12 DePuy Synthes Instructions for Use Universal Battery Charger II
Operating the Device
Checking and Refreshing Batteries
1. Battery Power Line and Colibri/SBD batteries
The charger enables Battery Power Line battery
(530.620) and Colibri/SBD batteries (532.003, 532.033)
to be refreshed and checked.
Itwillbeindicatedifthebatteryperformanceissuffi-
cient or if the battery needs to be replaced.
The following factors will affect the battery perfor-
mance:
•
Unused, new battery
•
Battery not used for a longer period
In these cases a battery will only reach its maximum
performance after several charging and discharging
cycles. The check and refresh function ensures that
the battery regains its maximum performance again.
•
Old battery
The battery performance declines with age and usage.
With the check function it can be seen if the battery
performanceissufficient.Insomecases,batteriesthat
are too old can no longer be refreshed.
When required, the check and refresh function is started
manually, as outlined below.
After the battery is inserted, the symbol
lights up
yellow. To refresh and check the battery, press the but-
ton with the exclamation mark for at least 2 seconds
(fig.1)untilthesymbol lightsupyellow(fig.2).The
device then carries out the process. The symbol is
yellow throughout this time.
Precautions:
•
To recharge batteries normally, do not press the
exclamation mark button .
•
Do not remove the battery from the charger bay as
long as the symbol is yellow. Wait until the pro-
cess has ended and the symbols
or light up.
Only then is the battery status clearly assessed.