VelaShape III User Manual Maintenance
PB81499EN
71
7. M
AINTENANCE
While the VelaShape III system should not be serviced by anyone other than an
authorized Syneron representative, the system requires regular preventive
maintenance and cleaning to ensure optimal system performance. This chapter
describes the routine maintenance procedures that operators may perform:
●
Cleaning the VelaShape III system.
●
Cleaning the VelaShape III applicator and its replaceable covers.
●
Cleaning the air filter.
●
Replacing the applicator's replaceable covers and filters when required
Note
Each patient should be assigned with
his/her own set of covers
in
order to provide safe and hygienic treatments. This also allows the
patient to keep track of the number of treatment hours administered.
Warning
The
symbol located near the power cable's connection port
indicates:
WARNING! DISCONNECT POWER SUPPLY CABLE BEFORE
MAINTENANCE OR SERVICE PROCEDURES TO AVOID
ELECTRIC SHOCK.
For continued protection against fire, replace the fuses only with ones
of the same type and rating.
Cleaning the Console
Clean the system at least once a week. Turn the system off and wipe all surfaces
with a soft, damp, non-abrasive cloth.
Clean the umbilical cable with pre-moistened wipes.
Warning
Be careful not to spill any liquids on the system.