52
A-96.730.121 / 160920
Chematest 35
Technical Data
13.3. Sensor Specifications
13.3.1 pH and Redox
Swansensor
pH CT
Measuring range:
Resolution:
1–13
0.01
Automatic temperature compensation with integrated sensor.
DIN/NIST or technical buffers.
Swansensor
Redox CT
Pt-redox electrode with integrated temperature sensor for consis
-
tent data records.
Measuring range:
Resolution:
−400–1200 mV
1 mV
Common
characteristics
Poisoning-protected reference systems with solid electrolyte and
robust open junction diaphragm.
Temperature range:
Pressure:
Conductivity of sample:
Reference system:
Solid electrolyte:
Diaphragm:
Connection:
0–50 °C
≤
2 bar
≥
100 µS/cm
Ag/AgCl
3.5 M KCl (AgCl-free)
open junction
1 m cable with 4-pin connector