CHAPTER 10
Troubleshooting
10-12
Sysmex XE-5000 Instructions for Use
R
e
vi
se
d A
p
ri
l 20
07
Error messages:
RBC Detector Temp High
RBC Detector Temp Low
Category:
Analysis Error
Probable Cause
1. Temperature of sample in RBC detector is not within 10 to
40°C.
Corrective Action
1. Check the ambient temperature:
Use sample at an ambient temperature of 15 to 30ºC
(Optimum: 25°C).
Error messages:
IMI Detector Temp High
IMI Detector Temp Low
Status:
Not Ready
Category:
Analysis Error
Probable Cause
1. Temperature of IMI detector is outside the regulation range.
Corrective Action
1. Wait until the temperature has stabilized inside the
regulation range:
If this error is still displayed 30 minutes after the power is
turned ON, there is probably something defective in the
system. Contact your Sysmex technical representative. In
this case, CBC and RET can be analyzed even if the error
persists.
Error messages:
IMI Detector Temp High HPC
IMI Detector Temp Low HPC
Status:
Not Ready
Category:
Analysis Error
Probable Cause
1. Temperature of IMI detector is outside the regulation range
for HPC analysis mode.
Corrective Action
1. Wait until the temperature has stabilized inside the
regulation range:
If this error is still displayed 30 minutes after switching to
HPC analysis mode, there is probably something defective
in the system. Contact your Sysmex technical
representative. In this case, it is possible to perform
analysis in normal mode even if the error persists.
Error messages:
40°C RH Temp High
40°C RH Temp Low
Status:
Not Ready
Category:
Analysis Error
Probable Cause
1. Temperature of 40°C reagent heater is outside the
regulation range.
Corrective Action
1. Wait until the temperature has stabilized inside the
regulation range:
If this error is still displayed 30 minutes after the power is
turned ON, there is probably something defective in the
system. Contact your Sysmex technical representative.
Error messages:
33°C RH Temp High
33°C RH Temp Low
Status:
Not Ready
Category:
Analysis Error
Probable Cause
1. Temperature of 33°C reagent heater is outside the
regulation range.
Corrective Action
1. Wait until the temperature has stabilized inside the
regulation range:
If this error is still displayed 30 minutes after the power is
turned ON, there is probably something defective in the
system. Contact your Sysmex technical representative.
Содержание XE-5000
Страница 2: ......
Страница 42: ...CHAPTER 3 Design and Function 3 18 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Страница 54: ...CHAPTER 5 Before Using 5 4 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Страница 96: ...CHAPTER 6 Operation 6 42 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Страница 128: ...CHAPTER 8 Calibration 8 10 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Страница 256: ...CHAPTER 11 Technical Information 11 38 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Страница 258: ...CHAPTER 12 Warranty 12 2 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Страница 262: ...CHAPTER 13 Index Revised April 2007 13 4 Sysmex XE 5000 Instructions for Use Blank page ...