Universal Battery Charger II Instructions for Use DePuy Synthes 23
The symbol is red when the char-
ger is switched on before the batter-
ies/power modules are inserted.
Symbol
flashesyellowduringthe
charging process.
Symbol does not light up yellow
when the button is pressed.
It is not possible to insert the
battery/power module into the slot.
Charger makes loud noises.
Problem
Possible cause
Solution
Charger bay is faulty.
Battery/power module is too hot.
Button was released too soon.
Charger bay is faulty.
Charger has an error.
Wrong slot.
Non-authorized battery/
power module.
Contacts in the slot bent.
Ventilation holes on the sides, back
or base are covered and/or the de-
vice is next to a heat source. Auto-
matic cooling is on full power.
Use another free charger bay. Send
the charger to the DePuy Synthes
affiliateforrepairsassoonaspossible.
Leave battery/power module in-
serted in the charger bay. Charger
automatically continues the charging
process once the battery/power
module has cooled down.
Hold button down for at least
2 seconds.
Select another free charger bay.
Send the device to the DePuy Synthes
represen tative for repairs as soon as
possible.
Switch off charger, then switch back
on after 5 seconds. If the ON/OFF
blueLEDflashes,sendthedeviceto
the DePuy Synthes representative for
repairs.
Select correct slot and re-insert
battery/power module.
Check battery/power module type.
Use another free charger bay. Send
the charger to the DePuy Synthes
representative for repairs as soon as
possible.
Expose ventilation holes and/or en-
sure that the device is not next to a
heat source.