5120A/5115A Operations and Maintenance Manual
73
To calibrate the internal reference oscillators:
1. Set the test set's system date and time if necessary, as calibration information is time
stamped.
To check current system time see
on page 21, to set system
time see
on page 22 for details.
2. Detach any signal connected to the Reference port.
3. Attach a calibration lab quality 10 MHz signal to the Input port.
4. Press
Settings and Options
, then
Options
, and finally
Internal Reference
to access the
calibration screen.
Note
This screen can be accessed to retrieve current calibration information regardless
of attached signals or collecting state.
5.
Start
data collection and wait at least 17 minutes for the 1000 second frequency average
used to calculate the internal oscillators' frequencies.
While the internal reference screen is active, the 5120A-01 periodically attempts
calibration data calculation. Calculation results or conditions that prevented the
calculation are reported in the Pending Calibration frame. When the Pending
Calibration frame presents the measured frequencies, the test set is ready to store the
oscillator calibration values (proceed to the next step)
6. Press
Calibrate
to store frequency calibration data.
Calibration frequencies are calculated, and stored. The
Previous Calibration
frame
updates to reflect newly calibrated frequencies and time of calibration.
8.1.1.2 Calibrating 5120A-01 Internal Reference Remotely
To retrieve current internal reference calibration data remotely:
1. Access the Command port. See
“7.1 Accessing the PNTS Remotely and Exiting the
2. Issue the
calinfo
command.
The 5120A-01 returns date and time of calibration and the calibrated frequencies of
both internal reference oscillators.
To calibrate the internal reference oscillators remotely:
1. Detach any signal connected to the Reference port.
2. Attach a calibration lab quality 10 MHz signal to the Input port.
3. Access the Command port. See
“7.1 Accessing the PNTS Remotely and Exiting the
4. Set the test set's system date and time if necessary, as calibration information is time
stamped.
Run the
show date
command to check, and
set date
command to set the system time.
Содержание 5115A
Страница 6: ...iv Contents Index 97 ...
Страница 12: ...6 1 Introduction ...
Страница 16: ...10 2 Setting Up a PNTS ...
Страница 26: ...20 3 The Screen and Basic Concepts ...
Страница 82: ...76 9 Maintaining a PNTS ...
Страница 92: ...86 Appendix A Specifications ...
Страница 98: ...92 Appendix B Theory of Operation ...
Страница 100: ...94 ...
Страница 102: ...96 ...
Страница 106: ...100 ...