CHAPTER 10
Troubleshooting
Sysmex XE-5000 Instructions for Use
10-11
R
e
vi
se
d A
p
ri
l 20
07
2. Temperature Errors
Error messages:
-0.07 MPa ERROR
Status:
Not Ready
Category:
Analysis Error
Probable Cause
1. Inadequate vacuum in the Pneumatic Unit
2. Pressure leakage from tube or nipple
Corrective Action
1. The Pneumatic Unit is probably defective. Contact your
Sysmex technical representative.
2. Check the tube connection and nipple looseness.
Error messages:
-0.04 MPa ERROR
Status:
Not Ready
Category:
Analysis Error
Probable Cause
1. -0.04 MPa vacuum adjustment error
2. There is fluid in the Pneumatic Unit trap chamber.
3. Pressure leakage from tube or nipple
Corrective Action
1. Adjust vacuum to -0.04 MPa.
(Refer to Chapter 9: 9.21 Adjusting pressure and vacuum.)
2. Drain fluid from the Pneumatic Unit trap chamber.
(Refer to Chapter 9: 9.2 Removing fluid from the Pneumatic
Unit trap chamber.)
3. Check the tube connection and nipple looseness.
Error messages:
Pressure Lower Error
Status:
Emergency Stop
Probable Cause
1. Pneumatic Unit power is OFF during the operation.
2. Pressure connection tube is disconnected.
Corrective Action
1. Check the Pneumatic Unit power switch and the power cord
connection.
2. Check the pressure connection tube.
Error messages:
Close RBC Detect Cover
Status:
Not Ready
Probable Cause
1. RBC detector cover is open.
2. The cover sensor of RBC detector has failed.
Corrective Action
1. Close the RBC detector cover. If the error persists, the
cover sensor is probably defective.
2. Replace the cover sensor of RBC detector:
Contact your Sysmex technical representative.
Error messages:
IMI Detector Cover Open
Status:
Not Ready
Probable Cause
1. IMI detector cover is open.
2. The cover sensor of IMI detector has failed.
Corrective Action
1. Close the IMI detector cover. If the error persists, the cover
sensor is probably defective.
2. Replace the cover sensor of IMI detector:
Contact your Sysmex technical representative.
Содержание XE-5000
Страница 2: ......
Страница 42: ...CHAPTER 3 Design and Function 3 18 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Страница 54: ...CHAPTER 5 Before Using 5 4 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Страница 96: ...CHAPTER 6 Operation 6 42 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Страница 128: ...CHAPTER 8 Calibration 8 10 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Страница 256: ...CHAPTER 11 Technical Information 11 38 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Страница 258: ...CHAPTER 12 Warranty 12 2 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Страница 262: ...CHAPTER 13 Index Revised April 2007 13 4 Sysmex XE 5000 Instructions for Use Blank page ...