Micro Dissection Attachment
MDA Micro Dissection Attachment
Note:
Instructions are written such that distal end of hand-
piece and attachment are pointed away from user.
1. Hold MDA by drive shaft. Insert dissection tool into MDA
attachment until dissection tool seats in attachment.
(Drops slightly proximally and stops rotating.) This is best
accomplished by pushing slightly while rotating.
2. Pull dissection tool to ensure proper engagement.
3. Move handpiece knurled knob to “RUN” (
) position.
4. Slide attachment over distal end of handpiece. Pull attach-
ment towards handpiece, and rotate attachment to right
approximately one-quarter turn until fully seated. Hand-
piece knurled knob moves up to seat next to attachment.
5. Attachment and dissection tool are secure.
Note:
Handpiece is now fully functional.
Disassembly
1. Move handpiece knurled knob to “SAFE” position.
2. Slowly rotate dissection tool in alternating directions while
removing distally.
3. Rotate attachment to left approximately one-quarter turn
and remove from distal end of handpiece.
Warning:
The Micro Dissection Attachment is to be used
only for bone dissection applications. Excessive force or side
loads during use can cause rapid temperature increase at the
distal end of the attachment. Always use copious irrigation
with this attachment.