A-96.730.101 / 151118
Chematest 30
Menu Explanations
3 Maintenance
3.1 Simulation
For test purposes a measurement can be made with one or both
LEDs. The results are displayed as raw values in volts. In case of
problems with the measurement, Swan Support may ask for these
values.
LED Green
Performs a test measurement with the green LED on.
LED IR
Performs a test measurement with the IR LED on.
Both LEDs
Performs three consecutive measurements:
both LEDs off
green LED on
IR LED on
3.2 Verification
Photometer
3.3 Set Time
Adjust date and time.
4 Settings
4.1 Sensors
Photometer
Enter the reference values of the verification cuvettes according to
the enclosed calibration certificate.
4.2 Identification
Optionally, each measurement can be assigned an ID (e.g. for
sampling points) and/or a user name. If activated, the ID or user is
selected from a list at the end of each measurement.
Attribute
Select which attributes are to be queried:
ID List
Enter up to 10 IDs.
User List
Enter up to 10 user names.
Attribute
None
ID
User
ID&User