31 DePuy Synthes Instructions for Use Universal Battery Charger II
Guidelines and manufacturer’s declaration – electromagnetic immunity
TheSynthesUniversalBatteryChargerIIisintendedforuseintheelectromagneticenvironmentspecifiedbelow.
The customer or the user of the Synthes Universal Battery Charger II should assure that it is used in such an envi-
ronment.
Immunity test
standard
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short interrup-
tions and voltage variations
on power supply lines
IEC 61000-4-11
Note:
U
T
is the A.C. mains voltage prior to application of the test level.
Power frequency (50/60 Hz)
magneticfield
IEC 61000-4-8
IEC 60601
test level
±8 kV contact
±15 kV air
±2 kV for power
supply lines
±1 kV line to line
±2 kV line to
earth
<5 % U
T
(0.5 cycle)
40 % U
T
(5 cycles)
70 % U
T
(25 cycles)
<5 % U
T
for 5 s
30 A/m
Electromagnetic environment – guidance
Floors should be wood, concrete or ceramic
tile.Ifthefloorsarecoveredwithsynthetic
material, the relative humidity should be at
least 30%.
Mains power quality should be that of a
typical commercial or hospital environment.
Mains power quality should be that of a
typical commercial or hospital environment.
Mains power quality should be that of a
typical commercial or hospital environment.
If the user of the Synthes Universal Batter
Charger II requires continued operation during
power mains inter ruptions, it is recommended
that the Synthes Universal Battery Charger II is
powered from an UPS.
Powerfrequencymagneticfieldsshouldbe
at levels characteristic of a typical location in a
typical commercial or hospital environment.
Table 2: Immunity (all devices)
Compliance
level
±8 kV contact
±15 kV air
±2 kV for power
supply lines
±1 kV line to line
±2 kV line to
earth
<5 % U
T
(0.5 cycle)
40 % U
T
(5 cycles)
70 % U
T
(25 cycles)
<5 % U
T
for 5 s
200 A/m
Technical Data