background image

21    DePuy Synthes  Instructions for Use  Universal Battery Charger II

The device should be sent to the DePuy Synthes 
 representative for repair if it is faulty or malfunctions.

The same applies if the ON/OFF blue LED does not light 
up­or­flashes­when­the­device­is­switched­on.

To prevent damage of the charger during shipping use 
the original packaging to return devices back to  
DePuy Synthes. If this is not possible, please contact the  
DePuy Synthes representative. 

When shipping Synthes batteries in aircraft carrier bays, 
please follow the instructions in the section 30% State 
of Charge of this IFU as well as the IFU of the device 
used. 

Precautions:

 

• 

The manufacturer shall assume no responsibility 
for damage resulting from unautho rized repair. 

• 

Users or third parties should not carry out repairs 
them selves. 

Care and Maintenance 

Repair and Technical Service

Содержание Universal Battery Charger II

Страница 1: ...Universal Battery Charger II Instructions for Use...

Страница 2: ......

Страница 3: ...Charging the battery 10 Temperature monitoring 11 Charging new batteries or batteries not recently used 11 Errors during charging 11 Checking and Refreshing Batteries 12 Battery Power Line and Colibri...

Страница 4: ...28 Electromagnetic Compatibility 29 Permissible Battery Types 33 Battery for Battery Power Line 33 Battery for Battery Power Line II 33 Power module for Trauma Recon System 33 Batteries for Colibri Sm...

Страница 5: ...s power modules Introduction General Information System Handpieces Battery Power Module Battery Power Line 530 605 530 620 14 4 V NiMH 530 610 530 615 Battery Power Line II 530 705 530 630 14 8 V Li I...

Страница 6: ...devices rather than the instru ments Specific adverse events side effects for the implants can be found in the respective Synthes implant instructions for use Patient Target Group There are no restri...

Страница 7: ...t the device for damage and wear e g unrecognizable markings missing or removed part numbers corrosion etc Do not use any component if dam age is apparent Precautions The device may only be used with...

Страница 8: ...The same environmental conditions apply for transport as for storage Warnings The device is designed to be operated and stored in closed rooms Do not use the device in the direct vicinity of radiator...

Страница 9: ...tery type 3 ON OFF blue LED 4 Control symbols for each charging bay 5 Ventilation holes Description of the Device Rear view 6 Ventilation holes 7 Power switch 8 Fuses 225 AT 250 V 9 Power cord connect...

Страница 10: ...must be sent in for control If the symbol for a single charger bay is red fig 3 before the battery is inserted this charger bay is faulty In this case batteries power modules can still be charged in...

Страница 11: ...r bay The device is fitted with four independent charger bays Each of these has three slots for the following batteries fig 1 1 Battery Power Line and Battery Power Line II batteries 530 620 530 630 2...

Страница 12: ...any objects other than authorized Synthes batteries into the charger as the contacts may otherwise be damaged Precaution If the symbol does not light up after the battery is inserted remove the batte...

Страница 13: ...Colibri SBD batter ies 532 003 532 033 that have not been used for a long period of time and that have not been stored in an acti vated charger do not reach their maximum performance for several char...

Страница 14: ...aximum performance again Old battery The battery performance declines with age and usage With the check function it can be seen if the battery performance is sufficient In some cases batteries that ar...

Страница 15: ...ocess refreshing and checking battery status takes around 10 hours and should only be carried out if there is enough time to do so A battery can be charged checked or refreshed inde pendently in each...

Страница 16: ...ion is started manually as outlined below After the battery is inserted the symbol lights up yellow To check the battery press the button with the exclamation mark for at least 2 seconds fig 5 until t...

Страница 17: ...posed of The entire process checking battery status takes around 3 hours and should only be carried out if there is enough time to do so A battery can be charged or checked independently in each charg...

Страница 18: ...hin the next 3 charging cycles This is done by pressing the ex clamation mark button for at least 2 seconds fig 10 The symbol light goes out and the symbol stops flashing and stays yellow fig 11 If th...

Страница 19: ...sary to store Trauma Recon System power module 05 001 202 Battery Power Line II battery 530 630 and Colibri II SBD II battery 532 103 in the charger Once charged they can be stored outside the charger...

Страница 20: ...r times 4x Fig 2 4 When the yellow symbols and light up the UBC II is performing the 30 SOC function This will take between 30 and 90 minutes depending on the charge status and type of battery Fig 3 5...

