Failures and remedies 

 

 

Failures and possible causes  

 
 

NOTE:

 

In case of malfunction on the external sensors, contact your  after 

sales service. 

 

 

Failure 

Cause 

Remède 

The displays do not light up 

after switching on. No lights are 

on. 

 

Primary power supply faulty  

 

Check the fuse of the primary 

power supply. 

 

Check the power cable. 

When powering, sensor (s) do 

not measure (s) or displayed 

values are crazy. 

 

Cable sensors offline 

 

Faulty sensor 

 

Check the sensor cable 

 

Check the BNC connector 

 

Change the sensor 

The sensor shows a maximum 

value at all times. 

 

Sensor cable cut 

 

Faulty sensor 

 

Check your connections or 

replace the faulty sensor 

Continuing instability of the 

measurement sensor. 

 

Worn or defective sensor 

 

Presence of air in the filtration 

 

External interference 

 

Change the sensor. 

 

Check the priming of filtration. 

Unable to calibrate the sensor. 

 

Poor electrode or unstable 

measurement 

 

Change the electrode and check 

the priming of filtration. 

Disturbed and unstable control. 

 

Incorrect parameters setting 

 

Check the program settings. 

 

Check the behaviour of the site 

and adjust the control parameters. 

Proportioning devices do not 

assay. 

 

Controller safekeeping 

 

Check the errors displayed. 

 

Check the operating limit 

exceeded. 

Содержание HYDR'ORP

Страница 1: ..._______________ HYDRO pH HYDR ORP HYDRO pH Duo and HYDRO pH ORP instructions HYDRO pH HYDR ORP HYDRO pH Duo and HYDRO pH ORP pH and ORP analysers controllers for private swimming pools Installation co...

Страница 2: ...pH ORP and pH ORP controllers for private swimming pools with DULCOFLEX pumps Operating and programming instructions of 2016 July the 02th Reference DOC0342 Publisher SYCLOPE Electronique S A Z I A r...

Страница 3: ...17 a Mounting procedure of pipe saddles 19 b Mounting procedure for the sensors connection kit 19 c Mounting procedure for pipes connection kit 20 d Mounting procedure of flexible sunction pipe 21 f...

Страница 4: ...ontroller 36 C Changing set points of pH and or ORP controller 37 D Faults reset 37 E Testing the metering pump s 38 F Change the brightness of display and LEDs 38 I Calibration of pH and or ORP with...

Страница 5: ...heir user friendliness and their remarkable technicality you will fully enjoy their many possibilities and will be assured of a perfect control and perfect monitoring of your pool water quality You wi...

Страница 6: ...ON and NOTE have the following meanings NOTE Indicates a special or important information WARNING Risk of injury or accident CAUTION Risk of malfunction DANGER Risk related to presence of power supply...

Страница 7: ...ff who is familiar with basic instructions on industrial safety and accident prevention is trained to use the device has read and understood these instructions warnings and rules for handling The user...

Страница 8: ...as not to exceed the maximum service temperature They must be protected from freezing See section wintering If potential winter these devices must be disassembled and placed in a dry place protected...

Страница 9: ...mm Options Level sensor for tank bottom Connections kit Packaging Standard solution pH pH sensor Product pH 7 sensor mounting kit injection kit Hydro pH controller Tank bottom 5 8mm tubing Box fixing...

Страница 10: ...ackaging Hydr ORP ORP ORP sensor Product controller sensor mounting kit injection kit Tank bottom Oxidizer pump Box fixing Wall screws valve tubing kit kit Codifications HYDR ORP standard controllers...

Страница 11: ...ller Fixation kit Kit of mural with power socket for box screws Codifications HYDR ORP socket controllers Code Reference HYDR ORP Socket electronic box Power supply Measuring sensor Dosing device Prot...

Страница 12: ...ble 1m 3x 0 75 mm2 230V 50 60Hz pH with 6m of cable Pumps 2x0 8l h IP54 HYD 1322 Standard equipment 50 with cable 1m 3x 0 75 mm2 230V 50 60Hz pH with 6m of cable Pumps 2x1 6l h IP54 HYD 1323 Standard...

Страница 13: ...DRO pH ORP electronic box Power supply Measuring sensors Metering devices Protection class HYD 1420 Standard equipment 50 with cable 1m 3x 0 75 mm2 230V 50 60Hz pH ORP cable 6m Pumps 2x 0 4 l h IP54 H...

