Chematest 35
Maintenance
A-96.730.121 / 160920
33
9.6. Cleaning the Conductivity Sensor
To clean the conductivity sensor, proceed as follows:
1
Unscrew the protection sleeve [C].
2
Clean the sensing tip [B] and the inner side of the protection
sleeve [C] with a soft brush and a household detergent.
3
Rinse the sensing tip and the protection sleeve with clean
water.
4
Screw the protection sleeve finger-tight onto the sensor.
9.7. Setting Time and Date
The correct setting of the clock is important for the identification of
stored measurements. It is therefore recommended to check the
date and time setting regularly.
The time and date can be set under the menu item <Menu>/
<Maintenance>/<Set Time>.
A
B
C
Sensor shaft
Sensing tip
Protection sleeve
A
C
B