Creating a new batch
Note
When you carry out a test measurement for a Spiral Plate batch in the Batch Designer, the
dilution selected in the dilutions menu affects the results, which show a count/ml value – see
Dilutions file format
Each line of the file should consist of a string representing the dilution (if the string is not
recognized or no dilution is given, ‘No Dilution’ is used).
The order of the lines in the file defines the order of the dilutions in the list.
For example:
No Dilution
1:5
1 in 1000
1 in 10^6
Note
aCOLyte3 does not
require
you to use unique dilutions.
Setting count restrictions
Note
The Count Restrictions button is hidden for Spiral Plate batches.
To place an upper limit on the count result:
1. Press the Measure tab
to display the measure controls, including the Count Restrictions button.
2. Press the Count Restrictions button to display the Count Restrictions controls:
3. Check Limit count per frame to enable count restrictions.
4. Either type the required limits directly into the Upper Count and Lower Count edit
boxes or press the arrow buttons to increase or decrease the values.
52
aCOLyte3
User Manual