Страница 21: ...based disinfectant each time Please follow the safety instructions given by the manual of the disinfec tant manufacturer which is selected Whenever it is cleaned the device should be checked to ensure...

Страница 22: ...he original packaging to return devices back to DePuy Synthes If this is not possible please contact the DePuy Synthes representative When shipping Synthes batteries in aircraft carrier bays please fo...

Страница 23: ...d of in accordance with the national application of the respective directive Faulty batteries may not be reused and should be dis posed of in an environmentally friendly manner and in accordance with...

Страница 24: ...arger bay and dispose of if necessary Send the charger to the DePuy Synthes representative for repairs Replace battery power module Check firmware version on the sticker which is visible on the under...

Страница 25: ...rce Auto matic cooling is on full power Use another free charger bay Send the charger to the DePuy Synthes affiliate for repairs as soon as possible Leave battery power module in serted in the charger...

Страница 26: ...excessive heat UBC II was washed disinfected or sterilized UBC II was short fused by metal objects UBC II fell on the floor Refresh battery see page 11 ff Only possible with Battery Power Line batter...

Страница 27: ...W Protection class I EN IEC 60601 1 Protection type by casing IPX0 EN IEC 60601 1 Fuses 225 AT 250 V Operating mode Continuous operating mode Sterilization Device must not be sterilized Performance Ch...

Страница 28: ...03 14 4 V Li Ion 11 0 sticker visible on the underside of the charger Battery Power Line II 530 630 14 8 V Li Ion 14 0 sticker visible on the underside of the charger 30 State of Charge 532 103 14 4 V...

Страница 29: ...ge Temperature 40 C 104 F 10 C 50 F 40 C 104 F 10 C 50 F Relative humidity 90 30 90 30 Atmospheric pressure 1060 hPa 500 hPa 1060 hPa 500 hPa Altitude 0 5000 m 0 5000 m Transportation Temperature Dura...

Страница 30: ...ANSI AAMI ES60601 1 2005 R 2012 CAN CSA C22 2 NO 60601 1 14 Medical electrical equipment Part 1 2 Collateral Standard Electromagnetic disturbances Requirements and tests IEC 60601 1 2 2014 Ed 4 0 EN...

Страница 31: ...uctuations flicker emissions IEC 61000 3 3 Compliance Group 1 Class B Class A Complies Electromagnetic environment guidance The Synthes Universal Battery Charger II uses RF energy only for its interna...

Страница 32: ...les 70 UT 25 cycles 5 UT for 5 s 30 A m Electromagnetic environment guidance Floors should be wood concrete or ceramic tile If the floors are covered with synthetic material the relative humidity shou...

Страница 33: ...lowing symbol Note 1 At 80 MHz and 800 MHz the higher frequency range applies Note 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflecti...

Страница 34: ...itter W 0 01 0 1 1 10 100 For transmitters rated at a maximum output power not listed above the recommended separation distance d in metres m can be estimated using the equation applicable to the freq...

Страница 35: ...apacity 2 Ah 28 8 Wh Battery NiMH Typical charging time 60 min Power module for Trauma Recon System Art no 05 001 202 Operating voltage rated 25 2 V Battery capacity 1 2 Ah 30 24 Wh Battery Li Ion Typ...

Страница 36: ...out and not longer available Subject to technical modifications Battery for Colibri II Small Battery Drive II Art no 532 103 Operating voltage rated 14 4 V Battery capacity 1 2 Ah 17 28 Wh Battery Li...

Страница 37: ...and to check Battery Power Line II Colibri II SBD II batteries or power modules Button to start the 30 state of charge feature Symbol is yellow the process to check and refresh batteries and to check...

Страница 38: ...tection require ments Please observe national and local regulations See section entitled Disposal Legal manufacturer Manufacturing date Product is UL Classified to the requirements of both the United...

Страница 39: ...ord three pole Switzerland 05 001 142 Power Cord three pole India South Africa 05 001 143 Power Cord three pole Italy 05 001 144 Power Cord three pole China 05 001 145 Power Cord three pole Japan 05 0...

Страница 40: ...nformation This publication is not intended for distribution in the USA For full user instructions warnings and precautions please consult the IFU of any additional devices used All Synthes implant in...

Отзывы: