24 DePuy Synthes Instructions for Use Universal Battery Charger II
Battery/power module performance
is low.
Battery/power module is visibly
damaged.
UBC II is visibly damaged.
Problem
Possible cause
Solution
Insufficientbattery/powermodule
status.
Expected battery/power module life
is reached.
Battery/power module is not ready
for use.
Power Tool or attachment is slug-
gish,i.e.asaresultofinsufficient
maintenance.
Battery/power module was exposed
to excessive heat.
Battery/power module was washed,
disinfected or sterilized.
Battery/power module was short-
fused by metal objects.
Battery/power module fell on the
floor.
UBC II was exposed to excessive
heat.
UBC II was washed, disinfected or
sterilized.
UBC II was short-fused by metal
objects.
UBCIIfellonthefloor.
Refresh battery (see page 11 ff). Only
possible with Battery Power Line
battery (530.620) and Colibri/SBD
batteries (532.003, 532.033).
Test battery/power module (see page
11 ff). If the red display lights up,
replace battery/power module.
Charge battery/power module until
symbol
is green.
Send Power Tool and attachments to
DePuy Synthes representative to be
checked.
Replace battery/power module.
Replace battery/power module.
Replace battery/power module.
Replace battery/power module.
Replace UBC II.
Replace UBC II.
Replace UBC II.
Replace UBC II.
Please also observe the instructions for use
of the corresponding Power Tools.
If the recommended solutions do not work,
please contact your DePuy Synthes representative.
Troubleshooting