4 Commissioning
9
EN
4
Commissioning
Check the aerosol suction as follows before using it/putting it into operation for the first time:
•
Has the suction power of the single place/central suction system been set correctly (min. > 300 l/min)?
•
Check how the headphones work with the treatment unit’s electric scale removal device by looking out for any
possible generation of heat.
•
Has the suction cannula been correctly fitted to the cannula adapter?
•
Has the cannula adapter been correctly connected to the suction hose on the spray mist suction unit?
•
Has the aerosol suction been correctly fastened to the headphones?
•
Are the headphones with aerosol suction correctly seated on the patient's head?
•
Are the protective covers on the headphone cups?
•
What is the level of wear on the suction cannulas? Do not use badly worn suction cannulas.
•
Device-specific kit; see the separate installation instructions for information on the additional aerosol suction.
5
Reprocessing
Reprocessing methods
– in accordance with ISO 17664:2018
Always perform all the reprocessing steps provided in these instructions for use as described and in the order
given.
Process the aerosol suction prior to using it for the first time and reprocess it after treating each patient.
The aerosol suction should be reprocessed as soon as possible following treatment. Also follow the processing
steps in the user manuals provided by the relevant treatment device manufacturers of the suction system.
The manufacturer guarantees effective reprocessing up to 250 cycles in accordance with these instructions for use.
The suction hoses of the various kits for treatment units are to be carried out according to the specifications of the
treatment steps from the operating instructions of the corresponding treatment unit manufacturers.
Reprocessing may only be conducted by qualified personnel.
No special servicing, maintenance, inspection or testing needs to be performed following the reprocessing.
Reprocessing restrictions
The following restrictions apply to the reprocessing of the
headphones
. Failure to observe the following may damage the
headphones or cause malfunctioning:
•
Do not clean inside the headphones
•
Do not use disinfectants inside the headphones
•
Do not perform automatic cleaning
•
Do not use thermal disinfection
•
Do not perform automatic disinfection
•
Do not perform ultrasonic cleaning
•
Do not submerge the headphones in liquid (e.g. cleaning agents/disinfectants)
Only clean the headphones using wipe-down disinfection as described in Chapter 5.3 Manual disinfection.