Страница 14: ...impedance input with free contact of potential b General functions Display Displays measured on pH or ORP in mV in physical quantity Human machine interface 4 keys of functions identified Calibration...

Страница 15: ...General dimensions of boxes and wall mounting dimensions Dimensions and fixing ratings of Hydro pH and Hydr Orp controllers Box dimension Internal fixation dimensions External fixation dimensions Dime...

Страница 16: ...t humidification of measurement tips NOTE The recyclable packaging of HYDRO devices must be eliminated according to the current rules Keep wintering packaging RECYCLING Electronic equipments shall be...

Страница 17: ...rance with the wall 8 Fix the upper screws and tighten the lower screws once in place 9 For a fixation without kit open the top door of the electronic box and tighten the 4 screws starting with the tw...

Страница 18: ...Example with HYDRO pH ORP 3rd case Example with HYDRO pH ORP Chemicals injection pipe rings are placed after the heater Sensors are placed between the filter output and the heater FILTER POOL SKIMMER...

Страница 19: ...performing this assembly to the vertical maintenance is easier to achieve Gasket Screw Nut Apply the pipe saddle either side of the tubing on a straight section of at least 15 cm Tighten the two side...

Страница 20: ...11 Step 12 Step 13 Step 14 Put teflon on the threads of the fitting and injection valve Screw the fitting on the pipe saddle Screw the injection valve on the fitting Teflon band Screw the cap with th...

Страница 21: ...d Mounting procedure of flexible sunction pipe Step 15 Step 16 Step 17 Step 18 Step 19 Step 20 Option Place the bottom tray valve weighted and adjust the depth Place the sensor level and its weight an...

Страница 22: ...the two sensors if necessary f Connecting the pH and or ORP sensors on the Hydro box Step 21 Step 22 Step 23 Hand tighten the nut of the sensor holder The sensor is ready Slide the nut on the sensor a...

Страница 23: ...the side of the device For safety controllers supply must be cut when the filtration is off CAUTION The electrical connection of the box must necessarily be coupled with the operation of the pool fil...

Страница 24: ...ral and a phase and Earth b Position of the pH control direction HYDRO pH and HYDRO pH ORP c Internal connections of the tank bottom detector Jumper pH pH pH pH Before opening the box cut the primary...

Страница 25: ...use replacement must be performed by a qualified technician CAUTION Deteriorated fuses must imperatively be replaced by fuses of identical intensity and technology NOTE If a fuse is destroyed it is im...

Страница 26: ...ey can damage your health and cause environmental damages HYDRO pH HYDR ORP and HYDRO pH ORP are devices of quantification of these products that meet the current standards Any mixture of chemicals ca...

Страница 27: ...performed to enter into specific operations 2 Display and LED Light Emitting Diode Depending on devices 4 or 8 LEDs can be used LED Set Yellow Indicates an operation is being LED Cal Yellow Indicates...

Страница 28: ...services intervention CAUTION In any case this switch can be used for working inside the controller Only the primary power failure allows this intervention 5 4 Safety breaker circuit uses in the HYDR...

Страница 29: ...mode This security is necessary for stabilization of measures before proportioning CAUTION In case of exceeding the operating range of the controller the pump s is are stopped and the warning light tu...

Страница 30: ...5 to 600 minutes per filter cycle or per day if the unit runs all day The Factory adjustment time is 120 minutes NOTE After the maximum time has elapsed the pH dosing pump is stopped and the error E01...

Страница 31: ...proportional bands The hidden mode for adjusting the proportional bands Lin or the On Off control with hysteresis working Tor of pH and ORP is reached by pressing simultaneously and keys when powerin...

Страница 32: ...H of 0 2 the metering pump runs at 100 of its time for a difference of 0 1 pH it runs at 50 of its time The basic operating time of the pumps is 240 seconds or 4 minutes C 2 Programming of the pH and...

Страница 33: ...ortional band Xp of the linear mode in mV The proportional band is validated to 50mV Remark The proportional band can be modified in a functional range of 5mV to 300mV The factory setting is fixed to...

Страница 34: ...n 8 The setting is made in steps of 0 1pH Factory value is 6 8 pH E Changing the maximum terminal of the pH control range The hidden mode for changing the higher or minimum terminal of the pH control...

Страница 35: ...Cal during 5 seconds after powering on 5s 5s Return to start The polarity of CAD input can be adjusted to 0 active contact open or 1 active contact closed Factory value is 0 G Record of the last repo...

