ......
Страница 42: ...CHAPTER 3 Design and Function 3 18 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Страница 54: ...CHAPTER 5 Before Using 5 4 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Страница 96: ...CHAPTER 6 Operation 6 42 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Страница 128: ...CHAPTER 8 Calibration 8 10 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Страница 256: ...CHAPTER 11 Technical Information 11 38 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Страница 258: ...CHAPTER 12 Warranty 12 2 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Страница 262: ...CHAPTER 13 Index Revised April 2007 13 4 Sysmex XE 5000 Instructions for Use Blank page ...