Страница 36: ...oller display the pH or ORP simply press successively the keys or This function is also necessary to identify operations of visualization modification of setpoints or calibration This function only co...

Страница 37: ...ersion ensure the controller Display mode Process Changing the pH set point Display of set point and or and NOTE The pH set point is adjustable in steps of 0 05 pH in the lower and upper limits set in...

Страница 38: ...display and LEDs To change the brightness of the display or indicators simply press keys Cal and then or Brightness increases or Brightness decreases Immediately after releasing the key and after a f...

Страница 39: ...version it is necessary to check the Display mode of the controller Automatic calibration of pH CAUTION To perform an automatic calibration of pH filtration must operate for several minutes the pH val...

Страница 40: ...r must not be faulty or disconnected The displayed value of ORP must be between 200mV and 800mV Process Adjust the display value to the value read with your standard reagent Value identical to the sta...

Страница 41: ...he filtration stop and the setting of a cap to replace the sensor to be calibrated NOTE The standard chemical reagents for calibration of pH or ORP and the cap are not supplied with the controller Ste...

Страница 42: ...H of your pool If it does not match perform the full calibration with pH 4 00 Process The value shown is the pH value to be corrected Press again on Cal during 5 seconds until you can see PH7 5s Calib...

Страница 43: ...isplay PH4 5s Calibration in progress Upon completion of calibration the controller resumes normal operation and displays pH 4 00 If calibration error the controller displays NOTE It is possible to ca...

Страница 44: ...ror the controller displays Step 6 Stop filtration and close the isolation gates Step 7 Replace the sensor in its sensor holder Step 8 Open the gates and turn the filtration on Zero ORP Calibration CA...

Страница 45: ...n of calibration the controller resumes normal operation and displays 0 If calibration error the controller displays Remove the short circuit and reconnect the ORP sensor Calibration of ORP 465mV Proc...

Страница 46: ...the display rd 0 then press again on Cal to display rd 5 5s Adjust the value of the display to the standard value of your reactive Calibration in progress Upon completion of calibration the controller...

Страница 47: ...o 750mV Setting of the operating band of the pH controller between 6 8 and 7 8pH Setting to x20 for the pH proportional band before the V2 12 realise Or setting to 0 8pH for the proportional band sinc...

Страница 48: ...surchlorations or deposits of chemicals can affect the operation or destroy sensors CAUTION Flocculation should never be made on direct contact with sensors If flocculation occurs in a skimmer so cont...

Страница 49: ...abnormal use it can be reduced to a few months If these precautions are not taken the sensors will not operate next season CAUTION The metering pumps should also be wintered Run clean water into the...

Страница 50: ...h pumps 2 4l h 230V 50Hz HYD 2423 HYDRO pH DUO single controller with pumps 0 4l h 230V 50Hz HYD 2320 HYDRO pH DUO single controller with pumps 0 8l h 230V 50Hz HYD 2321 HYDRO pH DUO single controller...

Страница 51: ...sensor Check the sensor cable Check the BNC connector Change the sensor The sensor shows a maximum value at all times Sensor cable cut Faulty sensor Check your connections or replace the faulty senso...

Страница 52: ...ror find the root default that make the error message Error code Error meaning Er1 Maximum daily dosing time for pH chanel overpass Er2 Maximum daily dosing time for ORP chanel overpass Er3 Calibratio...

Страница 53: ...Page 53 ________________________________________________________________________________________________ HYDRO pH HYDR ORP HYDRO pH Duo and HYDRO pH ORP instructions...

Страница 54: ...Page 54 ________________________________________________________________________________________________ HYDRO pH HYDR ORP HYDRO pH Duo and HYDRO pH ORP instructions NOTES...

Страница 55: ...Page 55 ________________________________________________________________________________________________ HYDRO pH HYDR ORP HYDRO pH Duo and HYDRO pH ORP instructions...

Страница 56: ...Page 56 ________________________________________________________________________________________________ HYDRO pH HYDR ORP HYDRO pH Duo and HYDRO pH ORP instructions...

Страница 57: ...ad Office Z I A ropole Pyr n es Rue du Bruscos 64 230 SAUVAGNON Tel 33 05 59 33 70 36 Num ris concentrated phone lines Fax 33 05 59 33 70 37 Email service technique syclope fr 2016 by SYCLOPE Electron...

Отзывы: